INTRATRACHEAL INSUFFLATION OF ETHER

Intratracheal insufflation consists in driving a current of air under

pressure, through a tube introduced by way of the mouth and larynx,

deeply into the trachea. The current of air which is continuous,

returns between the tube and the wall of the trachea, and escapes

through the mouth and nose.

Certain Physical Considerations.

The work of Meltzer and Auer has demonstrated that this insufflation of

air into the trachea under adequate pressure ventilates the pulmonary

alveoli, and enables the normal diffusion of gases to be carried out

for many hours, independently of all respiratory movements. If the

air in its passage under pressure is made to pass through a chamber

containing ether, we are enabled to introduce into the pulmonary

alveoli, ether vapour of varying strength, and, by this means, to

maintain surgical anæsthesia.

* * * * *

THE ACTUAL PROCESS OF EXTERNAL RESPIRATION consists in the absorption

of oxygen from the alveoli into the blood of the lung capillaries, and

the elimination of carbon dioxide from the lung capillaries into the

alveoli. The oxygen has to be brought from the outside to the alveoli,

and the carbon dioxide has to be conducted from the alveoli to the

outside. Between the outside and the alveoli is the long airshaft,

consisting of mouth and nose, pharynx, larynx, trachea, bronchi and

bronchioles. In natural respiration the conduction of oxygen inwards,

and of carbon dioxide outwards, is carried through by a complicated

pumping mechanism. In ordinary inhalation anæsthesia, this mechanism is

entrusted with the task of introducing ether vapour into the alveoli.

In intratracheal insufflation the work of this natural pumping

apparatus is taken over by an artificial mechanism. In considering the

justification for this, the following points are to be noted:--

(1) The patient is unconscious--not naturally so as in

sleep--but unnaturally as the result of drugs; there is

therefore a probability that the elaborate natural mechanism

may not work smoothly--especially is there a danger that the

free airway may be interfered with. Intratracheal insufflation

obviates this danger.

(2) By means of the artificial mechanism, air is brought with

some force to the mouth of the bronchi, and thus a more rapid

and more powerful diffusion of gases takes place.

(3) The mechanism ensures the maintenance of a current of air

blowing forcefully from the trachea and larynx through the

pharynx, mouth and nose. This re-current continuous air-stream

effectively prevents the entrance of blood or any infectious

material into the bronchi and air cells, and thus the danger of

septic lung troubles is obviated.

(4) In certain intrathoracic operations the normal respiratory

mechanism is deliberately interfered with. Intratracheal

insufflation by the constant maintenance of sufficient positive

pressure, prevents or regulates the collapse of the lung

which occurs when the thorax is opened, and thus obviates the

necessity for the somewhat elaborate positive and negative

pressure cabinets and masks which had been devised for

intrathoracic operations.

(5) Incidentally it may be noted that intratracheal insufflation

of air provides us with an excellent means for performing

artificial respiration. Many instances have now been recorded

of its utility in this respect. Dr Elsberg records the case

of a patient who had taken morphia with suicidal intent, on

whom artificial respiration by this method was kept up for 12

hours, without any respiratory movements taking place, recovery

ultimately ensuing.

The Apparatus.

This consists of (_see_ Fig. 30):--

1. Instrument for producing air current (A).

2. Ether chamber and various regulating taps (B).

3. Device for warming vapour (C).

4. Safety valve (D).

5. Manometer (E).

6. Intratracheal catheter, with rubber tubing linking up the

various parts of the apparatus (H).

[Illustration: FIG. 30. Diagram of intratracheal apparatus.]

1. THE AIR CURRENT is obtained either by means of an ordinary

glass-blower’s foot bellows, or an electric motor may actuate a

rotatory blower which produces a current of air. The blower may be made

to rotate at a speed varying from 50 to 1000 revolutions per minute

(_see_ Fig. 31).

The foot bellows is simple and inexpensive, and there is no reason why

it should not be efficient. The electrically rotated blower is more

efficient, ensures a smoother air current, and saves much labour.

