CHAPTER II.

THE PETROGRAPHICAL MICROSCOPE.[16]

The _petrographical microscope_ is essentially an ordinary

microscope[17] with the following important additional equipment. It

must be provided with: 1° a _polarizer_ (a piece of apparatus for giving

polarized light) placed below the stage; 2° an _analyzer_ (a piece of

apparatus for analyzing the rays of light after they have passed through

the polarizer and transparent section) placed between the objective and

the eye; 3° a _stage rotating_ about an axis which is the line of sight

of the microscope. A convenient type of microscope is that made by

Seibert of Wetzlar (No. 11 _a_), Fig. 2.

The reflector _a_ is usually fitted with a plane mirror on one side, and

a parabolic mirror on the other. The plane mirror should be used with

sunlight, and the parabolic mirror with artificial light, in order to

make the rays of light as nearly parallel as possible.

The _polarizer_ is in most cases a nicol prism, set in a suitable frame

_b_, and made as follows:

For the _nicol prism_ “a cleavage rhombohedron of calcite (the variety

Iceland spar is universally used in consequence of its transparency) is

obtained, having four large and two small rhombohedral faces opposite

each other. In place of the latter planes, two new surfaces are cut,

making angles of 68° (instead of 71°) with the obtuse vertical edges;

these then form the terminal faces of the prism. In addition to this,

the prism is cut through in the direction HH′, Fig. 3, the parts then

polished and cemented together again with Canada balsam. A ray of light,

_ab_, entering the prism, is divided into two rays polarized at right

angles to each other. One of these, _bc_, on meeting the layer of balsam

(whose refractive index is less than that of the ray _bc_), suffers

total reflection, and is deflected against the blackened sides of the

prism and extinguished. The other, _bd_, passes through and emerges at

_e_, a completely polarized ray of light, that is, a ray with vibrations

in one direction only, and that the direction of the shorter diagonal of

the prism.”[18] The vertical plane through the shorter diagonal may be

called the _plane_ of _vibration_[19] of the nicol.

[Illustration:

FIG. 2.

]

[Illustration:

FIG. 3.

]

The polarizer must be below the stage _c_, and is generally adjusted so

as to have its plane of vibration parallel to the N. and S. cross-wire

in the eye-piece _o_. It is important to know the direction of the plane

of vibration of the polarizer or lower nicol, as we can then determine,

when absorption of light occurs in a mineral, the direction in this

mineral parallel to which the absorbed rays are vibrating. The polarizer

slides in an outer shell or frame, and, by means of a lever _d_, can

easily be raised or lowered.

A convenient test for the location of this plane of vibration of the

polarizer is as follows: Make use of a section of biotite, cut at right

angles to the basal plane, hence showing the basal cleavage cracks.

Biotite has the property of absorbing to a marked extent the light

vibrating parallel to these cleavage cracks. Rotate such a section on

the stage of the microscope until the position of maximum darkness is

reached and when such is the case the plane of vibration of the

polarizer must be parallel to these cleavage cracks.

On the top of the nicol is placed the _condensing lens_ for getting

_convergent_ light, and the adjustments are so arranged that when the

nicol is up as far as it will go, the condensing lens[20] is brought

almost in contact with the lower surface of the transparent section

resting on the stage.

The _rotating_ circular _stage e_ is supported on a suitable frame _c_,

and arranged so that its axis of rotation coincides with the line of

sight of the microscope. The stage is graduated, and, by means of an

index fixed to the frame, the angular rotation can always be obtained.

It is also provided with two adjusting screws[21] _f_, by means of which

the axis of rotation can be accurately centered.

The method of centering is as follows: Bring some prominent mark in the

section exactly in coincidence with the intersection of the cross-wires

in the eye-piece. Rotate the stage 180°, and correct one half the error

by means of the centering screws, and the other half by moving the

section on the stage. Check the result by rotating the stage 180° again,

and if necessary repeat the corrections in the same way until the

adjustment is satisfactory.

The _objective_[22] _g_ screws into the collar _h_, which has a slot

_k_, in the upper portion, for the introduction of a sensitive color

plate, a one quarter undulation mica plate or a quartz wedge.[23]

The slot _k_ is so arranged that, when the sensitive color plates are

introduced, the vibration directions of these plates will make an angle

of 45° with the planes of vibration of the crossed nicols, and the

interference color will thus be at its maximum intensity.

A revolving _nose-piece_ is sometimes used which can be attached to the

collar _h_ and arranged to carry two or three objectives, which can thus

be very quickly brought into position for use. This is convenient in

passing rapidly from observations with parallel light to observations

with convergent light, which must be made with a high power objective.

The difficulty is that recentering is generally required. The modern

microscopes are provided with a clip for holding the objectives, instead

of the screw-thread collar _h_.

The _analyzer_[24] or upper nicol is contained in the frame _l_, which

is arranged so as to slide in and out of the tube of the microscope.

The plane of vibration of the analyzer is fixed by the instrument maker

so as to be at right angles to the plane of vibration of the polarizer,

hence in the Seibert microscope parallel to the E. and W. cross-wire in

the eye-piece. Consequently when the frame _l_ is pushed into the tube,

the analyzer is introduced in the line of sight between the objective

and the observer’s eye, with its plane of vibration at right angles to

the plane of vibration of the polarizer; that is the nicols are crossed.

When the nicols are crossed, if they are properly adjusted, no light can

pass through to the eye and the field of view should be dark.[25]

The _eye-piece_[26] _o_ fits into the top of the tube, and, by means of

a little projecting piece fitting into a slot in the frame, can always

be adjusted so as to have its cross-wires parallel to the planes of

vibration of the two nicols.[25]

Some instruments are provided with an additional slot in the tube

between the analyzer and the eye-piece for the introduction of a

Bertrand lens, which is used to magnify the interference figures

produced by convergent light.

The first approximate focusing is made by the screw _m_, and the fine

adjustment by the screw _n_.

In focusing always start with the objective very near the section, and

move it away until the right focus is obtained. Never move down towards

the section in obtaining the focus, as there is danger then of striking

the objective against the section.

For cleaning the microscope xylol can be used, as it will not injure the

lacquer. To lubricate any of the parts use a small quantity of soft

tallow, good clock oil or paraffin oil.