CHAPTER III

THERMO-ELECTRIC PYROMETERS

=General Principles.=—Seebeck, in 1822, made the discovery that

when a junction of two dissimilar metals is heated an electromotive

force is set up at the junction, which gives rise to a current

of electricity when the heated junction forms part of a closed

circuit. Becquerel, in 1826, attempted to apply this discovery to

the measurement of high temperatures, it having been observed that

in general the E.M.F. increased as the temperature of the junction

was raised. No concordant results were obtained, and the same fate

befell the investigations of others who subsequently attempted to

produce pyrometers based on the Seebeck effect. These failures were

due to several causes, but chiefly to the non-existence of reliable

galvanometers, such as we now possess. It was not until 1886 that the

problem was satisfactorily solved by Le Chatelier of Paris.

Although any heated junction of metals will give rise to an

electromotive force, it does not follow that any pair, taken at random,

will be suited to the purposes of a pyrometer. A junction of iron and

copper, for example, gives rise to an E.M.F. which increases with the

temperature up to a certain point, beyond which the E.M.F. falls off

although the temperature rises, and finally reverses in direction—a

phenomenon to which the name of “thermo-electric inversion” has been

applied. Evidently, it would be impossible to measure temperatures

in this case from observations of the electromotive force produced,

and any couple chosen must be free from this deterrent property.

Moreover, the metals used must not undergo deterioration, or alteration

in thermo-electric properties, when subjected for a prolonged

period to the temperature it is desired to measure. These and other

considerations greatly restrict the choice of a suitable pair of

metals, which, to give satisfaction, should conform to the following

conditions:—

1. The E.M.F. developed by the junction should increase uniformly as

the temperature rises.

2. The melting point of either component should be well above the

highest temperature to be measured. An exception to this rule

occurs when the E.M.F. of fused materials is employed.

3. The thermo-electric value of the couple should not be altered by

prolonged heating.

4. The metals should be capable of being drawn into homogeneous

wires, so that a junction, wherever formed, may always give rise

to the same E.M.F. under given conditions.

It is a further advantage if the metals which fulfil the above

conditions are cheap and durable.

The exacting character of these requirements delayed the production

of a reliable thermo-electric pyrometer until 1886, when Le Chatelier

discovered that a junction formed of platinum as one metal, and an

alloy of 90 per cent. of platinum and 10 per cent. of rhodium as the

other, gave concordant results. In measuring the E.M.F. produced, Le

Chatelier took advantage of the moving-coil galvanometer introduced by

d’Arsonval, which possessed the advantages of an evenly-divided scale

and a dead-beat action. This happy combination of a suitable junction

with a simple and satisfactory indicator immediately established the

reliability of the thermo-electric method of measuring temperatures.

As platinum melts at 1755° C., and the rhodium alloy at a still higher

temperature, a means was thus provided of controlling most of the

industrial operations carried out in furnaces.

So far, the effect of heating the junction has been considered without

regard to the temperature of the remainder of the circuit, and it is

necessary, before describing the construction of practical instruments,

to consider the laws governing the thermo-electric circuit, the

simplest form of which is represented in fig. 2. One of the wires is

connected at both ends to separate pieces of the other wire, the free

ends of which are taken to the galvanometer Two junctions, A and B, are

thus formed, which evidently act in opposition; for if on heating A the

direction of current be from A to B, then on heating B the direction

will be from B to A. Hence if A and B were equally heated no current

would flow in the circuit, the arrangement being equivalent to two

cells of equal E.M.F. in opposition. Thermal junctions are formed at

each of the galvanometer terminals, but the currents to which they give

rise, when the temperature changes, are opposed and cancel each other.

The law which holds for this circuit may be expressed thus:—

“If in a thermo-electric circuit there be two junctions, A and B,

the electromotive force developed is proportional to the

_difference_ in temperature between A and B.”

[Illustration: FIG. 2.—TWO-JUNCTION THERMO-ELECTRIC CIRCUIT.]

It is customary to refer to the two junctions as the “hot” and “cold”

junctions; but it is important to remember that fluctuations in the

temperature of either will alter the reading on the galvanometer or

indicator.

A second law, which applies to all thermo-electric circuits, is that

“the E.M.F. developed is independent of the thickness of the wire.”

This does not mean that the deflection of the galvanometer is the

same whether thin or thick wires are used to form the junction. The

deflection depends upon the current flowing through the circuit, and

this, according to Ohm’s law, varies inversely as the total resistance

of the circuit. Consequently, the use of thin wires of a given kind

will tend to give a less deflection than in the case of thick wires, as

the resistance of the former will be greater, and unless the resistance

of the galvanometer be great compared with that of the junction, the

difference in deflection will be conspicuous. The E.M.F., however, is

the same under given conditions, whatever thickness of wire be used.

Reference to fig. 2 will show that in order to realise this circuit in

practice, one of the wires forming the couple must be used in the form

of leads to the galvanometer. This can readily be done if the material

of the wire is cheap; but if platinum or other expensive metal be

used, and the galvanometer be some yards distant, the question of cost

necessitates a compromise, and the circuit is then arranged as in fig.

3. The wires forming the hot junction are brought to brass terminals T

T, from which copper wires lead to the galvanometer G. This arrangement

results in three effective junctions, viz. the hot junction A to B; the

junction A to brass, and the junction B to brass. It will be seen that

the two junctions of copper to brass are in opposition, and cancel

each other for equal heating; and the same applies to the galvanometer

connections. A circuit thus composed of three separate junctions does

not permit of a simple expression for the net E.M.F. under varying

temperature conditions, and to avoid errors in readings care must be

taken to prevent any notable change of temperature at the terminals T T

in a practical instrument arranged as in the diagram.

[Illustration: FIG. 3.—THREE-JUNCTION THERMO-ELECTRIC CIRCUIT.]

A point of practical utility in thermo-electric work is the fact that

if a wire be interrupted by a length of other metal, as indicated at C

in fig. 3, no current will be set up in a circuit if both joints are

equally heated, as the electromotive forces generated at each junction

are in opposition. It is thus possible to interrupt a circuit by a

plug-key or switch, without introducing an error; always provided that

an even temperature prevails over the region containing the joints.

Another useful fact is that if two wires be brought into contact, they

may be fastened over the joint by soldering or using a third metal,

without alteration of thermo-electric value, except in rare cases. Thus

a copper-constantan or iron-constantan junction may suitably be united

by silver solder, using borax as a flux, thus avoiding the uncertainty

of contact which must always occur when the wires are merely twisted

together. Welding, however, is preferable to soldering.

=Metals used for Thermal Junctions.=—Until recent years

it was customary to employ a platinum-rhodioplatinum or

platinum-iridioplatinum junction for all temperatures beyond the

scope of the mercury thermometer. The almost prohibitive price of

these metals has caused investigations to be made with a view to

discovering cheaper substitutes, with successful results up to 1000°

C. or 1800° F., thus comprehending the range of temperatures employed

in many industrial processes. Above this temperature the platinum

series of metals are still used for accurate working, but it will be

of great advantage if the range measurable by cheap or “base” metals

can be further extended. Promise in this direction is afforded by

the properties of fused metals when used in thermal junctions. An

investigation by the author has shown that in general the E.M.F.

developed by a junction does not undergo any sudden change when one

or both metals melt, but continues as if fusion had not occurred. By

making arrangements to maintain the continuity of the circuit after

fusion, it may be possible to read temperatures approximating to the

boiling points of metals such as copper and tin, both of which are over

2000° C. The base metals are not so durable as platinum and kindred

metals, but as the cost of replacement is negligible, this drawback is

of little importance. Moreover, base-metal junctions usually develop a

much higher E.M.F. than the platinum metals, which enables stronger and

cheaper galvanometers to be used as indicators.

THERMAL JUNCTIONS USED IN PYROMETERS.

─────────────────────────────────────────────────┬────────────────

│ Upper limit to

│ which Junction

│ may be used.

Couple. ├────────┬───────

│ Deg. │ Deg.

│ Cent. │ Fahr.

─────────────────────────────────────────────────┼────────┼───────

Platinum and rhodioplatinum (10 per cent. Rh) │ 1400 │ 2550

2 Rhodioplatinum alloys of different composition │ 1600 │ 2900

Platinum and iridioplatinum (10 per cent. Ir) │ 1100 │ 2000

Nickel and constantan │ 900 │ 1650

Nickel and copper │ 800 │ 1475

Nickel and carbon │ 1000 │ 1850

Nickel and iron │ 1000 │ 1850

Iron and constantan │ 900 │ 1650

Copper and constantan │ 800 │ 1475

Silver and constantan │ 800 │ 1475

2 Nickel chrome alloys of different composition │ │

(Hoskin’s alloys) │ 1100 │ 2010

Nickel-chrome alloy and nickel-aluminium alloy │ 1100 │ 2010

2 Iron-nickel alloys of different composition │ 1000 │ 1850

─────────────────────────────────────────────────┴────────┴───────

The electromotive force developed by a junction of any given pair of

metals when heated to a given temperature varies according to the

origin of the metals. It is not unusual, for example, for two samples

of 10 per cent. rhodioplatinum, obtained from different sources, to

show a difference in this respect of 40 per cent. when coupled with

the same piece of platinum. Equal or greater divergences may be noted

with other metals; and hence the replacement of a junction can only be

effected, with accuracy, by wires from the same lengths of which the

junction formed a part. As showing how platinum itself is not uniform,

it may be mentioned that almost any two pieces of platinum wire, if

not from the same length, will cause a deflection on a sensitive

galvanometer when made into a junction and heated. It is therefore

customary for makers to obtain considerable quantities of wire of a

given kind, homogeneous as far as possible, in order that a number of

identical instruments may be made, and the junctions replaced, when

necessary, without alteration of the scale of the indicator.

The alloy known as “constantan,” which figures largely in the foregoing

table, is composed of nickel and copper, and is practically identical

with the alloy sold as “Eureka” or “Advance.” It has a high specific

resistance, and a very small temperature coefficient, and is much used

for winding resistances. Couples formed of constantan and other metals

furnish on heating an E.M.F. several times greater than that yielded

by couples of the platinum series, and show an equally steady rise

of E.M.F. with temperature. This alloy has proved of great service in

connection with the thermo-electric method of measuring temperatures.

Couples formed of nickel-chrome alloys, known as “Hoskin’s alloys,”

have been introduced into Britain by the Foster Instrument Company,

which may be used continuously to 1100° C., and for occasional readings

up to 1300° C. Another couple, much used in America, consists of an

alloy of 90 per cent. nickel and 10 per cent. chromium, and an alloy

of 98 per cent. nickel and 2 per cent. aluminium, which may be used up

to 1100° C. Other couples, formed of alloys of nickel, chromium, iron,

aluminium, etc., have been introduced by different makers, but have not

proved so satisfactory as those mentioned above.

