CALCULATION FROM RESIDUAL ANALYSES.

The increment in weight of the absorbers for water and carbon dioxide

and the loss in weight of the oxygen cylinder give only an approximate

idea of the amounts of carbon dioxide and water-vapor produced and

oxygen absorbed during the period, and it is necessary to make

correction for change in the composition of the air as shown by the

residual analyses and for fluctuations in the actual volume. In order to

compute from the analyses the total carbon-dioxide content of the

residual air, it is necessary to know the relation of the air used for

the sample to the total volume, and thus we must know accurately the

volume of air passing through the gas-meter.

In the earlier apparatus 10-liter samples were used, and the volume of

the respiration chamber was so large that it was necessary to multiply

the values found in the residual sample by a very large factor, 500.

Hence, the utmost caution was taken to procure an accurate measurement

of the sample, the exact amounts of carbon dioxide absorbed, and

water-vapor absorbed. To this end a large number of corrections were

made, which are not necessary with the present type of apparatus with a

volume of residual air of but about 1,300 liters, and accordingly the

manipulation and calculations have been very greatly simplified.

While formerly pains were taken to obtain the exact temperature of the

air leaving the gas-meter, with this apparatus it is unnecessary. When

the earlier type of apparatus was in use there were marked changes in

the temperature of the calorimeter laboratory and in the water in the

meter which were naturally prejudicial to the accurate measurement of

the volume of samples, but with the present control of temperature in

this laboratory it has been found by repeated tests that the temperature

of the water in the meter does not vary a sufficient amount to justify

this painstaking measurement and calculation. Obviously, this

observation also pertains to the corrections for the tension of aqueous

vapor. It has been found possible to assume an average laboratory

temperature and reduce the volume as read on the meter by means of a

constant factor.

The quantity of air passing through the meter is so adjusted that

exactly 10 liters as measured on the dial pass through it for one

analysis. The air as measured in the meter is, however, under markedly

different conditions from the air inside the respiration chamber. While

there is the same temperature, there is a material difference in the

water-vapor present, and hence the moisture content as expressed in

terms of tension of aqueous vapor must be considered. This obviously

tends to diminish the true volume of air in the meter.

Formerly we made accurate correction for the tension of aqueous vapor

based upon the barometer and the temperature of the meter at the end of

the period, but it has now been found that the reduction of the meter

readings to conditions inside of the chamber can be made with a

sufficient degree of accuracy by multiplying the volume of air passing

through the meter by a fraction, _(h-t)/h_, in which _h_ represents the

barometer and _t_ the tension of aqueous vapor at the temperature of the

laboratory, 20° C. Since the tension of aqueous vapor at the laboratory

temperature is not far from 15 mm., a simple calculation will show that

there may be considerable variations in the value of _h_ without

affecting the fraction materially, and we have accordingly assumed a

value of _h_ as normally 760 mm., and the correction thus obtained is

(760 - 15)/760 = 0.98, and all readings on the meter should be

multiplied by this fraction.

On the one hand, then, there is the correction on the meter itself,

which correction is +1.4 per cent (see page 75); and on the other hand

the correction on the sample for the tension of aqueous vapor, which is

-2.0 per cent, and consequently the resultant correction is -0.6 per

cent. From the conditions under which the experiments are made, however,

it is rarely possible to read the meter closer than ±0.05 liter, as the

graduations on the meter correspond to 50 cubic centimeters. It will be

seen, then, that this final correction is really inside the limit of

error of the instrument, and consequently with this particular meter now

in use no correction whatever is necessary for the reduction of the

volume. The matter of temperature corrections has been taken up in great

detail in an earlier publication, and where there are noticeable

differences in temperature between the meter and the calorimeter chamber

the calculation is very much more complicated.

For practical purposes, therefore, we may assume that the quantity of

air passed through the meter, as now in use, represents exactly 10

liters measured under the conditions obtaining inside of the respiration

chamber, and in order to find the total amount of water-vapor present in

the chamber it is necessary only to multiply the weight of water found

in the 10-liter sample by one-tenth of the total volume of air

containing water-vapor.

The total volume of air which contains water-vapor is not far from 1,360

liters; consequently multiplying the weight of water in the sample by

136 gives the total amount of water in the chamber and the piping. The

volume of air containing carbon dioxide is that contained in the chamber

and piping to the first sulphuric-acid vessel plus 16 liters of air

above the sulphuric acid and connections in the first porcelain vessel,

and in order to obtain the amount of carbon dioxide from the sample it

is only necessary to multiply the weight of carbon dioxide in the sample

by 137.6.