2. THE ETHER CHAMBER.--Ether vapour may be produced by either of two

methods. Fig. 32 shows Kelly’s instrument where air is blown over the

surface of a considerable quantity of liquid ether. Dr Meltzer[5]

maintains that the effectiveness of the etherization is proportional

to the diameter of the ether bottle. In this connection it may be well

to recall certain points pertinent to the subject of etherization. A

satisfactory etherization depends on the establishment in the blood

and tissues of an ether tension of definite strength. Boothby[6]

states that this tension should correspond to about 15 per cent. of

ether vapour in the alveolar cells. “If when this tension has been

established, less than 15 per cent. ether vapour is administered,

outward diffusion occurs from the tissues and blood to the air and the

anæsthesia becomes lighter.”

[Illustration: FIG. 31.--Electric blower to supply current of

air for intratracheal anæsthesia.]

[Illustration: FIG. 32. Kelly’s intratracheal apparatus.]

In an apparatus such as we are describing the strength of the ether

vapour depends on and is influenced by a variety of factors, among

which the following may be specially noted:--

(_a_) _The diameter of the ether chamber._--The larger the

diameter the stronger will be the ether vapour. The chamber is

generally kept two-thirds full: the less the empty space in the

chamber the stronger will be the vapour.

(_b_) _The rapidity of the air-current._--The larger the amount

of air passing over the ether the more rapid will be the

vapouration, with the result that the temperature of the liquid

ether will rapidly fall and the strength of the ether vapour in

the air will be correspondingly lowered.

(_c_) _The temperature of the liquid ether._--The higher the

temperature of the liquid ether the stronger will be the

percentage of ether vapour in the air. If the ether chamber is

placed in a bath of water which is maintained steadily at an

adequate temperature and if the rate of the air flow remains

constant, a constant strength of ether vapour will be given off.

It is well in this connection to remember that the boiling

point of ether is low and that a very high percentage of ether

vapour is readily obtained if the temperature in the water

bath is allowed to rise beyond 80°F. For further information

regarding ether percentages _see_ Appendix I and II.

Such an instrument as Kelly’s requires at its head a _regulating tap_,

movement of which is capable of diverting part of the air stream direct

to the patient without coming in contact with the surface of the ether.

In this way the maximum strength of vapour may be diluted as and when

required.

In Fig. 33 is shown Shipway’s instrument. Here the drug is dripped

into a chamber, the floor of which is kept warm. The ether volatilises

at once and the vapour is carried away by the air stream passing

through the chamber. The strength of ether vapour in this instrument is

regulated solely by the rate of drip which is in complete control of

the administrator.

Whichever method of making ether vapour is utilised the instrument

should be capable of producing a maximum at least of 15 per cent. to 18

per cent., and means must be provided to reduce this percentage at will.

3. DEVICE FOR WARMING VAPOUR.--It is doubtful if the warming of

inspired or insufflated vapour has any appreciable influence on

the body temperature, but there is a fairly general consensus of

opinion that if the ether vapour is warmed there is less likelihood

of irritation of the respiratory mucous membrane. It is to be noted,

moreover, that in intratracheal insufflation the natural apparatus for

warming the inspired air is put out of action. This warming can be

effected by the simple device of carrying the tube through a chamber of

hot water after it emerges from the ether chamber.

4. SAFETY VALVE.--Dr Meltzer insists very strongly on the necessity

of having a safety valve capable of controlling the maximum pressure

under which the air may enter into the intratracheal tube. By a

simple device, any excess of desired pressure will cause the air to

bubble through mercury and thus never reach the lungs. In this way any

possibility of accident from undue intrapulmonary pressure is obviated.

[Illustration: FIG. 33.--Shipway’s intratracheal

apparatus.]

5. MERCURY MANOMETER to indicate the pressure in the tube in m.m.

of mercury. It has been experimentally established[7] “that the

pressure in the trachea and in the bronchi is only a small fraction

of the pressure in the manometer outside of the body, and that the

intratracheal pressure grows considerably less with the decrease of

the diameter of the intratracheal tube.”