=Changes in Thermal Junctions when constantly used.=—No metal

appears to be able to withstand a high temperature continuously without

undergoing some physical alteration; and for this reason the E.M.F.

developed by a given junction is liable to change after a period of

constant use. At temperatures above 1100° C., platinum, for example,

undergoes a notable change in a comparatively short period, but below

1000° C., the change is very slight, and if this range be not exceeded,

a platinum-rhodioplatinum or iridioplatinum junction may be used for

years without serious error arising from this cause. This liability

to change is one of the factors which restricts the range of thermal

junctions, which should never be used continuously beyond the

temperature at which the alteration commences to become large. A second

cause of discrepancy is the possible alteration in the composition

of an alloy, due to one of the constituents leaving in the form of

vapour, as is noted with iridioplatinum alloys, from which the iridium

volatilises in tangible quantities above 1100° C., causing a fall of

10 per cent. or more in the thermo-electric value of the junction of

these alloys with platinum. Constantan appears to be very stable in

its thermo-electric properties, and the various junctions in which

it plays a part show a high degree of stability if not overheated.

Rhodioplatinum alloys are very stable, and for temperatures exceeding

1100° C. a junction of two of these alloys, of different composition,

is more durable than one in which pure platinum is used. An extended

series of tests on base-metal junctions made in America by Kowalke

showed that continuous heating of couples as received from the makers

altered the E.M.F. considerably, the change in some cases representing

over 100° C. on the indicator. A stable condition, due to the relief of

strains or other change, was finally reached, and the conclusion drawn

that the materials should be thoroughly annealed before calibration. It

is desirable in all cases periodically to test the junctions at some

standard temperature, and if any conspicuous error be noted, to replace

the old junction by a new one.

In addition to the errors due to slow physical changes, a junction

may be altered considerably, if imperfectly protected, owing to the

chemical action of furnace gases, or of solids with which the junction

may come into contact. The vapours of metals such as lead or antimony

are very injurious; and platinum in particular is seriously affected

by vapours containing phosphorus, if in a reducing atmosphere. So

searching is the corrosive action of furnace gases that adequate

protection of the junction is essential if errors and damage are to be

avoided. When a wire has once been corroded, a junction made with it

will not develop the same E.M.F. as before.

=Electromotive Force developed by Typical Junctions.=—The

following table exhibits the E.M.F. generated by several junctions

for a range of 100° C., taken at the middle part of the working range

in each case. These figures are subject to considerable variation,

according to the origin of the metals.

──────────────────────────────────────────┬─────────────────────────

│ E.M.F. in millivolts for

Couple. │ a rise of 100° at middle

│ of working range.

──────────────────────────────────────────┼─────────────────────────

Platinum-rhodioplatinum (10 per cent. Rh) │ 1·1

Platinum-iridioplatinum (10 per cent. Ir) │ 1·2

Nickel-constantan │ 2·3

Copper-constantan │ 5·8

Nickel-copper │ 6·1

Iron-constantan │ 6·7

Hoskin’s alloys │ 7·4

──────────────────────────────────────────┴─────────────────────────

It will be noted that the base-metal junctions give much higher values

than the platinum series, and hence can be used with a less sensitive,

and therefore cheaper, indicator. Base-metal junctions are also, in

consequence of the greater E.M.F. furnished, capable of yielding more

sensitive readings over a selected range of temperature.

[Illustration: FIG. 4.—PRACTICAL FORM OF THERMO-ELECTRIC PYROMETER.]

=Practical Forms of Thermocouples.=—When expensive junctions are

employed, wires of the minimum thickness consistent with strength and

convenience of construction are used, a diameter of No. 25 standard

wire gauge being suitable. A common arrangement is shown in fig. 4, in

which J is the hot junction, the wires from which are passed through

thin fireclay tubes which serve as insulators (or through twin-bore

fireclay) to the reels R R, in the head of the pyrometer, upon which

a quantity of spare wire is wound to enable new junctions to be made

when required. Two brass strips, S, are screwed down on to the wires

at one end, and are furnished with screw terminals at the other

end, from which wires are taken to the galvanometer or indicator. A

protecting-tube, T, surrounds the wires and hot junction. The head,

H, may be constructed of wood, fibre, or porcelain, and should be an

insulator for electricity and heat. There are various modifications

in use, but the general method described is adopted by most makers.

In order to guard against errors arising from alterations in the

temperature of the cold junctions in the end of the pyrometer, some

firms construct the head so as to leave a hollow space, through which

cold water is constantly circulated (fig. 5), the arrangement being

known as a “water-cooled head.” In some forms the supply of spare wires

is made to take the form of two spiral springs in a hollow head, the

upper ends of the springs being taken to terminals.

[Illustration: FIG. 5.—PYROMETER WITH WATER-COOLED HEAD.]

The choice of a protecting-tube is a matter of considerable importance.

Obviously, such a tube should not soften at the highest temperature

attained, and when expensive metals are used to form the junction the

sheath should not be permeable to gases or vapours. It should also, if

possible, be a good conductor of heat, so that the junction may respond

quickly to a change of temperature in its surroundings, and should be

mechanically strong. It is difficult to secure all these properties

in any single material, and the choice of a sheath is decided by

the conditions under which the couple is to be used. The substances

employed, and their properties and special uses, may be enumerated as

follows:—

1. _Iron or Mild Steel._—For temperatures not exceeding 1100° C.

iron or mild steel covers are cheap and efficient from the standpoint

of conductivity, although liable to deteriorate owing to oxidation.

The tendency to oxidise is greatly diminished by “calorising” the

exterior by Ruder’s process, in which the iron is heated in a mixture

of metallic aluminium and oxide of aluminium, a surface alloy being

formed which resists oxidation. A result nearly as good may be obtained

by smearing the surface with fine aluminium powder, and bringing to a

white heat. This treatment greatly prolongs the life of an iron sheath.

Some makers employ an inner steel tube round the wires, and an outer

tube which comes into contact with the furnace gases, corrosion of the

latter being detected before the inner tube has given way and exposed

the junction. Some makers do not consider it safe to expose heated

platinum to an iron surface, with only air intervening, and hence use

an inner cover of silica or porcelain, which the outer iron or steel

tube protects from mechanical damage. For ordinary work seamless steam

or hydraulic steel tubing, with a welded end, is satisfactory; but for

dipping into molten lead or other metals the tube should be bored

from the solid. The great advantage of an iron or steel sheath is its

mechanical strength, which protects the couple from damage in case of

rough usage.

2. _Nichrom._—Certain alloys of nickel and chromium, and especially

that known as Nichrom II, may be kept at 1100° C. without oxidising

to any appreciable extent; and hence sheaths of this material may be

used up to the temperature named. In addition to being more durable

than iron, nichrom possesses the same advantages of strength and good

conductivity; on the other hand, it is more costly.

3. _Molybdenum._—This metal, which possesses a melting point of

about 2500° C., may be dipped in molten brass, bronze, copper, etc.,

without being attacked, and has been used to form the tip of a

protecting-tube designed to measure the temperature of molten alloys. A

junction covered only by a thin tube of molybdenum quickly attains the

temperature of its surroundings.

4. _Graphite and Graphite Compositions._—Carbon has the highest

melting point of all known substances, and in the form of artificial or

Acheson graphite may be easily machined to any desired shape. Graphite

sheaths are sometimes used for immersion in molten metals, but at 1000°

C. and higher Acheson graphite oxidises easily and becomes friable. It

is a good conductor of heat, but is easily broken. Compositions of

natural graphite and refractory earths, such as Morgan’s “Salamander,”

are inferior to pure graphite in conductivity, but are stronger and not

readily oxidised, and may be used to form sheaths for temperatures up

to 1400° C. or possibly higher, when penetration of furnace gases to

the junction is not of moment.

5. _Porcelain._—This material, in its best forms, may be used up to

1400° C., but must be efficiently glazed to prevent the ingress of

furnace gases to the junction. It is easily broken by a blow, and when

circumstances permit should be protected by an iron covering-sheath.

The variety known as “Marquardt” has been found very satisfactory for

high-reading thermal couples. Porcelain is not a good conductor of

heat, and a junction encased in it does not respond quickly to external

changes in temperature.

6. _Vitrified Silica._—This substance, which may be worked in the

oxy-hydrogen blowpipe, is largely used as a protecting-tube. It is not

advisable, however, to use it for continuous work above 1100° C., as

beyond this temperature devitrification occurs, and the tube becomes

porous. It is a fairly good conductor of heat, and withstands rapid

changes in temperature without cracking. It is very brittle, and for

this reason is generally encased in iron.

7. _Alundum._—This material is made from fused bauxite, and

has a melting point of 2050° C. A special form of alundum, used

for protecting-tubes, is non-porous up to 1300° C., and forms a

satisfactory covering. Alundum is a moderately good conductor of heat,

but is easily broken.

8. _Carborundum._—This is an electric furnace product, which may

be heated above 2000° C. without damage. For making into pyrometer

tubes, it is bonded with a suitable material, and baked after shaping.

Carborundum, and the amorphous variety known as “silfrax,” have proved

useful for protecting junctions at temperatures as high as 1600° C.

The thermal conductivity is relatively good, but the tubes are easily

broken.

9. _Magnesia._—Tubes of this material, which melts at a temperature

considerably above 2000° C., have been used for special work. Magnesia

is a poor conductor of heat, and has little mechanical strength.

10. _Zirconia._—This is a very refractory material, its melting point

exceeding 2500° C. It may be made into a vitreous variety, which is

non-porous and proof against sudden temperature changes. At present,

only a moulded form of pyrometer tube, made from zirconia powder, is

available, the material worked in this manner being termed “zirkite.”

Although zirconia is a bad conductor of heat, its other qualities

are such that it forms an excellent material for work at the highest

temperatures possible for thermal junctions; and when the vitreous

variety is available, may come into extended use.

[Illustration: FIG. 6. PYROMETER WITH SPECIAL COLD JUNCTION IN HEAD.]

It will be seen from the foregoing that the ideal protecting-tube has

yet to be found, and the user must choose the one which comes nearest

to his requirements. Special consideration must be given in cases when

chemical fumes are present, and a sheath selected which is not attacked

or penetrated by them.

Returning to the junction, it is advisable always to weld the wires,

and not to rely upon the contact resulting from twisting them together.

