CHAPTER XXVI

TRANSCRIBER'S NOTE:

In transcribing this book, the proofreaders found and corrected several

minor typographical errors which did not affect the sense of the text.

In the caption to Figure 541, the equation for the voltage of a Weston cell

at different temperatures was missing a digit "1" and this has been

corrected. There is a reference to a Figure 619 but no such figure exists

in the original text. There are references to a Figure 119 and a Figure

443; these presumably exist in one of the preceding volumes of the series.

Throughout, the use of italic type is indicated by _underscores_. In

equations, superscripts are indicated by a caret and braces, as in X^{2}

for "X squared". Subscripts are indicated by underscore and braces, as in

E_{t} to mean "E sub t".

THE THOUGHT IS IN THE QUESTION THE INFORMATION IS IN THE ANSWER

HAWKINS

ELECTRICAL GUIDE

NUMBER THREE

QUESTIONS

ANSWERS

&

ILLUSTRATIONS

A PROGRESSIVE COURSE OF STUDY

FOR ENGINEERS, ELECTRICIANS, STUDENTS

AND THOSE DESIRING TO ACQUIRE A

WORKING KNOWLEDGE OF

ELECTRICITY AND ITS APPLICATIONS

A PRACTICAL TREATISE

by

HAWKINS AND STAFF

THEO. AUDEL & CO. 72 FIFTH AVE. NEW YORK

COPYRIGHTED, 1914,

by

THEO. AUDEL & CO.,

New York.

Printed in the United States.

TABLE OF CONTENTS

GUIDE NO. 3.

GALVANOMETERS 431 to 464

Action of compass needle--simple galvanometer--difference between

galvanoscope and galvanometer--sensibility--action of short and long coil

galvanometers--classes of galvanometer--astatic galvanometer--tangent

galvanometer--graduation of tangent galvanometer scale--table of

galvanometer constants--mechanical explanation of tangent law--sine

galvanometer--table of natural sines and tangents--comparison of

sine and tangent galvanometers--differential galvanometer--ballistic

galvanometer--kick--damping effect--use of mirrors in galvanometers--lamp

and scale--damping--D'Arsonval galvanometer: construction, operation;

uses--galvanometer constant or figure of merit--shunts.

TESTING AND TESTING APPARATUS 465 to 536

Pressure measurement--Clark cell--Weston cadmium cell--pressure

measurement error with ordinary voltmeter--International volt--hydraulic

analogy of amperes--coulombs--current measurement--International

ampere--voltameters--Ohm's law and the ohm--International ohm--ohm

table--practical standards of resistance--various methods of resistance

measurement--direct deflection method--method of substitution--resistance

box--fall of potential method--differential galvanometer method--drop

method--voltmeter method--Wheatstone bridge--usual arrangement of

resistances of Wheatstone bridge--ratio coils of Wheatstone bridge--the

decade plan--two plug arrangement--"plug out" and "plug in" type of

resistance box--testing sets--direct deflection method with Queen Acme

set--ohmmeter--fall of potential method with Queen Acme set--apparatus

for measuring low resistances--how to check a voltmeter--Kelvin wire

bridge--internal resistance measurement--Evershed portable ohmmeter

set--L and N fault finder--ammeter test--diagram of Queen standard

potentiometer--diagrams illustrating loop testing--the Murray loop--the

Varley loop--special loop--the potentiometer--location of opens--to pick

out faulty wires in a cable--voltage of cell measurement with

potentiometer--care of potentiometer--location of faults where the

loop is composed of cables of different cross sections.

AMMETERS, VOLTMETERS, AND WATTMETERS 537 to 572

Definition of ammeter--classification of ammeter and voltmeters--moving

iron type instrument--Keystone voltmeter--winding in ammeters

and volts--connections for series and shunt ammeters--voltmeter

connections--Westinghouse ammeter shunts--various types of

instrument--plunger type instrument--magnetic vane instrument--inclined

coil instrument--Whitney hot wire instruments--principle of

electrostatic instruments--multipliers--portable shunts--Siemens

electro-dynamometer--station instruments--Thompson watt hour meter--how

to read a meter--installation of wattmeters--Westinghouse watt hour

meter--Thompson prepayment watt hour meter--how to test a meter--Sangamo

watt hour meter--Columbia watt hour meter--Duncan watt hour meter.

OPERATION OF DYNAMOS 573 to 596

Before starting a dynamo--adjusting the brushes--brush position--how

to set the brushes--method of soldering cable to carbon brush--brush

contact pressure--direction of rotation--method of winding cables with

marlin--method of assembling core discs--starting a dynamo--tinning block

for electric soldering tool--shunt dynamos in parallel--shunt dynamos

on three wire system--how to start a series machine--the term "build

up"--how to start a shunt or compound machine--"picking up"--indication

of reversed connections--how to correct reversed polarity--finding

the reversed coil--loss of residual magnetism--remedy for reversed

dynamo--attention while running--lead of brushes--method of taking

temperature--lubrication--oils--allowable degree of heating--attention to

brushes and brush gear.

COUPLING OF DYNAMOS 597 to 610

Series and parallel connections--coupling series dynamos in series; in

parallel--equalizer--shunt dynamos in series; in parallel--switching

dynamo into and out of parallel--to cut out a machine--dividing

the lead--compound dynamos in series; in parallel--equalizer

connection--switching a compound dynamo into and out of

parallel--equalizing the load--shunt and compound dynamos in parallel.

DYNAMO FAILS TO EXCITE 611 to 622

Various causes--brushes not properly adjusted--defective

contacts--incorrect adjustment of regulators--speed too low--testing for

break--insufficient residual magnetism; remedy--open circuits--test for

field circuit breakers--probable location of breaks--Watson armature

discs--Fort Wayne commutator truing device--short circuits--Watson

armature--wrong connections--reversed field magnetism.

ARMATURE TROUBLES 623 to 634

Causes--how avoided--various faults--short circuit in individual

coils--location of faulty coil--test for break in armature lead--bar to

bar test for open or short circuit in coil or between segments--short

circuits between adjacent coils--alternate bar test for short circuits

between sections--short circuits between sections through frame or

core of armature; between sections through binding wires--partial short

circuits in armatures--method of testing for breaks--burning of armature

coils--Watson field coils--grounds in armatures--method of locating

grounded armature coil--magneto test for grounded armatures--method of

binding armature winding--breaks in armature circuit.

