ANALYSIS OF RESIDUAL AIR.

The carbon-dioxide production, water-vapor elimination, and oxygen

absorption of the subject during 1 or 2 hour periods are recorded in a

general way by the amounts of carbon dioxide and water-vapor absorbed by

the purifying vessels and the loss of weight of the oxygen cylinder;

but, as a matter of fact, there may be considerable fluctuations in the

amounts of carbon dioxide and water-vapor and particularly oxygen in the

large volume of residual air inside the chamber. With carbon dioxide and

water-vapor this is not as noticeable as with oxygen, for in the 1,300

liters of air in the chamber there are some 250 liters of oxygen, and

slight changes in the composition of this air indicate considerable

changes in the amount of oxygen. Great changes may also take place in

the amounts of carbon dioxide and water-vapor under certain conditions.

In some experiments, particularly where there are variations in muscular

activity from period to period, there may be a considerable amount of

carbon dioxide in the residual air and during the next period, when the

muscular activity is decreased, for example, the percentage composition

of the air may vary so much as to indicate a distinct fall in the amount

of carbon dioxide present. Under ordinary conditions of ventilation

during rest experiments the quantity of carbon dioxide present in the

residual air is not far from 8 to 10 grams. There are usually present in

the air not far from 6 to 9 grams of water-vapor, and hence this

residual amount can undergo considerable fluctuations. When it is

considered that an attempt is made to measure the total amount of carbon

dioxide expired in one hour to the fraction of a gram, it is obvious

that fluctuations in the composition of residual air must be taken into

consideration.

It is extremely difficult to get a fair sample of air from the chamber.

The air entering the chamber is free from water-vapor and carbon

dioxide. In the immediate vicinity of the entering air-tube there is air

which has a much lower percentage of carbon dioxide and water-vapor than

the average, and on the other hand close to the nose and mouth of the

subject there is air of a much higher percentage of carbon dioxide and

water-vapor than the average. It has been assumed that the composition

of the air leaving the chamber represents the average composition of the

air in the chamber. This assumption is only in part true, but in rest

experiments (and by far the largest number of experiments are rest

experiments) the changes in the composition of the residual air are so

slow and so small that this assumption is safe for all practical

purposes.

Another difficulty presents itself in the matter of determining the

amount of carbon dioxide and water-vapor; that is, to make a

satisfactory analysis of air without withdrawing too great a volume from

the chamber. The difficulty in analysis is almost wholly confined to the

determination of water-vapor, for while there are a large number of

methods for determining small amounts of carbon dioxide with great

accuracy, the method for determining water-vapor to be accurate calls

for the use of rather large quantities of air. From preliminary

experiments with a sling psychrometer it was found that its use was

precluded by the space required to successfully use this instrument, the

addition of an unknown amount of water to the chamber from the wet bulb,

and the difficulties of reading the instrument from without the chamber.

Recourse was had to the determination of moisture by the absolute

method, in that a definite amount of air is caused to pass over

pumice-stone saturated with sulphuric acid. It is of interest here to

record that at the moment of writing a series of experiments are in

progress in which an attempt is being made to use a hair hygrometer for

this purpose.

The method of determining the water-vapor and carbon dioxide in the

residual air is extremely simple, in that a definite volume of air is

caused to pass over sulphuric acid and soda-lime contained in U-tubes.

In other words, a small amount of air is caused to pass through a small

absorbing-system constructed of U-tubes rather than of porcelain vessels

and silver-plated cans. Formerly a very elaborate apparatus was employed

for aspirating the air from the chamber through U-tubes and then

returning the aspirated air to the chamber. This involved the use of a

suction-pump and called for a special installation for maintaining the

pressure of water constant. More recently a much simpler device has been

employed, in that we have taken advantage of the pressure in the

ventilating air-system developed by the passage of air through the

blower. After forcing a definite quantity of air through the reagents in

the U-tubes, it is then conducted back to the system after having been

measured in a gas-meter.

This procedure is best noted from fig. 30. The connected series of three

U-tubes on the rack on the table is joined on one end by well-fitting

rubber connections to the tube leading from the mercurial manometer and

on the other end to the rubber tube A leading to the gas-meter. On

lowering the mercury reservoir E, the mercury is drained out of the tube

D and air passes through both arms of the tube and then through the

three U-tubes. In the first of these it is deprived of moisture, and in

the last two of carbon dioxide. The air then enters the meter, where it

is measured and leaves the meter through the tube B, saturated with

water-vapor at the room temperature. To remove this water-vapor the air

is passed through a tower filled with pumice-stone drenched with

sulphuric acid. It leaves the tower through the tube C and enters the

ventilating air-pipe on its way to the calorimeter.

The method of manipulation is very simple. After connecting the U-tubes

the pet-cock connecting the tube C with the pipe is opened, the mercury

reservoir E is lowered, and air is allowed to pass through until the

meter registers 10 liters. By raising the reservoir E the air supply is

shut off, and after closing the stop-cock at C the tubes are

disconnected, a second set is put in place, and the operation repeated.

The U-tubes are of a size having a total length of the glass portion

equal to 270 millimeters and an internal diameter of 16 millimeters.

They permit the passage of 3 liters of air per minute through them

without a noticeable escape of water-vapor or carbon dioxide. The

U-tubes filled with pumice-stone and sulphuric acid weigh 90 grams. They

are always weighed on the balance with a counterpoise, but no attempt is

made to weigh them closer than to 0.5 milligram.

Respiration Calorimeters for Studying the Respiratory Exchange and Energy Transformations of Man · 第 93 / 120 题