6. INTRATRACHEAL CATHETER.--It is essential to have an instrument of

adequate rigidity which can be satisfactorily sterilized. The ordinary

coudé catheters, or the silk web white enamelled cylindrical catheters,

are suitable. The size should be selected from a range of 18 to 25

French. It is preferable to err rather on the side of a small than a

large tube. An ordinary adult will require a tube of about size 22 to

24, a plethoric alcoholic, on the other hand, might need a 25.

A tube of too large a calibre interferes with the free return of air

and spontaneous respiration soon becomes too slow. Expiration is

prolonged, active, and laboured, and after a few minutes, respiratory

movements may cease entirely. The only way to meet such a situation is

to withdraw the tube and insert a smaller one.

TECHNIQUE OF ADMINISTRATION.--It is advisable in adults to administer

about three quarters of an hour before the operation, a hypodermic

injection of morphia (gr. ⅙) and atropine (gr. ¹⁄₁₀₀), or of

scopolamine (gr. ¹⁄₁₀₀) and morphia (gr. ⅙). The latter combination is

more efficacious in alcoholic subjects. In children, atropine alone

should be given.

It is to be remembered that intratracheal insufflation of ether is a

method of maintaining not of inducing anæsthesia. Induction is carried

out in the ordinary way. When this has been done, the catheter is

passed.

The introduction of the catheter does present some difficulty, but

this is largely overcome as skill and confidence are acquired with

practice. It may be carried out indirectly, or a view of the glottis

may be obtained by the aid of such an endoscope as Hill’s, and the

catheter inserted between the cords. The latter method is probably the

more satisfactory, but it is well to acquire the skill to pass the

catheter indirectly as in a certain small proportion of cases there are

obstacles to the use of the endoscope.

To facilitate catherization the pharynx and epiglottic region may

be cocainized with a 5 per cent. solution before induction. Hill’s

endoscope (Fig. 34) is distally illuminated by a small electric lamp,

which is connected with a small pocket battery. A useful modification

of this has been devised by Mr Dott. In it, the catheter is passed

along a separate compartment, so that the view of the glottis is

undisturbed. The point of the catheter comes into view at the distal

extremity of the endoscope and can be guided between the cords into the

trachea.

[Illustration: FIG. 34.--Hill’s Direct Laryngoscope.]

It is essential before attempting intubation, that there should be

thorough relaxation. The lower jaw should be so slack that a gag is not

required. The head is then placed in the occipito-shoulder position, or

is allowed to hang over the end of the table. The tongue is controlled

by forceps, and the endoscope passed slowly along its dorsum until the

epiglottis comes into view. The point of the endoscope is then passed

sufficiently far below the tip of the epiglottis to ensure that it will

not slip; too deep insertion must be avoided. The endoscope being held

in the left hand, the hyoid bone is lifted up by a tilting movement of

the hand. The glottis is thus brought into view. The catheter stiffened

by means of a probe is then passed through the glottis into the

trachea and the endoscope withdrawn. The bifurcation of the trachea in

the adult is at a distance of about 26 cm. from the incisor teeth. The

catheter should be marked accordingly, and inserted to a point just

short of this. The probe is then withdrawn, and connection made with

the air current.

It occasionally happens that unexpected difficulty is met with on

attempting to pass the catheter by direct vision. The larynx may be so

fixed that the glottis does not readily come into view; or in the case

of an intraoral neoplasm the view may be obstructed by the presence of

blood. In such cases the catheter can be introduced by the indirect

method. The middle finger of the left hand is passed along the dorsum

of the tongue until the epiglottis is felt. The index finger is then

used to guide the point of the catheter to the glottis through which

it is then passed. No undue force must be used. A stilette should be

inserted into the catheter which should be moulded almost to a right

angle at its terminal third. If the stilette is withdrawn when the

point of the catheter is over the mouth of the glottis, the instrument

will, as a rule, slip easily into the trachea.