CARE OF THE COMMUTATOR AND BRUSHES 635 to 652

Conditions for satisfactory operation--oil for commutator--attention

to brushes--Bissell brush gear--two kinds of sparking--commutator

clamp--causes of sparking--bad adjustment of brushes--rocking--bad

condition of brushes--brushes making bad contact--bad condition of

commutator--detection of untrue commutator--high segments--"flats"--causes

of flats; remedy--method of repairing broken joint between commutator

segment and lug--segments loose or knocked in--how to re-turn a

commutator--Bissell commutators--overload of dynamo--method of repairing

large hole burned in two adjacent bars of a commutator--operating dynamos

with metal brushes--indication of excessive voltage--method of smoothing

commutator with a stone--causes of excessive voltage--loose connections,

terminals, etc.,--breaks in armature circuit--sandpaper holder for

commutator--short circuits, in armature circuits; in field--breaks in

field--sandpaper block--short circuits in commutator.

HEATING 653 to 662

Various causes--how detected--procedure--heating, of connections;

of brushes, commutator and armature--excessive heating--ventilated

commutator--self-oiling bearing--some causes of hot bearing--effect of hot

bearings--points relating to hot bearings--operation above rated voltage

and below normal speed--forced system of lubrication--heating of field

magnets--causes of eddy currents in pole pieces--detection of moisture in

field coils--indication of short circuits in field coils.

OPERATION OF MOTORS 663 to 696

Before starting a motor--starting a motor--various starting

resistances--starting boxes--speed regulators--Cutler Hammer starter--time

required to start motor--how to start--sliding contact starters--series

motors on battery circuits--starting a shunt motor--multiple switch

starters--effect of reverse voltage--rheostat with no voltage and

overload release--failure to start--starting panel--Cutler Hammer starting

rheostats--Allen Bradley automatic starter--Monitor starter with relay

for push button control--a remote control of shunt motors--regulation of

motor speed; various methods--Monitor printing press controller--speed

regulation of series motor, by short circuiting sections of the field

winding--varying the speed of shunt and compound motors--Cutler Hammer

multiple switch starter--regulation by armature resistance--Compound

starter--regulation by shunt field resistance--Holzer Cabot instructions

for shunt wound motor--Reliance adjustable speed motor--Cutler Hammer

reversible starter--combined armature and shunt field control--selection

of starters and regulators--Watson commutators--organ blower speed

regulator--General Electric controller--speed regulation of traction

motors--controller of the Rauch and Lang electric vehicles--two motor

regulation--controller connection diagrams--stopping a motor.

CHAPTER XXVI

GALVANOMETERS

If a compass needle be allowed to come to rest in its natural position,

and a current of electricity be passed through a wire just over it from

north to south, the north seeking end of the needle will be deflected

toward the east. If the wire be placed under the needle and the current

continued from north to south the needle will be deflected toward the

west. Again, if the current be passed from north to south over the needle,

and back from south to north under the needle, as shown in fig. 504, the

magnetic effect will be doubled, and the needle deflected proportionately.

Upon these phenomena depend the working of galvanometers.

[Illustration: Fig. 503.--Effect of neighboring current upon a magnetic

needle. Above the needle and parallel to it is a conductor carrying an

electric current, the current flowing in the direction indicated by the

arrow. This causes the north pole of the needle to turn toward the east.

If the conductor be held _below_ the needle, its north pole will turn in

the opposite direction or toward the west. These movements are easily

determined by Ampere's rule as follows: _If a man could swim in the

conductor with the current, and turn to face the needle, then the north

pole of the needle will be deflected toward his left hand_.]

Ques. Describe a simple galvanometer.

Ans. It consists essentially of a magnetic needle suspended within a coil

of wire, and free to swing over the face of a graduated dial.

Ques. What is a galvanoscope and how does it differ from a galvanometer?

Ans. A galvanoscope, as shown in fig. 504, serves merely to indicate the

presence of an electric current without measuring its strength. It is an

indicator of currents where the movement of the needle shows the direction

of the current, and indicates whether it is a strong or a weak one. When

the value of the readings has been determined by experiment or calculation

any galvanoscope becomes a galvanometer.

[Illustration: Fig. 504.--Effect upon a magnetic needle of a neighboring

current in a loop. In this arrangement the same conductor is simply

carried back _beneath_ the needle and hence both the upper and lower

portions tend to turn it in the same direction, while the side branch or

vertical section is ineffective. In accordance with Ampere's swimming

rule, the _upper_ wire causes the N pole of the needle to turn to the

left, while if a man can imagine himself swimming in the lower wire in

the direction of the current, and facing the needle (that is, swimming

on his back), the N pole of the needle will turn to his left--that is to

the east. The effect of the loop then has double the effect of the single

wire in fig. 503.]

Ques. For what use are galvanometers employed?

Ans. They are used for detecting the presence of an electric current, and

for determining its direction and strength.

Ques. How is the direction and strength of the current indicated?

Ans. When a galvanometer is connected in a circuit, the direction of

the current is indicated by the side towards which the north pole of

the needle moves, and the current strength by the extent of the needle's

deflection.

[Illustration: Fig. 505.--Effect upon a magnetic needle of a neighboring

current in a coil. The coil as shown, is equivalent to several loops,

that is, the force tending to deflect the needle is equal to that of a

single loop multiplied by the number of turns. Hence, by using a coil

with a large number of turns, a galvanometer may be made very sensitive

so that the needle will be perceptibly deflected by very feeble currents.

An instrument, as shown in the figure is called a _galvanoscope_. When

it is accurately constructed, and supplied with a scale showing how many

degrees the needle is deflected it is then called a galvanometer.]

Ques. How should a galvanometer be set up before using?

Ans. When no current is flowing, the coil should be parallel to the

magnetic needle when at rest.

Ques. What is a "sensitive" galvanometer?

Ans. One which requires a very small current or pressure to produce a

stated deflection.

It does not follow that a galvanometer which is sensitive for current

measurement will also be sensitive for pressure measurement.

[Illustration: Fig. 506.--Bunnell simple detector galvanometer. It has

middle clamps and scale divided into degrees.]

Ques. Define the term "sensibility."

