CALIBRATIONS

_Calorimeter_

At the conclusion of these experiments, the accuracy of our

calorimetry apparatus was tested by burning an ethanol lamp in the

metabolism chamber. During these tests a CO_{2} analyzer was

incorporated into the system (Beckman, LB-2). Results demonstrated

that we measured 84% of the oxygen consumed by the lamp as well as 84%

of the water and CO_{2} it produced; standard deviation = +-2.6, +-5.0,

and +-3.6, respectively (n = 27). Average respiratory quotient (RQ)

calculated from these data was O.657 +-0.008 (n = 27), which is 99.5%

of that predicted (0.66). McNab (1988b) reports that the accuracy of

open-flow indirect calorimetry systems, such as ours, depends on the

rate of air flow through the animal chamber. If flow rates are too

low, there is inadequate mixing of air within the chamber, and the

rate of oxygen consumption, as calculated from the difference in

oxygen content of air flowing into and out of the chamber (Depocas and

Hart, 1957), is underestimated. At some critical rate of air flow,

which is unique to each combination of chamber and animal, this

situation changes such that measured rates of oxygen consumption

become independent of any further increase in flow rate (McNab,

1988b). In recent tests of our system, where we burned the ethanol

lamp at a variety of chamber flow rates, the efficiency of

measurement increased linearly as flow rate increased, and the

critical rate of air flow was about 6.7 L/min. This appeared to

explain why a flow rate of 3.0 L/min underestimated oxygen consumption

of the ethanol lamp.

Our earlier tests of the efficiency of our system indicated that

although we underestimated actual oxygen consumption of the ethanol

lamp, we did so with a fair degree of precision; probably because flow

rates were closely controlled. During our metabolic measurements,

chamber flow rates also were closely controlled at 3.0 L/min, and we

believe, therefore, that these measurements also were carried out with

a high degree of precision. Consequently, all measured values of

oxygen consumption and water production were considered to be 84% of

their actual value and were adjusted to 100% before being included in

this report.

_Body Temperature Transmitters_

The calibration of all temperature-sensitive radio transmitters

drifted over time. Transmitters were calibrated before they were

surgically implanted and again after they were removed from the

animals. Although the drift of each transmitter was unique, it was

also linear (S. Tomkiewicz, Telonics, Inc., pers. com.). All body

temperature measurements were corrected from timed extrapolations of

the difference between starting and ending calibrations.