Platinum and the platinum alloys may be welded readily by placing the

junction in a coal-gas blowpipe fed with oxygen instead of air. For

work at lower temperatures the platinum metals may be soldered by means

of a small quantity of gold, in the flame of a Bunsen burner.

When cheap metals are used for the junction the construction may

be considerably modified, and often with advantage. In fig. 6, for

example, which represents a thermocouple made by A. Gallenkamp & Co.,

the metals used are copper and constantan, and the hot junction,

fastened by silver solder, is supplemented by a cold junction of the

same metals located in the head. The copper wire from the hot junction

passes directly to a copper terminal, from whence a copper wire lead

is carried to the galvanometer; and the same procedure is carried out

with the copper wire from the cold junction, thus realising the circuit

shown in fig. 2. The cold junction is kept in oil, the temperature

of which is registered by a short thermometer, thus enabling (as

will be explained later) the correct temperature of the hot junction

to be deduced under any circumstances. In this instrument twin-bore

fireclay is used to insulate the wires, and the protecting-tube is

of iron—which suffices for the upper limit (800° C.) to which the

junction may be used. Iron and constantan could be used in this manner

by employing iron leads to the galvanometer.

Another type of instrument, rendered practicable by the use of cheap

metals, and which may be termed the “heavy type,” is constructed of

thick pieces of the metals welded together instead of wires, thus

ensuring greater strength and longer life. Messrs Crompton & Co. were

the first to introduce thermocouples of this type, consisting of a

heavy steel tube, to one end of which a nickel rod is welded, the other

end being free, and the length of the rod suitably insulated from the

steel tube; leads for the rod and tube being taken to the galvanometer.

Fig. 7 shows a couple of this kind, made by Paul, consisting of an iron

tube down the middle of which a constantan rod is passed, insulated

from the tube by magnesia. At the tapered end the two metals are welded

together, and at the free end a special cap, fitted over the tube and

rod, the contact parts being insulated from one another, serves to

enable leads to be taken to the galvanometer. Similar thermocouples are

made by the Foster Instrument Company (fig. 8), and are simple, cheap,

and reliable up to 900° C. with an iron-constantan couple, and to 1100°

C. with nichrom couples. When worn out they may be replaced, at a

trifling cost, by others made from the same batch of metal.

[Illustration: FIG. 7.—HEAVY TYPE, CHEAP-METAL PYROMETER.]

The drawback to the use of carbon as one of the materials for a

junction is the difficulty experienced in securing a good contact with

the metal with which it is coupled. In nickel-carbon junctions the

contact is sometimes ensured by the aid of a spring, which presses

the two substances together. Such an arrangement is evidently not so

reliable as one in which the materials are welded, and a defective

contact, arising from any cause, would lead to serious error. A

preferable plan is to screw both the nickel and carbon rods into a

cross-piece of either element.

[Illustration: FIG. 8.—FOSTER’S CHEAP-METAL PYROMETER.]

When applying a thermal junction to the measurement of surface

temperatures, such as steam-pipes or the exterior of furnaces, the

wires may be passed through a thin disc of metal, about ¼ in. in

diameter, and soldered at the back. Suitable materials are copper and

constantan, soldered to a thin copper disc with silver solder, and

brought to a cold junction in the head of the instrument as shown in

fig. 6. The terminal piece of the insulation may be made of hard wood,

with the holes countersunk so as to cover the solder and enable the

wood to touch the disc, which, when pressed on the hot surface, will

then rapidly acquire the temperature. The author has found, by trials

under varying circumstances, that this method of measuring surface

temperatures gives reliable and concordant results. For very high

surface temperatures a platinum disc, with one of the usual platinum

metal couples soldered to the disc with pure silver, and a piece of

twin-bore fireclay brought to the back of the disc, will be found to

suffice for most cases arising in practice. A small blowpipe flame is

best for soldering the wires to the disc, borax being used as flux in

the first case; but no flux is necessary in soldering the platinum

metals with pure silver.

In deciding upon the length of a thermocouple it must be remembered

that the temperature recorded is that prevailing in the region of

the hot junction. When the temperature of a furnace is uniform it is

sufficient to allow the end of the thermocouple to protrude about 12

inches into the interior, but when following the change of temperature

undergone by objects in a furnace the end must be located near the

objects. If the distance from the exterior of the furnace to the

objects exceed 2 feet, the thermocouple should be inserted through the

roof so as to hang vertically, as if placed through the side it would

droop by its own weight at high temperatures. The distance between

the exterior of the furnace and the cold junctions should be at least

15 inches in all cases in which the heating of the cold junction

is not automatically compensated. After inserting the couple the

opening through the furnace wall should be closed by means of suitable

luting-clay.

In certain instances, such as flues, it is necessary to use a long

instrument in a horizontal position. A rail may then be placed across

the flue, at a suitable place, to serve as a support and so to prevent

drooping.

=Liquid Element Thermocouples.=—An investigation by the author

and A. W. Grace has shown that the continuity of the E.M.F. produced

by a rising temperature is not interrupted by fusion, except in the

cases of bismuth and antimony, which both show an abrupt change in

thermo-electric properties at the melting point. It would therefore

appear feasible to measure temperatures by constructing a thermocouple

so as to retain the circuit after fusion, the advantage gained being

that the range is restricted by the boiling point of the metals instead

of the melting point and higher readings are rendered possible. The

boiling points of some of the common metals are appended:—

───────────┬───────────────────

Metal. │ Boiling Point.

───────────┼─────────┬─────────

│ Deg. C. │ Deg. F.

├─────────┼─────────

Aluminium │ 1800 │ 3270

Silver │ 1955 │ 3550

Tin │ 2270 │ 4120

Copper │ 2310 │ 4190

Nickel │ 2330 │ 4225

Iron │ 2450 │ 4440

───────────┴─────────┴─────────

From inspection of these figures, it will be seen that if a suitable

couple could be obtained, common metals might be used to measure

temperatures equalling or even exceeding the limit of the range covered

by wire junctions of metals of the platinum series. Instead of using

two metals, graphite might form one member of the couple, provided that

no objection to its use existed on other grounds.

[Illustration: FIG. 9.—LIQUID-ELEMENT THERMOCOUPLE.]

The form of thermocouple designed by the author to permit of the use of

molten elements is shown in fig. 9. A rod of refractory material, R, is

perforated longitudinally by two holes, down which are passed rods of

the thermo-elements, A and B. The lower ends of A and B are inserted in

a graphite block G, which is jointed on its upper face to R; the whole

being surrounded by the refractory cover C. On either or both of the

elements melting, the circuit is maintained through G, which serves

also to prevent the mixing of A and B when molten, whilst not affecting

the E.M.F. developed. In order to allow for the expansion of the metals

on melting, A and B are made to fit loosely in R. When inserted in a

furnace to a depth represented by EF, only the portion of the metals

adjacent to the closed end will melt, the outer parts remaining solid.

At present it has not been found possible to procure the refractory

parts in a form suited to commercial use, but when this obstacle

is overcome this type of thermocouple should prove of service for

measuring temperatures beyond the scope of ordinary base-metal

junctions.

=Indicators for Thermo-electric Pyrometers.=—As the electromotive

force developed by a single junction when heated is small, a sensitive

galvanometer is required to indicate the minute current flowing

through the circuit. Delicate millivoltmeters, of the moving-coil

type, are universally employed, as they possess the advantage of

an evenly-divided scale combined with the requisite degree of

sensitiveness. The original d’Arsonval galvanometer, consisting of a

coil suspended by a metallic strip between the poles of a horse-shoe

magnet, was used by Le Chatelier, who, by its aid, was enabled to

lay the foundations of this branch of pyrometry. Three forms of this

instrument are now in use, viz. (_a_) the suspended coil “mirror” type;

(_b_) the suspended coil “pointer” type; and (_c_) the pivoted type.

Examples of each will now be described.

[Illustration: FIG. 10.—HOLDEN-D’ARSONVAL MIRROR GALVANOMETER.]

Fig. 10 represents a mirror galvanometer working on the d’Arsonval

principle, designed by Gen. Holden, F.R.S. The horse-shoe magnet

is laminated, and an iron core, supported by a pillar, is placed

between the poles. The coil, which moves in the space between the

core and the poles of the magnet, is suspended by a thin, flat strip

of phosphor-bronze, which carries a small circular mirror. A similar

phosphor-bronze strip is fastened to the lower part of the coil,

and is continued to an adjusting-screw in the base. The ends of the

suspension strips communicate with the terminals of the galvanometer,

and a current entering at one terminal passes through the metallic

suspensions and the coil to the other. The effect of passing a current

through the coil, which is located in a powerful magnetic field is to

produce an axial movement tending to twist the suspension strips, which

movement is greatly magnified by a spot of light reflected from the

mirror on to a distant scale. When the current ceases, the untwisting

of the strip restores the coil to its former position. Galvanometers of

this type are remarkably “dead-beat” in action, that is, the movement

and restoration of the coil are accomplished without vibration. A

semi-transparent scale, placed at 1 metre distance, and 50 centimetres

long, is suitable for use with this galvanometer. When used in

workshops, it is necessary to protect a mirror galvanometer from the

vibrations produced by machinery, which would cause the spot of light

to become unsteady. The best method of effecting this is shown in fig.

11, which represents the mode of suspension devised by W. J. Lambert

for use in the Royal Gun Factory, Woolwich Arsenal. The usual supports

of the galvanometer are abolished, and the instrument suspended from

the ring of a brass tripod, so as to keep three springs partly in

compression. When suspended in this manner, a mirror galvanometer

is quite suited to commercial use; in the quiet of the laboratory

the ordinary supports may be employed. The advantage gained by using

the mirror type is that a much longer scale is possible than with

instruments furnished with a pointer, and hence greater accuracy in

determining temperature readings may be secured.

[Illustration: FIG. 11.—LAMBERT’S ANTI-VIBRATION STAND FOR

GALVANOMETERS.]

[Illustration: FIG. 12.—SIEMENS’ THERMO-ELECTRIC INDICATOR.]

In suspended coil instruments furnished with a pointer, the

construction differs only in detail from the foregoing. In place of

the mirror, a light pointer is attached to the suspension so as to

rest on the coil and a scale is furnished over which the pointer

moves. Fig. 12 is an example of this type, made by Messrs Siemens, the

suspension being contained in the tube which rises from the body of the

instrument. The maximum length of scale moved over by the extremity

of the pointer is about 6 inches, as a longer and therefore heavier

pointer would reduce the sensitiveness below the point requisite for

thermo-electric work.

In the double-pivoted type, the suspension is eliminated, and pivots

are fastened to each end of the moving coil which rest in bearings.