Ans. With reference to mirror reflecting galvanometers it may be defined

in three ways. First, in _megohms_, the sensibility being the number

of _megohms_ through which one volt will produce a deflection of one

millimeter with the scale at one meter distance. Second, in _micro-volts_,

the sensibility being the number of micro-volts which applied directly

to the terminals of the galvanometer will produce a deflection of one

millimeter on a scale one meter from mirror. The sensibility is best

stated in megohms for high resistance galvanometers and in micro-volts

for low resistance galvanometers, and is frequently given both for

galvanometers for intermediate resistance. Third, in micro-amperes,

the sensibility being the number of micro-amperes that will give one

millimeter deflection with scale at a distance of one meter.

Ques. Upon what does the sensibility depend?

Ans. 1, Upon the number of times the current circulates around the coil,

2, the distance of the needle from the coil, 3, the weight of the needle,

4, the current strength, and 5, the amount of friction produced by its

movement.

[Illustration: Fig. 507.--Breguet upright galvanometer with glass shade.]

[Illustration: Fig. 508.--Bunnell horizontal galvanometer. It has two

coils, one of which is of zero resistance and one of fifty ohms resistance

adapting it to a variety of test.]

The needle is usually quite small, and often a compound one. In very

sensitive galvanometers, the coils are wound with thousands of turns

of very fine wire, and shunts are generally used in connection with

them.

NOTE.--Strong currents must not be passed through very sensitive

galvanometers, for even if they be not ruined, the deflections of the

needle will be too large to give accurate measurements. In such cases the

galvanometer is used with a shunt, or coil of wire arranged so that the

greater part of the current will flow through it, and only a small portion

through the galvanometer.

Ques. What two kinds of coil are used?

Ans. The short coil and the long coil.

Ques. What is the difference between a short coil and a long coil

galvanometer?

Ans. A short coil galvanometer has a coil consisting of a few turns of

heavy wire; a long coil galvanometer is wound with a large number of turns

of fine wire.

[Illustration: Fig. 509.--Bunnell galvanometer for measurements of

instruments, lines, batteries, wires and any object from 1/100 to 10,000

ohms or more.]

Ques. What is the action of short and long coil galvanometers?

Ans. With a given current, the total magnetizing force which deflects

the needle is the same, but with a short coil, it is produced by a

large current circulating around a few turns, instead of a small current

circulating around thousands of turns as in the long coil. The short coil

being of low resistance is used to measure the current, and the long coil

with high resistance, is suitable for measuring the pressure. Hence,

a short coil instrument with its scale directly graduated in amperes

is an _ammeter_, and the long coil type with graduation in volts is a

_voltmeter_.

Classes of Galvanometer.--There are numerous kinds of galvanometer

designed to meet the varied requirements. According to construction,

galvanometers may be divided into two classes, as those having:

1. Movable magnet and stationary coil;

2. Stationary magnet and movable coil.

[Illustration: Fig. 510.--Astatic needles. Two magnetic needles of equal

moment are mounted in opposition on a light support. The whole system is

suspended by a delicate fibre, and when placed in a uniform magnetic field

such as that of the earth, there will be no tendency to assume any fixed

direction, the only restraining influence on the needles being that due

to torsion in the suspension fibre.]

Either type may be constructed with short or long coil, and there are

several ways in which the deflections are indicated. The principal forms

of galvanometer are as follows:

1. Astatic;

2. Tangent;

3. Sine;

4. Differential;

5. Ballistic;

6. D'Arsonval.

Astatic Galvanometer.--It has been pointed out how a compass needle is

affected when a wire carrying a current is held over or under it, the

needle being turned in one direction in the first instance, and in the

opposite direction for the second position of the wire.

[Illustration: Fig. 511.--Connections of single coil astatic needles. The

coil surrounds the lower needle and the direction of the current between

the two needles tends to turn them the same way.]

The earth's magnetism naturally holds the compass needle north and south.

The magnetic field encircling the wire, being at right angles to the

needle (when the wire itself is parallel therewith), operates to turn

it from its normal position, north and south, so as to set it partially

east and west. However, on account of the fact that the earth's magnetism

does exert some force tending to hold the needle north and south, it

is evident that no matter how strong the current, the latter can never

succeed in turning the needle entirely east and west. The accomplishment

of this is further prevented by the reason of the points of the needle,

where the magnetic effect is greatest, quickly passing out of the reach

of the magnetic field, where it is now practically operated on only in a

slight degree. Thus it would take quite a powerful current to hold the

needle deflected any appreciable distance. The use of a shorter needle

is, therefore, more desirable.

It is evident in this style of instrument that the effect of the current

cannot be accurately measured, because it acts in opposition to the

earth's magnetism, and as this is constantly varying, some method must

be employed which will either destroy the earth's magnetism or else

neutralize it.

In the astatic galvanometer, the earth's magnetism is neutralized by

means of _astatic needles_. These consist of a combination of two magnetic

needles of equal size and strength, connected rigidly together with their

poles pointing in opposite and parallel directions, as shown in fig.

510. As the north pole of the earth attracts the south pole of one of the

needles, it repels with equal strength the north pole of the other needle,

hence, the combination is independent of the earth's magnetism and will

remain at rest in any position.

[Illustration: Fig. 512.--Connections of double coil astatic needles. With

this arrangement, the direction of current in both coils will tend to turn

the system in the same direction, making the needles more sensitive than

with a single coil as in fig. 511.]

If one of the needles be surrounded by a coil, as shown in fig. 511,

the magnetic effect of the current will be correctly indicated by the

deflection of the needle.

Sometimes each needle is surrounded by a coil, as in fig. 512, the coils

being so connected that the direction of current in each will tend to

deflect the needles in the same direction.

Ques. For what use is the astatic galvanometer adapted?

Ans. For the detection of small currents.

It is used in the "nil" or zero methods, in which the current between

the points to which the galvanometer is connected is reduced to zero.

[Illustration: Fig. 513.--Queen reflecting astatic galvanometer. It

is mounted on a mahogany base with levelling screws. A plain mirror is

attached above the upper needle. The entire combination of mirror and

needles is suspended by unspun silk from the interior of a brass tube,

which also carries a weak controlling magnet. A dial 4 inches in diameter

and graduated in degrees, enables the deflections of the needle to be

accurately read. The mirror can be used with a reading telescope and

scale, or by means of a lantern, the image of a slit may be reflected from

the mirror to a screen. Resistance, .5 to 1,000 ohms.]