Since in the calculation of the total amount of residual oxygen volumes

rather than weights of gases are used, it is our custom to convert the

weights of carbon dioxide and water-vapor in the chamber to volumes by

multiplying by the well-known factors. The determination of oxygen

depends upon the knowledge of the true rather than the apparent volume

of air in the system, and consequently the apparent volume must be

reduced to standard conditions of temperature and pressure each time the

calculation is made. To this end, the total volume of air in the

inclosed circuit (including that in the tension-equalizer, amounting to

1,400 liters in all) is reduced to 0° and 760 millimeters by the usual

methods of computation. The total volume of air (which may be designated

as _V_) includes the volumes of carbon dioxide, water-vapor, oxygen, and

nitrogen. From the calculations mentioned above, the volumes of

water-vapor and carbon dioxide have been computed, and deducting the sum

of these from the reduced volume of air gives the volume of oxygen plus

nitrogen. If the volume of nitrogen is known, obviously the volume of

oxygen can be found.

At the beginning of the experiment, it is assumed that the chamber is

filled with ordinary air. By calculating the amount of nitrogen in the

chamber at the start as four-fifths of the total amount, no great error

is introduced. In many experiments actual analyses of the air have been

made at the moment of the beginning of the experiment. The important

thing to bear in mind is that having once sealed the chamber and closed

it tightly, no nitrogen can enter other than that admitted with the

oxygen, and hence the residual amount of nitrogen remains unaltered save

for this single exception. If care is taken to keep an accurate record

of the amount of nitrogen admitted with the oxygen, the nitrogen

residual in the chamber at any given time is readily computed. While

from an absolute mathematical standpoint the accuracy of this

computation can be questioned, here again we are seeking an accurate

record of differences rather than an absolute amount, and whether we

assume the volume of the air in the chamber to contain 20.4 per cent of

oxygen or 21.6 per cent is a matter of indifference. It is of importance

only to note the increases in the amount of nitrogen, since these

increases represent decrease in the residual oxygen and it is with the

changes in the residual oxygen that we particularly have to do.

INFLUENCE OF FLUCTUATIONS IN TEMPERATURE AND PRESSURE ON THE APPARENT

VOLUME OF AIR IN THE SYSTEM.

The air, being confined in a space with semi-rigid walls, is subjected

naturally to variations in true volume, depending upon the temperature

and barometric pressure. If the air inside of the chamber becomes

considerably warmer there is naturally an expansion, and were it not for

the tension-equalizer there would be pressure in the system. Also, if

the barometer falls, there is an expansion of air which, again, in the

absence of the tension-equalizer, would produce pressure in the system.

It is necessary, therefore, in calculating the true volume of air, to

take into account not only the apparent volume, which, as is shown

above, is always a constant amount at the end of each period, but the

changes in temperature and barometric pressure must also be noted. Since

there is a volume of about 1,400 liters, a simple calculation will show

that for each degree centigrade change in temperature there will be a

change in volume of approximately 4.8 liters. In actual practice,

however, this rarely occurs, as the temperature control is usually

inside of 0.1° C. and for the most part within a few hundredths. A

variation in barometric pressure of 1 millimeter will affect 1,400

liters by 1.8 liters.

In actual practice, therefore, it is seen that if the barometer falls

there will be an expansion of air in the system. This will tend to

increase the volume by raising the rubber diaphragm on the

tension-equalizer, the ultimate result of which is that at the final

filling with oxygen at the end of the period less is used than would be

the case had there been no change in the barometer. In other words, for

each liter expansion of air inside of the system, there is 1 liter less

oxygen required to bring the apparent volume the same at the end of the

period. Similarly, if there is an increase in temperature of the air,

there is expansion, and a smaller amount of oxygen is required than

would be the case had there been no change; and conversely, if the

barometer rises or the temperature falls, more oxygen would be supplied

than is needed for consumption. It is thus seen that the temperature and

barometer changes affect the quantity of oxygen admitted to the chamber.

INFLUENCE OF FLUCTUATIONS IN THE AMOUNTS OF CARBON DIOXIDE AND

WATER-VAPOR UPON RESIDUAL OXYGEN.

Any variations in the residual amount of carbon dioxide or water-vapor

likewise affect the oxygen. Thus, if there is an increase of 1 gram in

the amount of residual carbon dioxide, this corresponds to 0.51 liter,

and consequently an equal volume of oxygen is not admitted to the

chamber during the period, since its place has been taken by the

increased volume of carbon dioxide. A similar reasoning will show that

increase in the water-vapor content will have a similar effect, for each

gram of water-vapor corresponds to 1.25 liters and therefore influences

markedly the introduction of oxygen. All four of the factors, therefore

(barometric pressure, temperature, residual carbon dioxide, and residual

water-vapor), affect noticeably the oxygen determination.

Respiration Calorimeters for Studying the Respiratory Exchange and Energy Transformations of Man · 第 103 / 120 题