VARIATIONS IN THE APPARENT VOLUME OF AIR.

In the earlier form of apparatus the largest variation in the apparent

volume of air was due to the fluctuations in the height of the large

rubber diaphragms used on the tension equalizer. In the present form of

apparatus there is but one rubber diaphragm, and this is small,

containing not more than 3 to 4 liters as compared to about 30 liters in

the earlier double rubber diaphragms. As now arranged, all fluctuations

due to the varying positions of the tension-equalizer are eliminated as

each experimental period is ended with the diaphragm in exactly the same

position, _i. e._, filled to a definite tension.

In its passage through the purifiers the air is subjected to more or

less pressure, and it is obvious that if these absorbers were coupled to

the ventilating system under atmospheric pressure, and then air caused

to pass through them, there would be compression in a portion of the

purifier system. Thus there would be a contraction in the volume, and

air thus compressed would subsequently be released into the open air

when the absorbers were uncoupled. The method of testing the system

outlined on page 100 equalizes this error, however, in that the system

is tested under the same pressure used during an actual experiment, and

hence between the surface of the sulphuric acid in the first porcelain

vessel and the sulphuric acid in the second porcelain vessel there is a

confined volume of air which at the beginning of an experimental period

is under identically the same pressure as it is at the end. There is,

then, no correction necessary for the rejection of air with the changes

in the absorber system.

Respiration Calorimeters for Studying the Respiratory Exchange and Energy Transformations of Man · 第 98 / 120 题