The turning of the coil is made to compress a hair spring, made of

phosphor-bronze; and when the current ceases the unwinding of this

spring restores the coil to its former position. The coil carries

a pointer which moves over a scale. These instruments are not so

sensitive as those in which the coil is suspended, but can be made

sufficiently sensitive to work with any kind of junction in practical

use. The pivoted form is cheaper and stronger than the suspended type,

and is used whenever sufficiently sensitive.

The “Uni-pivot” galvanometer, made by R. W. Paul, is shown in figs. 13

and 17. The coil, which carries the pointer, is circular, and moves

round a spherical core of iron placed between the poles of the magnet.

A hole is drilled in the iron core, and the coil rests on a single

bearing at the bottom of this hole. A phosphor-bronze control-spring

serves to restore the coil to the zero position. The lessened friction

due to the use of a single pivot enables this instrument to be made

very sensitive when needed, so that a relatively small rise in the

temperature of a junction may cause the pointer to traverse the whole

length of the scale.

[Illustration: FIG. 13.—PRINCIPLE OF UNI-PIVOT GALVANOMETER.]

=Special Features of Indicators.=—All moving-coil instruments,

whether suspended or pivoted, are liable to alteration of the zero

point owing to what is termed “creep.” The suspension strip, when first

fixed in position, generally possesses a certain amount of initial

torsion, which comes into operation gradually and causes a slight

movement of the coil. Similarly, in a pivoted instrument, the strength

or shape of the control-spring undergoes a gradual alteration at first,

causing the pointer to move away from the zero position. For this

reason adjusting arrangements are fitted by means of which the spot of

light or pointer may be brought back to the zero. This creeping ceases

after a time—often requiring twelve months—and if not subjected to

any strain, error from this cause does not recur to any notable extent.

With a mirror galvanometer it is better to move the scale, or turn the

galvanometer round on its axis to restore the correct zero, rather than

to twist the coil back; but with a fixed scale and pointer the only

remedy is to turn the coil bodily round. In a single-pivot indicator

constantly used in the author’s laboratory, the creep amounted to a

movement of the end of the pointer through an angle of 2 degrees in

the first few months, since when, after the lapse of several years, no

further alteration has occurred. It is advisable to test the zero point

of an indicator from time to time by breaking the circuit, and if an

error be discovered the pointer should be re-set, or an allowance made

in taking a reading.

The resistance of an indicator should be so high that the readings

should not be perceptibly altered by any fluctuations in the resistance

of the circuit which may arise in practice. If leads of considerable

length were used to connect the pyrometer with the indicator, and were

subject to fairly large alterations of temperature, the consequent

changes in the resistance of such leads would be noticeable on a

low-resistance indicator; and similarly, if a pyrometer were inserted

at different depths in a furnace at separate times, thus heating up

varying lengths of the junction wires, a discrepancy would arise for

the same reason. The resistance of an indicator, however, cannot be

raised beyond a certain point without reducing the sensitiveness below

the required limit. A mirror galvanometer of the type described may

have a resistance—partly in the coil and partly in an added series

resistance—of 1000 ohms or more, and still be sufficiently sensitive;

and in the latest types of instruments provided with pointers the

resistance may be made as high as 1000 ohms, although it is more

usually 400 to 500 ohms. Many indicators are in use, however, in which

the resistance is 100 ohms or less. As, from Ohm’s law, the current

varies inversely as the total resistance in the circuit, any alteration

in resistance should be small relatively to the total to render the

error negligible. This point is made clear in the following example:—

_Example._—A thermocouple and leads have a resistance of 5

ohms and are subject to alterations amounting to 1 ohm. To

find the errors resulting when indicators of resistances

800, 400, and 50 ohms respectively are used.

From Ohm’s law, C = E/R, the variation in C, with E

constant, will be 1 in 805, 1 in 405, and 1 in 55

respectively. As the indications are proportional to the

current, the alterations caused will be approximately

⅛ per cent., ¼ per cent., and 2 per cent. The first two

may be ignored; the last may be quite serious and lead to

the failure of an operation.

It will be seen from the foregoing that low-resistance indicators

should only be used for fixed thermocouples and short leads not subject

to temperature changes, or, in other words, in a circuit of fixed

resistance.

The resistance of an indicator, when unknown, may be found by the

following method, suggested by the author:—A resistance box is joined

at one end to one terminal of the indicator. To the other terminal a

fairly stout iron wire, 18 inches long, is connected, and a similar

length of constantan wire is coupled to the other end of the resistance

box. The free ends of the wires are twisted into a junction which is

dipped into boiling water. The deflection obtained with no resistance

in the box (D_{1}) is noted, and resistances (R) are then unplugged

until the deflection (D_{2}) is approximately one-half of D_{1}. The

resistance (G) of the indicator, ignoring that of the wires, is then

given by the formula

D_{2}R

G = ───────────────

(D_{1} - D_{2})

as may readily be proved from Ohm’s law, E being constant. This method

is extremely simple and reasonably accurate.

Reliable indicators are now procurable from many instrument-makers at

a comparatively small cost, progress in this direction having been

most marked in recent years, particularly in the case of pivoted

instruments. The most convenient form for workshop use is made with

an edgewise scale (fig. 14) and may be placed in a suitable position

fixed to a bracket. The flat-scale pattern is preferable for use on a

laboratory table, or for a portable pyrometer. The sector pattern is

also good for workshop use, the dial being visible from a distance.

[Illustration: FIG. 14.—INDICATOR WITH EDGEWISE SCALE.]

=Standardizing of Indicators to read Temperatures directly.=—The

temperature scale of an indicator, for use with a given thermal couple,

is always marked by the maker in the case of instruments furnished

with a pointer, and, generally speaking, is correct within reasonable

limits. It is customary and necessary to send with the instrument a

statement of the cold-junction temperature for which the markings

are correct; say 20° C. or 60° F. The user should then endeavour to

maintain the cold junction at this specified temperature when taking

a reading, or otherwise a considerable error may be introduced. It is

highly desirable, however, that the user should be able to perform the

standardizing himself, if only for checking purposes; and when using a

mirror galvanometer as indicator it is necessary to standardize on the

spot at which the instrument is fixed. Ability to prepare a temperature

scale is further useful, inasmuch as any good millivoltmeter, of range

0 to 20 millivolts, may be used for thermo-electric work of all kinds,

and may be calibrated for different junctions, a suitable series

resistance being added to enable E.M.F.’s higher than 20 millivolts to

be measured. Such an instrument may thus be made extremely useful, both

in the workshop and laboratory.

Standardization may be effected either by subjecting the hot

junction to several known temperatures, and noticing the deflections

corresponding thereto; or by measuring the electromotive force

developed by the junction, and calculating the corresponding

temperature from a formula which is known to hold for the range

comprehended by the instrument. The former method is simpler; and if

carefully conducted is quite accurate. The latter method possesses the

advantage that readings in millivolts may be translated directly into

temperatures when the constants of a given thermal couple are known.

It is now usual to mark indicators with a double scale, one reading

millivolts and the other temperatures.

=Standardization by Fixed Points.=—Taking any millivoltmeter

which, with a maximum of 20 millivolts at the terminals, will give a

full scale deflection, the first step is to arrange that the pointer

(or spot of light) shall just remain on the scale at the highest

temperature to be attained by the junction. This may be done by placing

the hot junction in boiling water and noting the deflection obtained,

either in millivolts or equal arbitrary divisions, and also the

temperature of the cold junction. The deflection observed is due to

a difference of temperature (100-_t_) deg. C, where _t_ is the

temperature of the cold junction. If the highest temperature to be

measured is 10 times (100-_t_), the deflection should be rather

less than 1/10 of the scale, and similarly for any other required

temperature limit. If the observed deflection exceed this proportion, a

series resistance should be added until the correct value is obtained.

This resistance is then permanently installed in the circuit for use

with the junction under trial.

Before proceeding further it is necessary to consider whether the

pyrometer is to possess a single cold junction of ascertainable

temperature (as in fig. 6), or whether it will be arranged with two

cold junctions in the head, as in fig. 4. In the former case it is

simpler to prepare a “difference” scale; that is, one which reads

differences of temperature between the hot and cold junctions, from

which the temperature of the hot end may be obtained by adding to the

difference that of the cold junction. In the latter case the cold end

should be kept by artificial means at the temperature likely to be

attained in practice—say 25° C.—a water-bath being suitable for this

purpose. It is advisable to remove the shield of the pyrometer when

standardizing, so as to expose the hot junction, as closer readings can

then be taken.

A number of materials—preferably cheap—of known boiling points or

melting points are then selected from a table of fixed points (page

16) so as to give about six points, distributed fairly evenly over the

scale. As an example, if it were desired to prepare a temperature scale

from 0° to 1000° C., the following might be chosen:—

════════════════════════════════════════════════╤═══════════════════

Substance and Condition. │ Temperature.

────────────────────────────────────────────────┼─────────┬─────────

Water at boiling point │ 100° C. │ 212° F.

Tin at melting point │ 232 │ 449

Zinc at melting point │ 419 │ 786

Antimony at melting point │ 631 │ 1167

Common salt at melting point │ 800 │ 1472

Copper at melting point (covered with graphite) │1084 │ 1983

════════════════════════════════════════════════╧═════════╧═════════

The hot junction is allowed to attain these temperatures successively,

and the corresponding deflection in each case is noted. It is then

possible to divide up the whole of the scale to read temperatures

directly.

The first reading is taken by placing the junction in a vessel of

boiling water, and for a locality near sea level it is not necessary

in ordinary work to take account of fluctuations in the boiling point

due to alterations of atmospheric pressure. To ensure that the other

readings are taken when the substances are exactly at the melting

point, the procedure is as follows: about 2-3 lb. of the substance are

melted in a salamander crucible, and a small fireclay tube, closed at

one end, is inserted in the molten mass. The hot junction is placed

in the fireclay tube, and the intervening space filled with asbestos

fibre. Great care must be taken not to let the junction touch the fused

substance. The crucible is now allowed to cool, and a reading of the

deflection taken every half-minute. When the substance is exactly at

its solidifying point—identical in general with the melting point—the

deflection remains stationary for several consecutive readings, owing

to the liberation of latent heat of fusion in sufficient quantity

to balance the loss by radiation. This stationary reading is noted

for each substance, and represents the deflection given when the hot

junction is at the temperature corresponding to the melting point, and

the cold junction or junctions at the temperature existing when the

observation is made. For melting the materials, a Davies furnace with

a large Teclu or Meker burner is convenient up to 850° C.; but to melt

the copper a blast lamp is requisite. The molten mass may be allowed to

cool in the furnace.