Ques. Upon what does the movement of the needles depend?

Ans. Upon the combined effect of the magnetic attraction of the current

which tends to deflect the needles, and the torsion in the suspension

fibre which tends to keep the needle at the zero position.

Ques. Does the astatic galvanometer give correct readings for different

values of the current?

Ans. When the deflections are _small_ (that is, less than 10° or 15°),

they are very nearly proportional to the strength of the currents that

produce them.

Thus, if a current produce a deflection of 6° it is known to be

approximately three times as strong as a current which only turns the

needle through 2°. But this approximate proportion ceases to be true

if the deflection be more than 15° or 20°.

[Illustration: Fig. 514.--Central Scientific Co. tangent galvanometer.

A 9 inch brass ring is mounted on a mahogany base which rotates on a

tripod provided with levelling screws. The needle has an aluminum pointer

and jewelled bearing. The winding consists of 300 turns of magnet wire

so connected to the plugs in front that 20, 40, 80, or 160 turns or any

combination of these numbers may be used. For heavy currents a band of

copper is used by connecting to the extra pair of binding posts in the

rear of the instrument.]

Ques. Why does the instrument not give accurate readings for large

deflections?

Ans. The needles are not so advantageously acted upon by the current,

since the poles are no longer within the coils, but protrude at the side.

Moreover, the needles being oblique to the force acting on them, part only

of the force is turning them against the directive force of the fibre;

the other part is uselessly pulling or pushing them along their length.

[Illustration: Fig. 515.--Bunnell tangent galvanometer. This instrument

is mounted on a circular hard rubber base, 7-3/8 inches diameter, provided

with levelling screws and anchoring points. The galvanometer consists of a

magnetized needle 7/8 inch in length, suspended at the center of a rubber

ring six inches in diameter, containing the coils. There are five coils of

0, 1, 10, 50 and 150 ohms resistance. The first is a stout copper band of

inappreciable resistance; the others are of different sized copper wires,

carefully insulated. Five terminals are provided, marked, respectively,

0, 1, 10, 50 and 150. The ends of the coils are so arranged that the plug

inserted at the terminal marked 50 puts in circuit all the coils; marked

at the terminal 50--all except the 150 ohm coil; and so on, till at the

zero terminal only the copper band is in circuit. Fixed to the needle,

which is balanced on jewel and point, is an aluminum pointer at right

angles, extending across a five inch dial immediately beneath. One side

of the dial is divided into degrees; on the other side, the graduations

correspond to the tangent of the angles of deflection.]

Ques. How may correct readings be obtained?

Ans. The instrument may be calibrated, that is, it may be ascertained by

special measurements, or by comparison with a standard instrument, the

amounts of deflection corresponding to particular current strengths.

Thus, if it be once known that a deflection of 32° on a particular

galvanometer is produced by a current of 1/100 of an ampere, then a

current of that strength will _always_ produce on that instrument the

same deflection, unless from any accident the torsion force or the

intensity of the magnetic field be altered.

[Illustration: Fig. 516.--Tangent galvanometer. It consists of a short

magnetic needle suspended at the center of a coil of large diameter and

small cross section. In practice, the diameter of the coil is about 17

times the length of the needle. If the instrument be so placed that, when

there is no current in the coil, the suspended magnet lies in the plane

of the coil, that is, if the plane of the coil be set in the magnetic

meridian, then _the current passing through the coil is proportional to

the tangent of the angle by which the magnet is deflected from the plane

of the coil_, or zero position--hence the name: "tangent galvanometer."]

The Tangent Galvanometer.--It is not possible to construct a galvanometer

in which the _angle_ (as measured in degrees of arc) through which the

needle is deflected is proportional throughout its whole range to the

strength of the current. But it is possible to construct a very simple

galvanometer in which the _tangent of the angle of deflection_ shall be

accurately proportional to the strength of the current.

[Illustration: Fig. 517.--Horizontal section through middle of tangent

galvanometer, showing magnetic whirls around the coil and corresponding

deflection of needle.]

[Illustration: Fig. 518.--Diagram of forces acting on the needle of a

tangent galvanometer.]

A simple form of tangent galvanometer is shown in fig. 516. The coil of

this instrument consists of a simple circle of stout copper wire from

ten to fifteen inches in diameter. At the center is delicately suspended

a magnetized steel needle not exceeding one inch in length, and usually

furnished with a light index of aluminum. When the galvanometer is in use,

the plane of the ring must be vertical and in the magnetic meridian. A

horizontal section through the middle of the instrument is shown in fig.

517. For simplicity, the coil is supposed to have but a single turn of

wire, the circles surrounding the wire representing the magnetic lines

of force. By extending the lines of force until they reach the needle,

it will be seen that with a short needle, the deflecting force acts in an

east and west direction when the galvanometer is placed with its coil in

the magnetic meridian.

If, in fig. 518, _ab_ represent the deflecting force acting on the N end

of the needle, the component of this force that acts at a right angle to

the needle will be

_ab_ cos _x_

in which, _x_ is the angle of the deflection.

The controlling force is

_ad_ = H

and when the needle is in equilibrium, the component _ae_ = H sin _x_ is

equal and opposite to _ac_, hence

_ab_ cos _x_ = H sin _x_

from which

_ab_ = H(sin _x_ / cos _x_) = H tan _x_

Since _ab_ is proportional to the current,

_ab_ = _k_ C = H tan _x_

in which _k_ is a constant depending upon the instrument. For any other

current C',

_k_ C' = H tan _x'_

hence

C: C' = tan _x_ : tan _x'_

This means that the currents passing through the coil of a tangent

galvanometer are proportional, not to the angle of deflection, but to the

tangent of that angle.

[Illustration: Fig. 519.--Diagram illustrating the tangent law.

This is the law of the combined action of two magnetic fields upon a

magnetic needle. If two magnetic fields be at right angles in direction

as indicated in the figure, the resultant field is obtained by the

parallelogram of forces and it makes an angle [theta] with one of the

component fields such that tan [theta] = M + H where M and H are the

strengths of the component fields. In the tangent galvanometer this

principle is employed in the measurement of currents. A magnetic needle is

pivoted in a field of known strength. The current to be measured is passed

round a coil (or coils) which generates a field at right angles to the

original field. The needle then lies along the direction of the resultant

field, and by finding the tangent of its angle of deflection, and knowing

the field strength produced by unit current in the coil, the current

strength can be found.]

