AUTOMATIC CONTROL OF OXYGEN SUPPLY.

The use of the reduction-valve has made the automatic control of the

oxygen supply much simpler than in the apparatus formerly used. The

details of the connections somewhat schematically outlined are given in

fig. 32, in which D is the oxygen cylinder, K the supporting band, A the

reduction-valve, and J the tension-equalizer attached to one of the

calorimeters. Having reduced the pressure to about 2 pounds by means of

the reduction-valve, the supply of oxygen can be shut off by putting a

pinch-cock on a rubber pipe leading from the reduction-valve to the

calorimeters. Instead of using the ordinary screw pinch-cock, this

connection is closed by a spring clamp. The spring E draws on the rod

which is connected at L and pinches the rubber tube tightly. The tension

at E can be released by an electro-magnet F, which when magnetized

exercises a pull on the iron rod, extends the spring E, and

simultaneously releases the pressure on the rubber tube at L. To make

the control perfectly automatic, the apparatus shown on the top of the

tension-equalizer J is employed. A wire ring, with a wire support, is

caused to pass up through a bearing fastened to the clamp above J. As

the air inside of the whole system becomes diminished in volume and the

rubber cap J sinks, there is a point at which a metal loop dips into two

mercury cups C and C', thus closing the circuit, which causes a current

of electricity to pass through F. This releases the pressure at L,

oxygen rushes in, and the rubber bag J becomes distended. As it is

distended, it lifts the metal loop out of the cups, C and C', and the

circuit is broken. There is, therefore, an alternate opening and closing

of this circuit with a corresponding admission of oxygen. The exact

position of the rubber diaphragm can be read when desired from a pointer

on a graduated scale attached to a support holding the terminals of the

electric wires. More frequently, however, when the volume is required,

instead of filling the bag to a definite point, as shown by the

pointer, a delicate manometer is attached to the can by means of a

pet-cock and the oxygen is admitted by operating the switch B until the

desired tension is reached.

[Illustration: FIG. 32.--Part of the oxygen cylinder and connections to

tension-equalizer. At the left is shown the upper half of the oxygen

cylinder with a detail of the electro-magnet and reducing-valve. D is

the cylinder; K, the band supporting the oxygen cylinder and

electro-magnet arrangement; F, the electro-magnet; E, the tension

spring; and L, the rubber tubing at a point where it is closed by the

clamp. The tension-equalizer and the method of closing the circuit

operating it are shown at the right. C and C' are two mercury cups into

which the wire loop dips, thus closing the circuit. B is a lever used

for short-circuiting for filling the diaphragm J. G is a sulphuric-acid

container; H, the quick-throw valve for shutting off the tension

equalizer J; M, part of the ingoing air-pipe; N, a plug connecting the

electric circuit with the electro-magnet; and O, a storage battery.]

In order to provide for the maximum sensitiveness for weighing D and its

appurtenances, the electric connection is broken at the cylinder by

means of the plug N and the rubber tube is connected by a glass

connector which can be disconnected during the process of weighing.

Obviously, provision is also made that there be no leakage of air out of

the system during the weighing. The current at F is obtained by means of

a storage battery O. The apparatus has been in use for some time in the

laboratory and has proved successful in the highest degree.

Respiration Calorimeters for Studying the Respiratory Exchange and Energy Transformations of Man · 第 90 / 120 题