From these observations a calibration curve may be drawn either for

differences between hot and cold junctions, or for a steady temperature

of the cold junctions. Two sets of data are appended to illustrate the

procedure.

════════════╤═════════════════════════════════╤═════════════════════

Temperature │ Pyrometer 1. Iron-constantan. │ Pyrometer 2.

of Hot │ (Series resistance in │ Platinum-

Junction. │ galvanometer circuit.) │ iridioplatinum.

├───────────┬─────────┬───────────┼───────────┬─────────

│Deflection.│ Cold │Difference.│Deflection.│ Cold

│ │Junction.│ │ │Junction.

────────────┼───────────┼─────────┼───────────┼───────────┼─────────

100° C. │ 8·9 │ 15° C.│ 85° C. │ 5·5 │}

232 │ 21·8 │ 17 │ 215 │ 15·6 │}Constant

419 │ 40·6 │ 19 │ 400 │ 29·4 │} at

631 │ 63·8 │ 19 │ 612 │ 45·5 │} 25° C.

800 │ 83·0 │ 20 │ 780 │ 59·0 │}

1084 │ ... │ ... │ ... │ 82·0 │}

════════════╧═══════════╧═════════╧═══════════╧═══════════╧═════════

[Illustration: FIG. 15.—CALIBRATION CURVES FOR TWO THERMO-ELECTRIC

PYROMETERS.]

Fig. 15, A, is a calibration curve for thermocouple 1, connecting

deflections with corresponding differences between the temperatures

of the hot and cold junctions. In order to read from this curve the

temperature of the hot end, the reading corresponding to the observed

deflection is added to the existing temperature of the cold junction.

Thus if a deflection of 56 divisions were obtained with the cold

junction at 25°, the temperature of the hot junction would be (540 +

25) = 565° C. The advantage of this method of calibration is that it is

unnecessary to take precautions to keep the cold junction at a steady

temperature; and when a single cold junction is used, as in fig. 6,

this plan should always be followed. It will be noted that this curve

passes through zero, as no deflection represents no difference of

temperature.

Fig. 15, B, represents the calibration curve for pyrometer 2, and

is such that direct readings may be obtained corresponding to any

given deflection, for a cold junction temperature of 25°. This curve,

therefore, cuts the axis of zero deflection at 25°, as no deflection

corresponds to the condition when both hot and cold junctions are at

25°. This method of calibration may be used with advantage for couples

of the type shown in fig. 4, where two cold junctions exist in the

head, and the simple rule of adding the cold junction temperature

does not apply. Many suggestions have been made for correcting for

alterations in the temperature of the cold end of such a couple,

but none are accurate, and it is necessary to keep this part at the

temperature of standardization to secure correct readings. In both of

the above calibrations the galvanometer used possessed a scale divided

into 100 equal arbitrary divisions.

In making permanent temperature scales from these curves to attach

to the existing galvanometer scale, intervals of 100° may be taken

and marked opposite to the corresponding divisions on the existing

scale. Each 100° may then be equally subdivided into as many parts as

the length of scale permits, and numbered at suitable intervals. If

the junction used yield a calibration curve departing greatly from a

straight line, every 50° interval should be taken, or, if necessary,

every 25°. In the examples given both curves are nearly straight lines

in the working region, viz. 400° to 800° for the iron-constantan

junction, and 500° to 1100° for the platinum-iridioplatinum.

One precaution necessary in standardizing an indicator by this method

is to ensure that the metals used are pure, as impurities lower the

melting points. If ordered as “pure” from any dealer of repute, the

metals will generally be found satisfactory. The common salt used

should be the ordinary salt sold in blocks, and not a prepared table

salt. A second precaution, when observing melting points, is to guard

against a possible error due to the substance becoming “surfused” or

“overcooled”; in which case the temperature falls below the ordinary

freezing point before solidification commences. When freezing occurs,

however, the temperature rises to and remains at the true melting

point, and an increase of deflection following a gradual fall always

indicates overcooling. The higher deflection then attained is the true

freezing point. Antimony frequently overcools to 600° before freezing,

but on setting rises to the correct figure—631°. All metals and salts

are liable to overcooling occasionally.

=Standardization by Measurement of E.M.F.=—It has been found,

as the result of experiments, that the relation between the E.M.F.

developed by a junction and its temperature—under constant conditions

of the cold junction—may be expressed approximately by a formula as

under:—

log E = A log _t_ + B (Holman’s formula),

where E = electromotive force in microvolts, _t_ = temperature in

Centigrade degrees, and A and B are constants depending upon the

junction. With certain junctions this formula may be applied over

the working part of the scale with an error not exceeding 2° C., but

with others the discrepancy is greater. In order to determine the

constants A and B, it is necessary to measure the E.M.F. at two known

temperatures, which should be chosen as far apart as possible in the

working region. When these constants are known, a measurement of E

enables the temperature _t_ to be found by calculation.

_Example._—Le Chatelier found that a junction at the

temperature of melted aluminium (657° C.) gave 6200

microvolts; at the melting point of copper in air (1062° C.)

the figure was 10580. Applying in the above formula

log 6200 = A log 657 + B

and log 10580 = A log 1062 + B,

the value of A is 1·2196 and of B 0·302, as may be found

by taking logarithms and solving for A and B.

The values of the constants A and B vary for different junctions, and

also for different melts of what are reputed to be the same materials.

When once determined for a quantity of homogeneous wires, to which

the formula applies with sufficient accuracy, it is evident that an

indicator with a millivolt scale may be made to read temperatures

directly without any necessity for further experiment, although it is

always advisable to take one check reading at a fixed point in the

working range.

[Illustration: FIG. 16.—POTENTIOMETER METHOD OF MEASURING E.M.F.]

In order to determine the E.M.F. of a junction at different

temperatures, the potentiometer method is used, in which the E.M.F.

of the test-couple is balanced against the known E.M.F. furnished by

a constant cell. The circuit is shown in fig. 16, in which B is an

accumulator which sends a current through the resistances R_{1}, R_{2},

and the calibrated wire DE. The cold ends of the couple are attached

at P so as to be in opposition to B, and in this branch of the circuit

are included a sensitive galvanometer G and a portion of the wire DE. A

standard cadmium cell, S, is connected between R_{1} and R_{2} at one

end, and may be put in circuit with the galvanometer through the switch

A. In commencing, S is connected to the galvanometer and R_{1} adjusted

until no deflection is obtained on G. The switch A is now moved over

to the circuit of the couple, and the terminal F moved along the wire

until zero deflection is again obtained. The E.M.F. of the couple is

determined from the relation

E of junction Resistance of DF

────────────────── = ───────────────── .

E of standard cell R_{2}

By exposing the hot end of the junction to successive standard

temperatures, and maintaining the cold ends at a known constant

temperature, the necessary data for inclusion in a formula may be

obtained.

In fixing a permanent temperature scale, calculated from the formula,

to a millivoltmeter, it must be remembered that the values given by

the experiment are absolute, and independent of the resistance of the

circuit composed of the thermo-element and galvanometer. On the other

hand, a millivoltmeter is marked to read difference of potential at

its terminals; and if in series with a junction and leads of notable

resistance, its indications will not be the E.M.F. of the junction. An

example will make this point clear.

_Example._—A millivoltmeter has a resistance of 100

ohms, and is marked to read P.D. at its terminals. A

thermocouple and leads connected to the millivoltmeter

have a resistance of 5 ohms. To find the relation between

the true E.M.F. of the junction and the readings of the

indicator.

If E = the E.M.F. developed by the junction, and V, the

reading of the millivoltmeter, = P.D. at its terminals,

then the current in the circuit = E/105 = V/100; and V =

(100/105)E. That is, the readings are lower by 5 per cent.

than the true E.M.F. of the junction. In the same way a

low resistance voltmeter, if applied to a cell of high

resistance, shows a lower reading than the E.M.F. of the

cell.

This example indicates how a table connecting true E.M.F.’s with

reading in millivolts may be calculated when the resistances concerned

are known. It is presumed, in preparing a scale in this manner, that

the resistance of the couple will not be subject to such alterations as

to affect the reading.

The advantages of this method of calibration are manifest when a number

of junctions are being made from a given batch of wires, as it is only

necessary to divide the scale of the indicator so as to represent

millivolts—a simple operation—and then to attach a temperature scale.

This procedure is much more expeditious than standardizing each

indicator at several fixed points when a number are concerned, but for

a single junction the fixed point method is easier. The potentiometer

method of measuring E.M.F. may also be used to determine temperatures

in place of an indicator, and is of great service in cases where very

accurate readings are specially required, being far more delicate in

detecting small differences of temperature than an indicator. Special

potentiometers for thermo-electric work are made by the Cambridge

and Paul Instrument Company, Siemens, and others, and are useful in

conducting accurate research, but are too elaborate for workshop or

ordinary laboratory practice.

=Cold Junction Compensators.=—The necessity for paying attention

to the cold junction has led to various attempts to compensate

automatically for changes of temperature at this part of the pyrometer.

A thermometer located near the cold junction, as in fig. 6, is all that

is needed to correct a two-junction circuit; but when a three-junction

circuit is used a correct reading is not secured by adding the excess

temperature of the thermometer over the calibration temperature to

the reading on the indicator. In Bristol’s arrangement a mercury

thermometer, with a large bulb and wide stem, is stationed at the cold

junction, and participates in any temperature change. In the stem is

placed a loop of thin platinum wire, which forms part of the pyrometer

circuit. When the mercury is heated it expands up the stem and

short-circuits a portion of the loop, thereby diminishing the

resistance of the pyrometer circuit, and tending to increase the

deflection on the indicator. Simultaneously the cold junction will

be heated, tending to diminish the current, and so to cause a less

deflection. By adjustment these two tendencies may be counterbalanced,

so that the reading is unaffected, but such adjustment will only

apply to a given E.M.F., and therefore to one temperature of the hot

junction. Hence this method fails in general application.

Peake’s compensated leads are intended to remedy cold-junction errors

by transferring this junction, in effect, to the galvanometer. They

are used for pyrometers in which the platinum metals are employed, and

consist of wires of two different alloys of copper and nickel, which

connect the cold end to the indicator. These alloys are such that

the electromotive forces set up at the junctions in the head—Pt and

Cu-Ni 1, and Pt-Ir with Cu-Ni 2—are equal and opposite at all working

temperatures, and hence changes at the cold junctions do not affect the

reading. At the indicator, however, temperature changes would cause an

alteration in deflection; but as the indicator is generally placed well

away from the furnace, and is not liable to notable heating or cooling,

the possible errors are greatly reduced by the use of these leads. They

are obviously of no value for use with base-metal pyrometers, as the

wires used in such may be prolonged to the indicator, with an identical

result.