Occasionally the tube passes into the œsophagus. With care and adequate

relaxation such a mistake should not occur, but the possibility of it

should be kept in mind. If the operator will abstain from attempting to

pass the catheter until such time as he has a satisfactory view of the

glottis, mistakes of this kind will seldom occur. The essentials are a

good illumination and an adequate relaxation.

Mild glottic spasm may supervene on the passage of the catheter but

this rapidly passes off. At first the degree of concentration of the

vapour should be low or irritation will result, evidenced by spasm and

coughing. The strength of the vapour is gradually increased until the

necessary concentration is attained. The pressure should vary according

to the requirements of the case and should range between 10 mm. to 25

mm. Hg. The safety valve must be set so as to make any pressure above

this impossible.

In the majority of cases the course of anæsthesia is smooth and

uneventful; the colour remains a rosy pink, the pulse is good, and the

respirations quiet and regular. It is undesirable that the respiratory

movements should be abolished altogether; their presence indicates that

neither the central nor the peripheral respiratory mechanism is being

overdosed with ether.

Theoretically, the constant plus-pressure in the lungs might be

thought to interfere with the circulation in the large veins, and

in the pulmonary vessels themselves. It is therefore well to reduce

the pressure in the catheter to zero every minute by opening the tap

provided for the purpose for a second or two.

At the conclusion of the operation, before withdrawing the catheter

it is well to flush out the lungs with air so as to remove any ether

vapour that is present. In a certain number of cases, notably in

big alcoholic subjects, difficulty may be experienced in securing

a sufficiently deep anæsthesia with good relaxation. It is seldom,

however, that patience and the careful introduction of a stronger

vapour will not suffice to overcome this. In alcoholic subjects, as

previously suggested, preliminary medication with scopolamine and

morphia will help. It really becomes a question, if one may put it

so, of coaxing the unconscious patient to tolerate an ether vapour of

adequate strength.

Advantages and Special Indications.

The general opinion of anæsthetists appears strongly to favour the view

that the absence of strain and the perfect æration in intratracheal

ether insufflation tend to lessen the shock of operation. The

post-operative history of patients also suggests that there is a

lessened liability to pulmonary complications as compared with cases

in which ether has been administered by other methods. Dr Elsberg[8]

of New York in this connection writes: “The absence of any pulmonary

complications has led us to use this method of anæsthesia in all

patients in whom pulmonary complications were to be feared after an

anæsthesia or operation. Thus on all asthmatics, in patients with

chronic bronchitis and emphysema, in patients who require gastric

resection and the like, we no longer, during two years, have seen the

much dreaded post-operative pneumonia wherever intratracheal anæsthesia

was used.”

In addition, there is ample evidence[9] that the introduction of a

catheter into the trachea and its presence therein does not tend,

as might have been expected, to set up any irritation at the time

or predispose to subsequent trouble. Apart from these general

considerations, which suggest the advantage of a somewhat extended use

of intratracheal ether insufflation, the method has obvious advantages

in all operations about the mouth, such as those for excision of the

upper jaw, those undertaken for the removal of nasopharyngeal growths

and various plastic operations involving the nasal and buccal cavities.

In such cases the danger of aspiration of blood, mucus, etc., is

obviated and the anæsthetist is well out of the surgeon’s way, while at

the same time an even interrupted delivery of ether vapour is effected.

In operations for removal of glands in the neck the surgeon has the

field to himself, and is not hampered nor is his asepsis endangered by

the proximity of the anæsthetist’s mask.

In such operations as laminectomy and nephrectomy the postural

difficulties with which the anæsthetist has to contend, and which also

tend to interfere with free respiration, are eliminated.

The great advantage of the method in intrathoracic operations has

already been referred to.

The introduction of intratracheal insufflation of ether was rendered

possible by the pioneer work of Drs Elsberg, Meltzer, and Auer. The

writer would like to acknowledge his indebtedness to their writings, of

which he has made free use.