TESTING THE CHAMBER FOR TIGHTNESS.

As already suggested, the walls are constructed of the largest possible

sheets of copper with a minimum number of seams and opportunities for

leakage. In testing the apparatus for leaks, the greatest precaution is

taken. A small air-pressure is applied and the variations in height of a

delicate manometer noted. In cases of apparent leakage, all possible

sources of leak are gone over with soapsuds when there is a slight

pressure on the chamber. As a last resort, which has ultimately proven

to be the best method of testing, an assistant goes inside of the

chamber, it is then hermetically sealed, and a slight diminished

pressure is produced. Ether is then poured about the walls of the

chamber and the odor of ether soon becomes apparent inside of the

chamber if there is a leakage. Many leaks that could not be found by

soapsuds can be readily detected by this method.