[Illustration: FIG. 17.—DARLING’S COMPENSATOR, FITTED TO GALVANOMETER.]

An automatic compensator for use with base-metal pyrometers has been

devised by the author, and is illustrated in figs. 17 and 18. A spiral

made of a compound strip of two metals is attached to the needle of the

indicator, and coils or uncoils when cooled or heated, thereby moving

the pointer over the scale. The length of the spiral is such that an

alteration of a given number of degrees in its temperature moves the

pointer by the same number of degrees on the scale—or, in other words,

the temperature scale of the pyrometer is identical with that of the

spiral. The metals forming the junction are continued, in the form of

wires, to the interior of the galvanometer, where a cold junction is

formed, which will always possess the same temperature as the spiral.

The scale is constructed to represent differences of temperature between

the hot and cold junctions, and before coupling up the pyrometer the

pointer indicates the temperature of the spiral; that is, of the cold

junction. On connecting the thermocouple the pointer is moved by the

coil of the indicator through an amount represented by the difference

in temperature between the two junctions, and therefore finally

indicates the temperature of the hot junction.

[Illustration: FIG. 18.—INDICATOR FITTED WITH DARLING’S COMPENSATOR.]

_Example._—If the cold junction were at 20°, the pointer,

before connecting the couple, would indicate 20° on the

scale. If the hot junction were 580° hotter than the cold,

then on completing the circuit the pointer would move 580

additional degrees along the scale, so that the figure

indicated would be (20 + 580) = 600°, the temperature of

the hot junction. If now the indicator were heated by 10°,

the spiral would tend to augment the deflection by 10°,

but simultaneously the deflection due to the junctions

would fall off by 10°, and the reading would still be 600°.

This method of compensation renders the readings independent of the

cold junction, and, in addition to its use for high temperatures,

enables ordinary and low temperatures to be read simply and correctly,

as will be shown later. The spiral is located in the tower rising from

the top of the indicator in fig. 18.

In Paul’s method of compensation the thermocouple and indicator

are placed across a Wheatstone bridge, two arms of which contain

resistances of copper, whilst the resistances in the other two arms are

of manganin. Any change in temperature at the cold junction is shared

by these four resistances, and, whilst affecting the resistance of the

copper parts, no change is caused in the manganin parts, as this alloy

has a negligible temperature coefficient. If, therefore, the bridge

were initially balanced at 20° C., and the temperature rose to 30°,

the increased resistance of the copper would destroy the balance, and

permit of a small current passing through the indicator. A fall to

10°, by diminishing the resistance of the copper, would cause an equal

current to pass through the indicator in the opposite direction. The

amount of this current is arranged so as to add the rise in temperature

of the cold junction to the reading of the indicator in the one case,

and to subtract the fall in the other, thus retaining true readings for

the cold-junction temperature at which the couple was standardized.

=Constant Temperature Cold Junctions.=—If the cold junction

can be kept at a steady temperature, compensators are unnecessary,

but no good practical means of achieving this end has yet been

devised. Water-cooled heads have already been referred to; but in many

situations the connecting-pipes entailed would be objectionable, and

hence this arrangement is not greatly used. An alternative method,

suggested by Prof. A. Zeleny, is to bury the cold junction in the

ground. Recent experiments, conducted at Cambridge by R. S. Whipple,

showed that a junction buried 10 feet deep did not vary in temperature

by more than 2° C. over a period of three years. This has led to the

adoption of buried junctions in special cases; but it is probable that

much greater variations would be experienced in the ground beneath

large furnaces, in which case the advantages of this procedure would

be lost. A common workshop method is to locate the cold junction in a

thermos flask filled with oil, when a temperature constant to 2° C. may

be secured, although the changes in the temperature of the surrounding

atmosphere may be as great as 150 C. For special work, ice may be

used in the thermos flask, thus securing absolute constancy; but this

procedure is not feasible in ordinary works practice.

=Special-Range Indicators.=—When the working range of a pyrometer

is from 600° C. upwards, it is evident that the part of the scale

occupied by the first 600° is useless, and that it would be an

advantage if the whole scale could be utilised for the special working

range, so as to secure more exact readings. This may be accomplished

by a “set-up” against the movement of the pointer caused by the

thermocouple, so as to prevent any motion over the scale until an

assigned temperature is reached. For example, a junction developing

12 millivolts at 1000° C. may be coupled to an indicator in which the

full-scale deflection of the pointer is produced by 6 millivolts. If an

E.M.F. of 6 millivolts be opposed to the junction, no deflection will

occur until the temperature at which the couple develops 6 millivolts

is reached—when the opposing E.M.F. will be overcome. This temperature

may be 500° C., so that the whole scale may be divided up between

500° and 1000°. The length of the indicator scale is thus effectively

doubled; and by using different values for the set-up, it is evident

that any desired range may be obtained within the limits of sensitivity

of the indicator. The method of procuring the opposing E.M.F. varies

with different makers. The Cambridge and Paul Instrument Company employ

a dry cell and a series resistance, connected so as to oppose the

thermocouple; and by adjusting the resistance any desired set-up may be

obtained, the value of which, in degrees, may be read off by connecting

the cell and resistance to the indicator, the couple having been

switched out of the circuit. Thus, to adjust for a range of 500°-1000°

on an indicator giving full-scale deflection for 500°, the resistance

is regulated so that the cell alone causes the pointer to move to the

end of the scale. The method adopted by Paul consists of suitable

resistances inserted in a Wheatstone bridge, which may be thrown off

the balance, and thus cause an opposing E.M.F. of the correct amount at

the terminals of the indicator.

A mechanical set-up has been introduced by the Cambridge and Paul

Instrument Company, the indicator in this case having a suspended

coil. By turning a milled-head a twist may be given to the suspending

strip, and by the turning of a second head the pointer may be brought

back to zero, retaining the initial twist, which is opposed to that

produced by the current due to the couple. Thus, if the imposed twist

were such as to move the pointer to the 400° mark on the scale, the

temperature indicated by the junction would be the observed reading

plus 400. By this method it is possible to obtain any desired range

within the limits of the indicator. The danger of producing errors due

to “creeping” is said to be negligible.

[Illustration: FIG. 19.—CIRCUIT OF NORTHRUP’S “PYROVOLTER.”]

=Potentiometer Indicators.=—The advantage of measuring E.M.F.

by the potentiometer method is that the result is independent of the

resistance of the circuit under test, whereas an indicator is affected

by changes in the resistance of the circuit in which it is inserted.

When long leads are used to connect a couple to its indicator, notable

errors may be caused by the varying resistance of the leads, due

to changing temperature; and, in addition, the resistance of the

couple-wires varies according to temperature and depth of insertion in

the furnace. Attempts have therefore been made to produce indicators

based on the potentiometer principle, suitable for workshop use, and

one form, known as Northrup’s “Pyrovolter,” is arranged as shown in

fig. 19, A. A cell D sends a current through a rheostat R, a copper

coil C, and a manganin coil S. The copper coil has the same resistance

as the copper winding of the indicator G. The couple is connected,

with G in circuit, across the manganin coil S, the resistance of this

material being unaffected by temperature. By adjusting R until no

deflection is shown on G, the drop of volts across S is made equal

to the E.M.F. of the couple. To measure this drop, a key is pressed,

altering the circuit as shown in B, the indicator being now in series

with S and the couple detached. The value of the current passing

through S is unchanged, as the indicator coil has the same resistance

as the copper coil C, which it now replaces. The deflection on G

indicates the value of this current, and, as the drop of volts across

S is proportional to the current, G may be marked off to read E.M.F.

and the corresponding temperature of the junction. The advantages

claimed are that the indicator may be used with any type of junction,

and is unaffected by temperature changes in the circuit. A similar

instrument is made by the Brown Company of Philadelphia. Up to the

present potentiometer indicators have not been adopted to any extent

in Britain, and the adjustments necessary to obtain a reading must be

accounted a distinct drawback from a workshop standpoint.

=Recorders for Thermo-electric Pyrometers.=—It is frequently of

importance to know not only the existing temperature of a furnace, but

also the fluctuations to which it is subject. Continuous observation of

a pyrometer would involve too much labour, and it is therefore evident

that an automatic recorder would possess many advantages in such cases.

A continuous record shows whether the attendant has maintained the

temperature between the prescribed limits, and furnishes a permanent

history of a given operation, which often serves as a guide to future

procedure.

The first successful recorder, suggested by Sir W. Roberts-Austen and

designed by Gen. Holden, F.R.S., was used in conjunction with a mirror

galvanometer. In its original form, the spot of light from the mirror

was made to fall on a sensitized plate, to which a gradual vertical

motion was conveyed by connecting the dark slide to a water-float by

means of a chain and pulley. The float was placed in a tank of water,

which was gradually emptied through a tap, causing the float to sink

and the plate to rise. If the deflection of the spot of light remained

steady, a vertical straight line was traced on the plate, fluctuations

producing a sinuous line. Trials at known temperatures enabled a

standard plate to be obtained, divided into degrees, which could be

superposed on a trial plate, and the temperatures thus determined. Much

valuable work was accomplished with this recorder by Roberts-Austen

for the Alloys Research Committee of the Institution of Mechanical

Engineers.

[Illustration: FIG. 20.—ROBERTS-AUSTEN RECORDER.]

In its modern form (fig. 20) the photographic plate is replaced by a

sheet of sensitized paper wound round a drum which rotates at a known

rate—say, once in 12 hours—by means of internal clockwork, shown to

the left of the figure. The galvanometer is placed at the opposite

end, and the mirror is illuminated by means of an electric lamp placed

externally, the rays from which are reflected from a prism in the

interior on to the mirror. The ray of light leaving the mirror is

broken into two portions, one of which passes through a narrow slit on

to the sensitized paper, whilst the other portion is reflected on to

a ground-glass scale on the lid, divided so as to read temperatures.

In this manner the arrangement serves not only as a recorder, but

also indicates the existing temperature without necessitating the

examination of the sensitized paper. The whole arrangement is made

impervious to light, so that it may be used in daylight. A dark room is

necessary for fixing the records. When desired, records of two or more

pyrometers may be taken on the same sheet, a clockwork device being

used to switch each instrument in turn on to the galvanometer for a

given period, an external dial indicating which pyrometer is for the

time being in circuit.

Whilst it is a drawback to the use of this recorder that the record is

not visible, the use of a mirror galvanometer confers a high degree of

sensitiveness to the instrument, not possessed by the recorders to be

described subsequently.

