Mass Spectrometry

The mass spectrometer measures isotopic abundances using a magnetic

field to sort electrically charged particles into groups according to

their masses. It works this way: A small drop of material to be analyzed

is placed on a metal filament and dried. The filament, in its holder, is

placed inside the mass spectrometer, and heated electrically in a

vacuum, like the filament in a light bulb. As the wire begins to glow,

some of the sample begins to radiate, or “boil off”, losing an electron

or two in the process. In other words, some of the atoms will be changed

into positive IONS.

An alternative method is to introduce the sample material into the

vacuum chamber in the form of a gas (like argon, for example), and then

bombard the gas with electrons streaming from a hot filament. The

electron stream will knock some electrons off the gas molecules and this

also will produce positive ions. Either process of ion production is

satisfactory, depending on the problem to be tackled, but the mass

spectrometers for the two methods are naturally quite different.

Whichever way the ions were produced, they are next exposed to a strong

electric field, accelerated, and electrostatically focused into a beam.

These charged particles are directed into a magnetic field between the

pole faces of an electromagnet. The magnet does the analyzing by the

principle of magnetic deflection that was known to André Ampere and

Michael Faraday more than a century ago. Any moving electric charge has

a magnetic field associated with it. This field interacts with the field

of the analyzing magnet to impress a deflecting force on the charge. The

force acts at right angles to the direction the charge travels and also

at right angles to the direction of the impressed magnetic field. The

pull of this force depends only on the electric charge and the speed of

each particle: A light single-charged particle will be deflected more

than a heavier particle with the same charge. In this way, the ions in

the beam are sorted out into a number of separate beams, each made up of

particles of the same charge/mass ratio. Each beam contains one isotope

of the original material, because isotopes differ on the basis of their

mass. By adjusting the current in the electromagnet we can direct these

separate beams into a “collector” and electrically measure their

intensity one by one. This gives the relative abundance of the separate

isotopes in the sample.