[Illustration: Fig. 520.--Graduation of tangent galvanometer scale with

divisions representing tangent values. In the figure let a tangent OT

be drawn to the circle, and along this line let any number of equal

divisions be set off, beginning at O. From these points draw lines back

to the center. The circle will thus be divided into a number of spaces, of

which those near O are nearly equal, but which get smaller and smaller as

they recede from O. These unequal spaces correspond to equal increments

of the tangent. If the scale were divided thus, the readings would be

proportional to the tangents.]

Ques. Upon what does the sensibility of a tangent galvanometer depend?

Ans. It is directly proportional to the number of turns of the coil and

inversely proportional to the diameter of the coil.

Ques. How may the tangent galvanometer be used as an ammeter?

Ans. The strength of the current may be calculated in amperes by the

formula given below when the dimensions of the instrument are known.

The needle is supposed to be subject to only the earth's magnetism and to

move in a horizontal plane. The current is calculated as follows:

(1) amperes = ((H × _r_)/N) tan _x_

in which

H = constant from table below;

r = radius of coil;

N = number of turns of coil;

x = angle of deflection of needle.

The constant H, given in the following table represents the horizontal

force of the earth's magnetism for the place where the galvanometer is

used. Each value has been multiplied by (2[pi])/10 so that the formula (1)

for amperes is correct as given.

Table of Galvanometer Constants.--Values of H.

Boston | .699|

Chicago | .759|

Denver | .919|

Jacksonville | 1.094|

London | .745|

Minneapolis | .681|

New York | .744|

New Haven | .731|

Philadelphia | .783|

Portland, Me. | .674|

San Francisco | 1.021|

St. Louis | .871|

Washington | .810|

[Illustration: Fig. 521.--Mechanical explanation of the tangent law.

Construct an apparatus as shown in the figure. The short wooden block,

NS, represents the magnetic needle. This piece of wood turns around its

center, C, which may be an ordinary nail. It will now be seen that two

different forces act upon N; namely, the weight, G (one or two ounces),

and the changeable weights which are placed in the scoop, W (made of

cardboard). The height of the roll, or wheel, R, is such that the cord,

RN, runs horizontally, when NS stands vertically, i.e., when there is no

weight in the little scoop. If the wheel, R, be placed sufficiently far

from NS, the string RN, will always remain almost horizontal, even if NS

be deviated. The thin hand on NS moves over a vertical scale, which is

divided into equal parts, as shown. This scale may be made of cardboard.

If the hand point to division 1 when one ounce is placed in the scoop,

it will point to 2 for two ounces, to 3 for three ounces, etc. At 45°

the needle is deviated at its greatest angle, and this is, therefore, the

sensitivity angle of the tangent galvanometer. The deviating values are,

therefore, proportionate to the scale parts 01, 02, and 03, and so on;

and, inasmuch as these themselves are tangents, the tangent law will hold

good.]

Ques. How is the tangent galvanometer constructed to give direct readings?

Ans. To obviate reference to a table, the circular scale of the instrument

is sometimes graduated into tangent values, as in fig. 520, instead of

being divided into equal degrees.

[Illustration: Fig. 522.--Queen tangent and sine galvanometer. This

instrument properly adjusted can be used as a standard instrument for

laboratory work. The brass ring is 12 inches in diameter, and the grooves

in which the wire is wound are carefully turned so as to be of true

rectangular cross section, thus allowing the constant of the instrument

to be accurately calculated and compared with the constant as obtained by

other methods. The compass box is 5 inches in diameter and is so held in

position that it may be raised or lowered, rotated on its vertical axis,

shifted out of the plane of the coil, etc., thus enabling the operator

to acquire proficiency with the instrument and to meet all cases of

derangement possible. The dial is graduated to single degrees, and the

needle is suspended by a very light cocoon fibre. The whole instrument

can be turned about its vertical axis, and a quadrant graduated in degrees

upon the base allows the amount of rotation to be accurately measured, and

the laws of the sine galvanometer investigated. The instrument is wound

to measure .25 ampere to 8 amperes.]

Ques. What is the objection to the scale with tangent values?

Ans. It is more difficult to divide an arc into tangent lines with

accuracy than into equal degrees.

Ques. What disadvantage has the tangent galvanometer?

Ans. The coil being much larger than the needle, and hence far away from

it, reduces the sensitiveness of the instrument.

The Sine Galvanometer.--This type of instrument has a vertical coil which

may be rotated around a vertical axis, so that it can be made to follow

the magnetic needle in its deflections.

In the sine galvanometer, the coil is moved so as to follow the needle

until it is parallel with the coil. Under these circumstances, the

strength of the deflecting current is proportional to sine of angle

of deflection.

[Illustration: Fig. 523.--Central Scientific Co. universal tangent

galvanometer. This instrument may be used as a tangent, Gaugain,

Helmholtz-Gaugain, sine, cosine, Wiedemann or detector galvanometer. The

coils, which slide on a beam parallel to the one carrying the needle box,

are wound on brass rings 12 inches in diameter. On each ring are wound two

coils of 48 turns each, connected to separate binding posts, and double

wound so as to be of equal resistance. The coils and needle box are each

provided with an indicator for reading their position on the scale. The

needle box is swivelled and removable and one coil may be rotated about

its vertical axis and its position read on a disc graduated in degrees.

Currents may be measured ranging from .000002 ampere to 100 amperes.]

Ques. Describe the construction of a sine galvanometer.

Ans. A form of sine galvanometer is shown in fig. 524. The vertical wire

coil is seen at M. A needle of any length less than the diameter of the

coil M, moves over the graduated circle N. The coil M, and graduated

circle N may be rotated on a vertical axis, and the amount of angular

movement necessary to bring the needle to zero, measured on the graduated

circle H.

Ques. How is the current strength measured?

Ans. It is proportional to the _sine_ of the angle measured on the

horizontal circle H, through which it is necessary to turn the coil M,

from the plane of the earth's magnetic meridian to the plane of the needle

when it is not further deflected by the current.