[Illustration: FIG. 21.—PRINCIPLE OF THREAD RECORDER.]

=The Thread Recorder.=—In this instrument an intermittent record

is secured in ink, possessing the advantages of visibility during

the period over which readings are taken, and of permanence without

subsequent treatment of the chart. The principle is shown in fig.

21, where A is a boom terminating in a V-shaped piece of ivory, and

attached to the galvanometer suspension B. By means of a cam E, rotated

by clockwork, a bar D is made to descend at stated intervals, pressing

the end of A on to an inked thread G, and causing the thread to touch

a paper wound round the drum C. This drum rotates on its axis once in

25 hours by the action of internal clockwork. The continued rotation

of the cam E alternately raises and depresses the boom A, leaving it

free for a sufficient time to enable it to attain the position it would

occupy if the mechanism were absent. The thread G is passed over

pulleys, and is wound round through an ink-well, so that the portion

opposite A is always moist. With the bar D descending every two

minutes, the successive dots form a nearly continuous line. The paper

on C is divided horizontally into temperatures, and vertically into

time units, so that the temperature existing at any given time may

readily be ascertained. The front of the bar D, or a separate strip

parallel to it, is divided so as to enable temperatures to be read

without reference to the chart. The actual instrument is shown in fig.

22. When several simultaneous records are required, the drum C is

extended, and other galvanometers introduced, to which the separate

pyrometers are connected. Several records can be taken on one chart by

introducing a clockwork mechanism to couple each pyrometer in turn to

the one galvanometer.

[Illustration: FIG. 22.—THREAD RECORDER.]

[Illustration: FIG. 23.—SIEMENS’ RECORDER.]

=The Siemens Recorder.=—In this instrument (fig. 23) the boom

from the galvanometer terminates in a knife-edge, and moves over a

thin horizontal rail, the top of which is rounded. Between the rail

and the boom are placed an inking ribbon and a paper chart, which is

moved forward by clockwork. A chopper-bar, also actuated by clockwork,

descends at about half-minute intervals, and depresses the end of the

galvanometer boom, thus producing a small dot on the chart. The paper

is 12 cms. wide and 40 yards long; it is divided into time and

temperature units, and moves forward at the rate of 2 cms. per hour.

Levelling screws are fixed to the base of the recorder.

[Illustration: FIG. 24.—FOSTER’S RECORDER.]

=Foster’s Recorder.=—Foster’s recorder (fig. 24) is designed

for use with base-metal couples of the nickel-chromium type, known as

Hoskin’s alloys, which yield an E.M.F. about five times as large as

a platinum-rhodioplatinum couple. The force available in this case

enables the coil of the galvanometer to be pivoted in a horizontal

position, the pointer being vertical, and yet to be sufficiently

sensitive. The chart is mounted on a vertical plate which rotates on

its axis, the time ordinates taking the form of concentric circles,

which are cut at an angle by the temperature ordinates. At the terminus

of the pointer is placed a small capillary tube, fitted with an inked

wick, which, when pressed upon the chart, makes a mark. The presser-bar

is curved to the same radius as the pointer, and carries a pad wetted

with ink, so that at each depression the supply of ink to the wick is

replenished by an amount equal to that imparted to the chart. This

recorder is sometimes fitted with special contacts, so that when the

correct temperature exists an electric lamp with a white bulb remains

lighted; whereas when too low or too high a green or red lamp is lit

up, and an alarm thus given. Such an addition involves the use of a

relay circuit, but is advisable in cases where expensive articles

might suffer if overheated. It can be modified to permit of several

simultaneous records being taken, and possesses the advantage that the

whole chart is visible at any time. On the other hand, the circular

coordinates may be accounted a drawback by some, as not being quite

so familiar to read as charts in which the lines are straight. Robust

construction is a feature of this recorder.

[Illustration: FIG. 25. PAUL’S RECORDER.]

=Paul’s Recorder.=—In the recorders previously described, the

motive power is furnished by clockwork. R. W. Paul has introduced an

instrument in which all the moving parts are actuated by a motor driven

with power from the mains. This recorder is shown in fig. 25. The

motor is furnished with a special type of governor to ensure constant

speed, and is connected by suitable gearing to the mechanisms moving

the chart, presser-bar, and inking ribbon, provision being made to vary

the speeds of these movements by changing the gear. The galvanometer

is of uni-pivot pattern, and the pointer is pressed at intervals on to

a typewriter ribbon which lies above the chart. Immediately beneath

the ribbon is placed a thin metal rod over which the paper passes, and

the result of the contact is to produce a small dot. As in the thread

recorder, the chart is divided into rectilinear coordinates, the ribbon

in this case serving the same purpose as the thread in the former

instrument. The lower part of the recorder is prolonged so as to

display a considerable length of the chart, which is in the form of

a roll, and is drawn forward by the mechanism. When two records are

taken simultaneously the ribbon consists of two strips, one moistened

with black ink, and the other with red; and it is arranged that each

strip in turn is over the thin rod on to which the pointer is pressed,

so that the records appear in separate colours. This recorder can also

be arranged for multiple records, or fitted with a scale-control.

With a view to workshop use, all the covers are fitted with faced

metal joints, which are much better for keeping out dust than wooden

ones. A further useful feature is that the various units in the

recorder—galvanometer, motor, feed and record mechanism, and reducing

gear—are all separate and interchangeable. By introducing a suitably

divided chart this recorder will also serve for a radiation pyrometer,

or, as will be shown later, for a resistance pyrometer.

[Illustration: FIG. 26.—LEEDS-NORTHRUP RECORDER.]

=The Leeds-Northrup Recorder.=—The Leeds and Northrup Company,

of Philadelphia, manufacture a recorder which is largely used in the

United States. As in Paul’s recorder, all the mechanism is motor

driven; but the other arrangements are entirely distinct. Instead of

measuring the deflection of the pointer, a zero deflection method is

used. The pyrometer forms part of a potentiometer circuit, and the

function of the mechanism is to oppose an E.M.F. equal to that of the

pyrometer, from which the temperature is known. This has the advantage

that the measurement is independent of the resistance of the leads, and

is capable of great accuracy. The manner in which the adjustment of the

opposing E.M.F. is controlled may be understood from fig. 26, in which

the galvanometer coil is shown at the top of the figure. The shaft from

the motor carries four cams, B, C, D, D, and at each revolution the cam

B raises the bar (5) so as to press it against an arm attached to the

galvanometer coil. At the same moment the cam C pushes against the bar

(3), and thereby releases a clutch (2) from the disc beneath. As shown,

the boom from the coil is to the right of the central position, and

is gripped between a bar (5) and the lever (4) when the former rises,

causing an angular movement of the clutch-arm (2). As the rotation

continues the cam C leaves the bar (3), which then springs back and

engages the clutch on the disc. The cam D then descends and presses

on the projection of the clutch-arm to the left, causing the disc to

rotate. The movement of the disc is conveyed to an arm which moves over

the slide wire of the potentiometer; and this movement continues until

the galvanometer boom is in the central or zero position, when neither

of the levers 4, 4 is gripped, and consequently the disc is not fed in

either direction. If the boom swing to the left, the movement of the

disc will evidently be in the converse direction to that described.

In this recorder considerable power is available to drive printing or

other mechanisms. The arm moving over the potentiometer wire carries a

pen which marks the moving chart, or, when several records are taken

simultaneously, a stamping machine is used which impresses the number

of the pyrometer on the chart. The same galvanometer mechanism serves

also for use with resistance pyrometers, as will be explained later.

=Control of Furnace Temperatures.=—Many attempts have been made

to secure the automatic regulation of furnace temperatures by means of

mechanisms controlled by an indicator or recorder. In the arrangement

employed by the Brown Company of Philadelphia, movable stops are

provided, which may be brought to any part of the scale, the mark

between the stops representing the temperature it is desired to

maintain. The indicator (or recorder) is provided with a presser-bar

which descends periodically; and if the temperature be too low the

depressed pointer completes a circuit through the inner stop, whilst

if too high the circuit is through the outer stop. Both circuits

contain a relay which brings a mechanism into operation, the result

being to increase the supply of electricity or gas if the temperature

be too low, or to diminish the supply when too high. When correct,

the depression of the pointer fails to complete either circuit, and

in this manner control between small limits may be ensured. In the

case of large furnaces the relay circuits are employed to light lamps

of different colours, the adjustment then being made by the man in

charge of the furnace. Arrangements of this kind effect a considerable

saving in fuel by preventing unnecessary heating, and are particularly

valuable in cases where overheating would be deleterious to the

articles in the furnace. The future will probably witness considerable

developments along these lines.

=Contact-Pen Recorders.=—The force with which the pointer of an

indicator is urged over the scale is relatively small, particularly in

the case of pyrometers in which the platinum series of metals are used,

as these furnish only a low E.M.F. If, therefore, the pointer terminate

in a pen which is in continuous contact with the record-paper, the

friction thus occasioned interferes considerably with the free movement

of the pointer. When cheap-metal pyrometers are used, which yield a

much higher E.M.F., the use of the pointer as contact-pen becomes

more feasible, and if uniform friction at all parts of the paper can

be ensured, records may be taken in this manner; and a recorder so

constructed is simpler and cheaper than those of the intermittent

type. Contact-pen recorders are used in America to some extent, being

made by Bristol, Brown, and others; but so far British makers have not

developed the manufacture of these instruments. At present, contact-pen

recorders must be considered less accurate and reliable than those in

which the contact is intermittent.

=Installations of Thermo-electric Pyrometers.=—When a number of

furnaces in the same establishment are to be controlled, considerable

economy may be effected by making one indicator serve for all the

couples, which in this case must necessarily be made up of wires

identical in thermo-electric value. Such an arrangement is shown in

fig. 27, in which H^1 and H^2 represent two couples, one wire from

each being connected to one of the terminals of the galvanometer G.

The other terminal of the galvanometer is connected to the arm D of

a switch, and the remaining thermocouple leads are connected to the

points 1 and 2 respectively on the circumference. As shown, H^1 is

connected to the galvanometer, and by turning the arm D to the point

2 the other couple would then be connected. Any number of junctions

may thus be arranged with a single indicator. When this arrangement

is adopted in a workshop, it is advisable to construct a small wooden

building at a spot convenient for most of the furnaces, in which the

indicator and switchboard are kept, and which could also contain a

recorder if necessary; a spot as free as possible from vibration being

preferable. Separate indicators are only necessary when a furnace is

used for special work.

[Illustration: FIG. 27.—CONNECTIONS FOR AN INSTALLATION OF

PYROMETERS.]

In some instances a second indicator is kept in the shop office, to

which all the pyrometers are wired, and which serves as a standard.

