TABLE III.--_Solution of Complete Calibration._

+------------+------+------+------+------+------+------+------+------+------+------+

| Step No. | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | 9 | 10 |

+------------+------+------+------+------+------+------+------+------+------+------+

| 1 | 0 | - 5 | +11 | +20 | +34 | +25 | + 7 | +26 | +23 | +32 |

| 2 | + 5 | 0 | +16 | +23 | +39 | +29 | +12 | +31 | +28 | +37 |

| 3 | -11 | -16 | 0 | + 8 | +24 | +13 | - 4 | +15 | +13 | +22 |

| 4 | -20 | -23 | - 8 | 0 | +15 | + 5 | -12 | + 7 | + 4 | +13 |

| 5 | -34 | -39 | -24 | -15 | 0 | - 9 | -26 | - 8 | -10 | - 2 |

| 6 | -25 | -29 | -13 | - 5 | + 9 | 0 | -17 | + 2 | - 1 | + 8 |

| 7 | - 7 | -12 | + 4 | +12 | +26 | +17 | 0 | +19 | +16 | +26 |

| 8 | -26 | -31 | -15 | - 7 | + 8 | - 2 | -19 | 0 | - 3 | + 6 |

| 9 | -23 | -28 | -13 | - 4 | +10 | + 1 | -16 | + 3 | 0 | + 9 |

| 10 | -32 | -37 | -22 | -13 | + 2 | - 8 | -26 | - 6 | - 9 | 0 |

+------------+------+------+------+------+------+------+------+------+------+------+

| Error of | | | | | | | | | | |

| step. | -17.3| -22.0| - 6.4| + 1.9| +16.7| + 7.1| -10.1| + 8.9| + 6.1| +15.1|

+------------+------+------+------+------+------+------+------+------+------+------+

|Corrections.| +17.3| +39.3| +45.7| +43.8| +27.1| +20.0| +30.1| +21.2| +15.1| 0|

+------------+------+------+------+------+------+------+------+------+------+------+

The advantages of this method are the simplicity and symmetry of the

work of reduction, and the accuracy of the result, which exceeds that

of the Gay Lussac method in consequence of the much larger number of

independent observations. It may be noticed, for instance, that the

correction at point 5 is 27.1 thousandths by the complete calibration,

which is 2 thousandths less than the value 29 obtained by the Gay

Lussac method, but agrees well with the value 27 thousandths obtained

by taking only the first and last observations with the thread of 5

deg. The disadvantage of the method lies in the great number of

observations required, and in the labour of adjusting so many

different threads to suitable lengths. It is probable that

sufficiently good results may be obtained with much less trouble by

using fewer threads, especially if more care is taken in the

micrometric determination of their errors.

The method adopted for dividing up the fundamental interval of any

thermometer into sections and steps for calibration may be widely

varied, and is necessarily modified in cases where auxiliary bulbs or

"ampoules" are employed. The Paris mercury-standards, which read

continuously from 0 deg. to 100 deg. C., without intermediate

ampoules, were calibrated by Chappuis in five sections of 20 deg.

each, to determine the corrections at the points 20 deg., 40 deg., 60

deg., 80 deg., which may be called the "principal points" of the

calibration, in terms of the fundamental interval. Each section of 20

deg. was subsequently calibrated in steps of 2 deg., the corrections

being at first referred, as in the example already given, to the mean

degree of the section itself, and being afterwards expressed, by a

simple transformation, in terms of the fundamental interval, by means

of the corrections already found for the ends of the section.

Supposing, for instance, that the corrections at the points 0 deg. and

10 deg. of Table III. are not zero, but C deg. and C' respectively,

the correction C_n at any intermediate point n will evidently be given

by the formula,