[Illustration: Fig. 524.--Sine galvanometer. It differs from the tangent

galvanometer in that the vertical coil and magnetic needle are mounted

upon a standard free to revolve around a vertical axis, with provision

for determining the angular position of the coil. The needle may be of any

length shorter than the diameter of the coil. In the figure the parts are:

M, coil; N, graduated dial of magnetic needle; H, graduated dial by which

the amount of rotation necessary to bring the needle to zero is measured;

E, terminals of the coil; O, upright standard carrying coil and graduated

dial of magnetic needle; C, base with levelling screws.]

Ques. How is the sine galvanometer operated?

Ans. In using the instrument, after the needle has been set to zero,

the current is sent through the coil, producing a deflection of the

needle. The coil is then rotated to follow the motion of the needle, the

current being kept constant, the rotation being continued until the zero

on the upper dial again registers with the needle. The current then is

proportional to the sine of the angle through which the coil has been

turned, as determined by the lower dial.

Ques. Has the sine galvanometer a large range?

Ans. For a given controlling field, it does not admit of a very large

range of current measurement, since, for large deflection, on rotating

the coil the position of instability is soon reached.

TABLE OF NATURAL SINES AND TANGENTS

Angle| Sin.| Tan.|

0°| .0000| .0000|

1 | .0175| .0175|

2 | .0349| .0349|

3 | .0523| .0524|

4 | .0698| .0699|

5 | .0871| .0875|

6 | .1045| .1051|

7 | .1219| .1228|

8 | .1392| .1405|

9 | .1564| .1564|

10°| .1736| .1763|

11 | .1908| .1944|

12 | .2079| .2126|

13 | .2250| .2309|

14 | .2419| .2493|

15 | .2588| .2679|

16 | .2756| .2867|

17 | .2924| .3057|

18 | .3090| .3249|

19 | .3256| .3443|

20°| .3420| .3640|

21 | .3584| .3839|

22 | .3746| .4040|

23 | .3907| .4245|

24 | .4067| .4452|

25 | .4226| .4663|

26 | .4384| .4877|

27 | .4540| .5095|

28 | .4695| .5317|

29 | .4848| .5543|

30°| .5000| .5774|

31 | .5150| .6009|

32 | .5299| .6249|

33 | .5446| .6494|

34 | .5592| .6745|

35 | .5736| .7002|

36 | .5878| .7265|

37 | .6018| .7536|

38 | .6157| .7813|

39 | .6293| .8098|

40°| .6428| .8391|

41 | .6561| .8693|

42 | .6691| .9004|

43 | .6820| .9325|

44 | .6947| .9657|

45 | .7071| 1.0000|

46 | .7193| 1.0355|

47 | .7314| 1.0724|

48 | .7431| 1.1106|

49 | .7547| 1.1504|

50°| .7660| 1.1918|

51 | .7771| 1.2349|

52 | .7880| 1.2799|

53 | .7986| 1.3270|

54 | .8090| 1.3764|

55 | .8192| 1.4281|

56 | .8290| 1.4826|

57 | .8387| 1.5399|

58 | .8480| 1.6003|

59 | .8572| 1.6643|

60°| .8660| 1.7321|

61 | .8746| 1.8040|

62 | .8829| 1.8807|

63 | .8910| 1.9626|

64 | .8988| 2.0503|

65 | .9063| 2.1445|

66 | .9135| 2.2460|

67 | .9205| 2.3559|

68 | .9272| 2.4751|

69 | .9339| 2.6051|

70°| .9397| 2.7475|

71 | .9455| 2.9042|

72 | .9511| 3.0772|

73 | .9563| 3.2709|

74 | .9613| 3.4874|

75 | .9659| 3.7321|

76 | .9703| 4.0108|

77 | .9744| 4.3315|

78 | .9781| 4.7046|

79 | .9816| 5.1446|

80°| .9848| 5.6713|

81 | .9877| 6.3138|

82 | .9903| 7.1154|

83 | .9925| 8.1443|

84 | .9945| 9.5144|

85 | .9962| 11.43|

86 | .9976| 14.30|

87 | .9986| 19.08|

88 | .9994| 28.64|

89 | .9998| 57.29|

Ques. What is the position of instability?

Ans. The position of the needle beyond which the rotation of the coil will

cause it to turn all the way round.

Ques. How may the range be increased?

Ans. By an adjustable controlling field or a shunt.

Ques. What advantage has the sine galvanometer over the tangent instrument?

Ans. Its advantage is in the case where the relative values of two

or more currents are required to be measured, or where the constant

of the instrument is obtained by comparison with a standard measuring

instrument and not calculated from the dimensions of the coil, because all

galvanometers thus used follow the sine law independently of the shape of

the coil, while only circular coils will follow the sine law.

[Illustration: Fig. 525.--Differential galvanometer. It consists of two

coils of wire, so wound as to have opposite magnetic effects on a magnetic

needle suspended centrally between them. The needle of a differential

galvanometer shows no deflection when two equal currents are sent through

the coils in opposite directions, since, under these conditions, each

coil neutralizes the effect of the other. Sometimes the current is so

sent through the two coils, that each coil deflects the needle in the

same direction. In this case the instrument is no longer differential in

action. If, when this condition obtains, the magnetic needle be suspended

at the exact center of the line which joins the centers of the coils, the

advantage is gained by obtaining a field of more nearly uniform intensity

around the needle. When the needle is suspended by a silk fibre, a final

and most delicate adjustment can be obtained by raising or lowering one

of the levelling screws slightly, so as to tilt the needle nearer to or

farther from one of the coils.]

The Differential Galvanometer.--This is a form of galvanometer in which

a magnetic needle is suspended between two coils of equal resistance so

wound as to tend to deflect the needle in opposite directions. The needle

of a differential galvanometer shows no deflection when two equal currents

are sent through the coils in opposite directions, since under these

conditions, each coil neutralizes the other's effects. Such instruments

may be used in comparing resistances, although the _Wheatstone bridge_,

in most cases, affords a preferable method.

Ques. What is the special use of the differential galvanometer?

Ans. It is used for comparing two currents.

Ques. What is the method of comparing currents?

Ans. If two equal currents be sent in opposite directions through the

coils of the galvanometer, the needle will not move; if the currents be

unequal, the needle will be deflected by the stronger of them with an

intensity corresponding to the difference of the strengths of the two

currents.

Ques. How are the coils adjusted?