The scale of the office indicator is checked daily at one point;

and by connecting a given couple first with the shop indicator, and

immediately afterwards with the office standard, any errors can be

detected. It is also possible to ascertain the temperature of any given

furnace in the office at any time, and so to control the whole. In

fixing up such an arrangement it is necessary that each couple and its

leads, up to the indicator, should possess the same resistance, or

should not differ by an amount sufficient to affect the readings. The

general experience of a properly managed installation is that the cost

is saved in a few months in fuel alone; and, in addition, the work is

carried out to much better advantage owing to complete control from the

office.

=Management of Thermo-electric Pyrometers.=—Generally speaking,

thermo-electric pyrometers give little trouble in practice, but the

management should always be placed in skilled hands. It is advisable

to test each instrument periodically at a fixed point near the working

temperature, by the method explained on page 57; and if two or three

pounds of material be used, the protecting shield need not be removed.

A useful material for checking pyrometers near the critical range of

steel is an alloy of 60 per cent. of copper and 40 per cent. of tin,

which gives a well-defined freezing point at 738° C., and which may be

used indefinitely in a reducing atmosphere Any serious error is easily

detected by observing that the indications differ widely from those

generally obtained under the same working conditions. If an error of

20° C. or more is noted, it is advisable to form a new junction, as the

discrepancy will probably become greater, being due to a change at the

hot junction. A small error, of the nature of 5 or 10° C., may be due

to “creep” in the indicator, which may be adjusted accordingly, or

a numerical correction may be made when taking a reading. An iron

protecting sheath may be saved from rapid oxidation by black-leading

once per week, which greatly prolongs its useful life, but should be

replaced immediately it becomes dangerously thin in any part. Coating

with aluminium powder also greatly prolongs the life of an iron sheath.

When used in lead baths, the immersed part, if of iron or steel, should

be bored from the solid, and left thick at the portion opposite the

surface of the lead, where most corrosion occurs. A graphite tube,

or one made of a composition containing graphite, is often useful in

cases where iron is readily corroded, and can be used to much higher

temperatures.

When a number of instruments are in use, it is advisable to keep a

standard pyrometer for checking purposes, preferably one which has been

certified by the National Physical Laboratory. In conducting a test,

the couples, with protecting-tubes removed, may be placed in the tube

of an electric furnace of the type shown in fig. 29, in close proximity

with the standard junction. On raising the temperature gradually, the

readings of each working instrument may be compared with the standard,

and the necessary corrections discovered. Care must be taken to prevent

contact with the furnace tube, and this may be accomplished by passing

the wires through an asbestos stopper fitted into the end of the tube.

When recorders are used the attendant should make himself thoroughly

conversant with the details of the mechanism, so as to be able to

remedy any minor ailments, which are, as a rule, easily cured. On no

account should an unskilled workman be trusted with recorders; it is

better and safer to keep these in the office, where they will not be

likely to be damaged or tampered with. All records should be kept

for future reference, properly dated, and labelled according to the

operations represented.

=Laboratory Uses of Thermo-electric Pyrometers.=—Numerous

operations carried out in muffle furnaces at prescribed temperatures

require no special precautions beyond those previously given.

In determining the melting points of metals or alloys, however,

a porcelain or silica sheath is inadvisable, as they are easily

corroded. An iron sheath is proof against some metals, but not

against others, and it is always safer to fix a thin fireclay sleeve,

closed at the end, over the part immersed. A sheath of graphite or

graphite composition may be used for temperatures above 1100° C.; and

occasionally a sheath bored from a thick arc-lamp carbon, coupled to

an iron tube beyond the heated part, will be found useful at high

temperatures. Alundum is useful up to 1600° C, and for temperatures of

this order the higher refractories such as silfrax and zirkite may also

be used to advantage.

[Illustration: FIG. 28.—DIFFERENTIAL METHOD FOR DETERMINING CRITICAL

POINTS OF STEEL.]

The determination of the “critical” points of steel call for special

mention. In cooling down a mass of steel the fall of temperature is

arrested at one or more points, observations of which are frequently

of service in deciding the subsequent treatment of the steel. A

method commonly employed is known as the “differential method,” and

is indicated in fig. 28. The sample of steel, A, is placed side by

side with a piece of nickel, B, of equal dimensions, in the tube of

an electric furnace. A naked junction, C, is placed in a hole drilled

in A, and is connected to the galvanometer G, which is calibrated to

read temperatures. A two-junction circuit, formed of a junction D

placed in the hole in A, and another junction E located in the hole in

B, are connected to a delicate galvanometer H. The furnace is heated

until the galvanometer G indicates 900° C., when the arrangement is

allowed to cool. As A and B, under normal circumstances, cool at an

equal rate, the junctions D and E will be at the same temperature, and

no deflection will be observed on H. When, owing to recalescence, the

cooling of A is arrested, B, not being thus affected, will continue to

cool, thus producing a difference between the temperatures of D and E,

and consequently a deflection on H. The temperature of A at the time

this occurs is read off on G.

[Illustration: FIG. 29.—ELECTRIC TUBE-FURNACE.]

The furnace illustrated in fig. 29 is suited to this determination. It

consists of a silica tube 1 foot long, wound with a special resistance

wire and efficiently lagged, and may be heated in safety to 1000° C.

for long periods, and to 1200° for a short time. It may be placed

across the electric mains directly, and reaches 900° C. in less than

half an hour. It consumes 600 to 700 watts at the highest temperatures,

and the cost of re-winding is small. This furnace is useful as a

general laboratory appliance, and may be kept at a given steady

temperature by the use of an external resistance.

The wires in this experiment should be platinum and iridioplatinum or

rhodioplatinum, or a good pair of base metals, and the junctions in A

should be separated from each other and from the specimen by asbestos;

the same precaution being taken to prevent the junction E from touching

B. A thin layer of mica should be used beneath A and B, to avoid

contact with the furnace tube, which, when hot, allows of leakage of

current from the heating coil. Both A and B may be 1½ in. long, ¾ in.

diameter, with a hole ¼ in. diameter drilled to a depth of ¾ inch.

An alternative method is to insert a junction in a hole in the

specimen, and to take direct readings as the temperature slowly

rises or falls, when an arrest in the movement of the pointer of the

indicator shows that the change-point has been reached. Special sets

are made for this purpose.

=Measurement of Lower Temperatures by the Thermo-electric

Method.=—Many cases arise in practice in which a thermal junction

and a sensitive galvanometer are preferable to a mercury thermometer;

and below -39° C., at which temperature mercury freezes, a thermal

junction is frequently better to employ than an alcohol or pentane

thermometer. A number of practical examples of the use of thermal

junctions for ordinary and low temperatures will now be considered.

=Measurement of Surface Temperatures.=—A mercury thermometer,

when laid on a hot surface, only touches along a line, and does not

show the true surface temperature. The construction of a thermal

junction suitable for this purpose is described on page 41, and for

steam-pipe surfaces, hot plates, and the exterior of furnaces, a

specially calibrated millivoltmeter, giving a full-scale deflection

with 20 millivolts, may be used. In making the temperature scale,

boiling water (100°C.), boiling aniline (184° C.), and melting tin

(232° C.) are convenient standards. If the surface temperature be less

than 100° C. a mirror galvanometer should be used, and the junction

standardized in paraffin wax (freezing point usually about 50° C.,

but should previously be determined with an accurate thermometer),

absolute alcohol at boiling point (79° C.) and boiling water. The

author has found that this method yields excellent results in the case

of steam-pipes, the exterior of rotary cement kilns, and hot surfaces

generally.

=Measurement of Low Temperatures.=—Junctions of iron and

constantan, Hoskin’s alloys, copper and German silver, or copper and

constantan, are suited to these measurements. In a laboratory the

cold junction may be kept in ice in a Dewar vessel, the mechanically

protected form known as the “Thermos” flask being very useful for this

purpose. With a good mirror galvanometer precise readings may be

secured, 1/10 of a degree C. being easily detected. Calibration between

-40° and +40°C. may be effected by comparison with a standard mercury

thermometer, a water-bath being used above 0°, and alcohol surrounded

by a freezing mixture of ice and calcium chloride crystals below

zero. For very low temperatures (-200°C. or less) the junction maybe

calibrated in solid carbon dioxide (-78°C.) and liquid air (-184°C.).

Dewar has found that copper and German silver form a reliable junction

for very low temperatures, and the author has successfully used a

couple of Hoskin’s alloys for special work down to -200°C., a pivoted

indicator being employed. No couples tested show a linear relation

between E.M.F. and temperature at these low ranges.

Owing to the magnitude of the error caused by changes in the cold

junction, the thermo-electric method is not suited to the measurement

of atmospheric temperatures, or for explosive magazines or cold stores.

In such cases instruments of the resistance type, to be described

later, are used.

=Temperature of Steam, Exhaust Gases, etc.=—For measuring the

temperature of ordinary or superheated steam, the exhaust gases from

internal combustion engines, etc., iron-constantan junctions, with

suitable indicators, are satisfactory. When placed in a pipe the

junction should be as nearly as possible in the centre, so as to

avoid the cooling effect of the walls. Several junctions, situated in

different parts of the pipe, may be used with a single indicator and

suitable switchboard. The above remarks also apply to the hot-blast for

blast furnaces, and similar instances where the temperature does not

exceed 900° C.

=Measurement of Differences of Temperature.=—Cases frequently

arise in practice in which the difference in temperature between two

points is required, and if this difference be subject to rapid changes,

a mercury thermometer, from its large mass, would not respond with

sufficient rapidity to indicate these changes. In such cases a circuit

is made after the manner of fig. 2, one junction being located at

each point; thin wires of iron and constantan being used. For small

differences—1° C. or less—a mirror galvanometer should be used.

Calibration may be performed by placing one junction in hot water and

the other in cold, the water temperatures being read with an accurate

thermometer.

=Advantages of the Thermo-electric Method of Measuring

Temperatures.=—Compared with other methods, the thermo-electric

possesses the following points of superiority:—(1) Simplicity, no

special experiment being necessary to obtain a reading; (2) cheapness

of outfit; (3) adaptability to a variety of purposes; (4) ease of

repair in case of damage; (5) robustness, not being liable to get

out of order under workshop conditions; and (6) suitability to the

purpose of a centrally controlled installation. The drawbacks are:—(1)

Liability to error owing to fluctuations in the cold junction (which

may be avoided with care); and (2) lack of sensitiveness at very high

temperatures compared with the resistance method—a point seldom of

great practical importance, as the limit of accuracy is usually within

the amount by which an ordinary furnace fluctuates in temperature under

working conditions.