Ans. This is done by coupling them in series in such a way that they tend

to turn the needle in opposite directions, and when a current is passing

through them, they are moved nearer to the needle or farther from it until

the needle stands at zero with any current.

If the coils be not movable, a turn or more can be unwound from the

coil giving the greatest magnetic effect until a balance is obtained,

the wire so unwound can then be coiled in the base of the instrument.

Ballistic Galvanometer.--This type of galvanometer is designed to measure

the strength of momentary currents, such for instance, as the discharge

of a condenser. In construction the magnetic system is given considerable

weight, and arranged to give the least possible _damping effect_.

The term "damping effect" means the offering of a retarding force to

control swinging vibrations, such as the movements of a galvanometer

needle, and to bring them quickly to rest.

If a momentary current be passed through a ballistic galvanometer, the

impulse given to the needle does not cause appreciable movement to the

magnetic system until the current ceases, owing to the inertia of the

heavy moving parts, the result being a slow swing of the needle.

[Illustration: Fig. 526.--Queen dead beat and ballistic reflecting

galvanometer. As illustrated, the coils are easily removable and enclose a

heavy block of copper fixed in a central fork. In a cylindrical hole bored

in this block hangs the bell magnet which with its mirror is suspended

by a long cocoon fibre, and the eddy currents induced in the copper bring

the system quickly to rest after a deflection. By lifting the copper block

out of the frame the instrument is made ballistic. The instrument is made

with coils of any desired resistance up to 1,000 ohms.]

Ques. What name is given to the swing of a ballistic galvanometer needle?

Ans. It is called the _kick_.

Ques. How is the current measured?

Ans. As the needle swings slowly around it adds up, as it were, the

varying impulses received during the passage of the momentary current, and

_the quantity of electricity that has passed is proportional to the sine

of half the angle of the first swing or kick_.

If a reflecting method be used with a straight scale, the observed

deflection depends upon the tangent of twice the angle of movement of

the needle. For small deflections, however, the change of flux can be

taken as directly proportional to the observed deflection.

[Illustration: Fig. 527.--Thompson galvanometer with mirror reflecting

system for reading the deflections of a galvanometer needle by the

movements of a spot of light reflected from a mirror attached to the

needle or movable magnetic system.]

Use of Mirrors in Galvanometers.--In order that small currents may be

measured accurately, some means must be provided to easily read a small

deflection of the needle. Accordingly, it is desirable that the pointer

be very long so that a large number of scale divisions may correspond to

small deflections. In construction, since sensitive galvanometers must

be made with the moving parts of little weight, it would not do to use a

long needle, hence a ray of light is used instead, which is reflected on

a distant scale by a small mirror attached to the moving part.

In the Thompson mirror reflecting galvanometer, as shown in fig. 528,

a small vertical slit is cut in the lamp screen below the scale, and

the ray of light from the lamp, passing through the slit, strikes

the mirror which is about three feet distant, and which reflects the

beam back to the scale. It should be noted that the angle between

the original ray of light and the reflected ray is twice the angle

of the deflection of the mirror; the deflections of the ray of light

on the scale, however, are practically proportional to the strength

of currents through the instrument. The mirror arrangement as shown

in fig. 528, requires a darkened room for its operation, but such is

not necessary when a telescope is used as in fig. 529. Here the scale

readings are reflected in the mirror and their value observed by the

telescope without artificial light.

[Illustration: Fig. 528.--Telescope method of reading galvanometer

deflections by reflection of scale reading in mirror. Here two mirrors

are used, but in most cases the telescope is pointed directly toward the

mirror on galvanometer shown in fig. 527, because the two mirror system,

as illustrated in the figure, is used on portable galvanometers since it

is the more compact.]

Damping.--This relates to the checking or reduction of oscillations.

Thus, a galvanometer is said to be damped when so constructed that any

oscillations of the pointer which may be started, rapidly die away.

Galvanometers are frequently provided with damping devices for the purpose

of annulling these oscillations, thus causing the moving part to assume

its final position as quickly as possible.

Sometimes the instrument is fitted with a damping coil, or closed coil so

arranged with respect to the moving system that the oscillations of the

latter give rise to electric currents in the closed coil, whereby energy

is dissipated. Again, air vanes are employed, but anything in the nature

of solid friction cannot be used.

[Illustration: Figs. 529 and 530.--Galvanometer lamp and scale for

individual use. The scale is etched on a ground glass strip 6 centimeters

wide by 60 centimeters long with long centimeter divisions and short

millimeter divisions the entire length, reading both ways from zero in the

center. It is mounted in an adjustable wooden frame. A straight filament

lamp (110 volts) is enclosed in a metal hood japanned black to cut out all

reflected light. This form of filament makes a single brilliant line on

the scale, enabling closer readings than the "spot of light" arrangement.

The lamp hood is adjustable to any desired height on the support rod.]

D'Arsonval Galvanometer.--This instrument has a movable coil in place of a

needle, and its operation depends upon the principle that if a flat coil

of wire be suspended with its axis perpendicular to a strong magnetic

field, it will be deflected whenever a current of electricity passes

through it.

Ques. Describe the construction of a D'Arsonval galvanometer.

Ans. The essential features are shown in figs. 532 and 533. The coil,

which is rectangular in section is wound upon a copper form, and suspended

between a permanent magnet by fine wires to the points A and B. The magnet

has its poles at N and S. It is a soft iron cylinder fixed between the

poles in order to intensify the magnetic field across the air gaps in

which the coil moves.

[Illustration: Fig. 531.--Queen reading telescope. This arrangement is

utilized to measure the deflections of a galvanometer having suspended

mirror moving system. It consists of a reading telescope mounted as

illustrated with a millimeter scale, having a length of 50 centimeters.

In use, the image of the scale is seen in the galvanometer mirror through

the telescope. The eye piece of the telescope has a cross hair which acts

as a reference line so that by noting the particular division on the scale

when the galvanometer is at rest, the amount of deflection can be readily

observed when the galvanometer is deflected. The instrument has all the

necessary adjustments to set it up quickly and for bringing the cross hair

and scale in focus. It is generally placed at a distance of one meter from

the galvanometer mirror.]

Ques. Explain its operation.

Ans. An enlarged horizontal cross section of the galvanometer on line XY

is shown in fig. 533. The current is flowing in the coil as in fig. 532,

up on the left side and down on the right. The position of the coil when

no current is flowing is indicated by _n' s'_. By applying the law of

mutual action between magnetic poles, it is seen that when the current is

applied, the poles developed at _n' s'_ will move into the position _n''

s''_. See fig. 119.

Ques. How is the coil affected by a change in the direction of the current?

Ans. The polarity of the coil is reversed and consequently the direction

of the deflection.

[Illustration: Figs. 532 and 533.--Diagrams showing essential features of

construction and principle of operation of D'Arsonval galvanometer.]

Ques. Upon what does the sensitiveness of the instrument depend?

Ans. Upon the strength of the field of the permanent magnet, the number

of turns in the suspended coil, and the torsion of the wires by which it

is suspended.

Ques. When is this galvanometer called "dead beat"?

Ans. When the construction is such that the moving part comes quickly to

rest without a series of diminishing vibrations.

[Illustration: Figs. 534 to 536.--Queen horizontal magnet D'Arsonval

galvanometer with telescope and scale. It is very sensitive and is used

in many electrical measurements, including commercial testing, such as

measuring insulation of cables, fault location, etc. It is not affected

by surrounding magnetic disturbances, and may, therefore, be used in

proximity to dynamos and switchboards. The instrument has a pair of

binding post terminals, one of which connects to a bottom spiral of the

system and the other forms a junction with the top of the tube holding the

system, forming a complete circuit through the coil. The tube containing

the system may be readily removed from the magnet and another tube having

a different system inserted as is required for various kinds of electrical

measurement. The entire system with its suspension may be inspected by

the removal of a thumb screw. To inspect interior of tube first be sure

that the screw B is turned so that the coil is clamped. Entirely remove

screw C, and, holding the outside tube near the window, press firmly with

the finger on the extreme top of the suspension support. The inside rib,

with complete suspension, will draw from the tube, and the working parts

can be fully inspected. Carefully return same to its original position in

tube, setting tight the screw C. The galvanometer is designed so that the

coil is clamped in position when the galvanometer must be transported.

The insulation of the galvanometer terminals and binding posts is such

as to guard against any possible leakage. As a further protection, each

levelling screw is provided with a hard rubber insulator. This feature

is essential since, in making insulation measurements, the operator

wishes to be assured that the deflection being obtained is the result

of leakage upon the cable or wire being measured and not leakage between

the galvanometer terminals. The galvanometer is provided with an attached

telescope and scale for noting the deflections. The deflections produced

by this galvanometer are proportional to the current. To facilitate

quickly setting up the instrument, two way levels are provided.]

Ques. What causes this?

Ans. The instrument is made dead beat by winding the coil on a copper or

aluminum frame, so that when in operation, currents are induced in the

frame by the motion of the coil in the magnetic field; these currents

oppose the motion of the coil.

Ques. For what service is the D'Arsonval galvanometer adapted?

Ans. It is desirable for general use as it is not much affected by

changes in the magnetic field. It may be made with high enough period and

sensibility to be satisfactory as a ballistic instrument, but for extreme

sensibility an instrument of the astatic type is more generally used.

Galvanometer "Constant" or "Figure of Merit."--In order that a

galvanometer shall be of value as a measuring instrument, the relation

between the current and the deflection produced by it must be known. This

may be obtained experimentally by determining the value of the current

required to produce one scale division. The galvanometer constant then may

be defined as _the resistance through which the galvanometer will give a

deflection of one scale division when the current applied is at a pressure

of one volt_.

Accordingly, the deflection as indicated on the scale must be multiplied

by its constant or figure of merit, in order to obtain the correct

reading. If the scale readings be not directly proportional to the

quantity to be measured, the law of the instrument must also be

considered.

Thus in a tangent galvanometer as previously explained

I = K tan [phi]

where I = current, [phi] the deflection or scale reading, and K the

galvanometer constant.

[Illustration: Fig. 537.--Diagram showing method of connecting

galvanometer shunt. By the use of a shunt the range of measurement of a

galvanometer can be greatly increased.]

[Illustration: Fig. 538.--Diagram of a form of universal shunt box for use

with galvanometers of widely different resistances. The galvanometer, as

indicated at G, is connected across the ends of a series of resistances

AB. The main wires are connected, one to end A of the series and the other

to a travelling point whose position is varied by means of plugs or by a

dial switch.]

Galvanometer Shunts.--The sensitiveness of a galvanometer used for

measuring current may be reduced to any desired extent by connecting a

resistance of known value in parallel with it. Thus, if it be desired

to measure a current greater than can be measured directly by the

galvanometer, a part of the current can be sent through the resistance or

shunt, and the total value of the current calculated.

A galvanometer shunt bears a definite ratio to the resistance of the

galvanometer, being usually adjusted so that only .1, .01, or .001 part

of the current passes through the galvanometer.

The degree in which a shunt increases the range of deflection of a

galvanometer is called its "multiplying power."

[Illustration: Fig. 539.--Ayrton-Mather universal shunt. This shunt may

be used with any galvanometer. The total resistance is 10,000 ohms, with

shunt powers of 1, 5, 10, 50, 100, 500, and 1,000. It is also fitted with

positions in which the galvanometer is shorted and off. The coils are of

constantan wire.]

If .1 of the current flowing, passed through the galvanometer and

.9 through the shunt, then the current in the circuit would be ten

times that through the galvanometer. Accordingly the current in the

galvanometer must be multiplied by the multiplying power of the shunt

to obtain the true value of the current in the circuit.

In order to determine the resistance necessary to be used with a certain

galvanometer, the resistance of the latter _is to be divided by the

multiplying power desired less one_.

EXAMPLE.--What must be the resistance of a shunt for a galvanometer of

2,000 ohms resistance where only one fifth of the current is to pass

through the galvanometer?

The multiplying power less one is

5 - 1 = 4

and the required resistance is

2,000 ÷ 4 = 500 ohms.

When it is essential that the total resistance of the circuit should not

be altered by an alternation of the galvanometer shunt, a compensating

box should be used which automatically inserts a resistance for each shunt

in series with the shunted galvanometer to bring the total resistance up

equal to the unshunted value. Thus the current in the main circuit is not

altered.

Hawkins Electrical Guide v. 03 (of 10) Questions, Answers, & Illustrations, A progressive course of study for engineers, electricians, students and those desiring to acquire a working knowledge of electricity and its applications · 第 1 / 10 章