CONNECTIONS TO THERMAL-JUNCTION SYSTEMS.

Since heretofore it has been deemed unwise to attempt to use both

calorimeters at the same time, the electrical connections are so made

that, by means of electrical switches, either calorimeter can be

connected to the apparatus on the table.

The thermal-junction measurements are made by a semicircular switch

S_{7}. The various points, I, II, III, IV, etc., are connected with the

different thermal-junction systems. Thus, by following the wiring

diagram, it can be seen that the connections with I run to the different

binding-posts of the switch S_{10}, which as a matter of fact is placed

beneath the table. This switch S_{10} has three rows of binding-posts.

The center row connects directly with the apparatus on the observer's

table, the outer rows connect with either the chair calorimeter or the

bed calorimeter. The points marked _a_, _b_, _d_, _e_, _f_, etc.,

connect with the bed calorimeter and A, B, D, etc., connect with the

chair calorimeter. Thus, by connecting the points _g_ and _i_ with the

two binding-posts opposite them on the switch S_{10}, it can be seen

that this connection leads directly to the point I on the switch S_{7},

and as a matter of fact this gives direct connection with the

galvanometer through the key on S_{7}, thus connecting the

thermal-junction system on one section of the bed calorimeter between

_g_ and _i_ directly with the galvanometer. Similar connections from the

other points can readily be followed from the diagram. The points on the

switch S_{7} indicated as I, II, III, IV, correspond respectively to the

thermal-junction systems on the top, rear, front, and bottom of the

chair calorimeter.

By following the wiring diagram of the point V, it will be seen that

this will include the connections with the thermal junctions connected

in series and thus give a sum total of the electromotive forces in the

thermal junctions. The point VI is connected with the thermal-junction

system in the air system, indicating the differences in temperature

between the ingoing and outgoing air. It will be noted that there are

four sections in the chair calorimeter, while in the bed calorimeter

there are but three, and hence a special switch S_{3} is installed to

insure proper connections when the bed calorimeter is in use.

This system of connecting the thermal junctions in different sections to

the galvanometer makes possible a more accurate control of the

temperatures in the various parts, and while the algebraic sum of the

temperature differences of the parts may equal zero, it is conceivable

that there may be a condition in the calorimeter when there is a

considerable amount of heat passing out through the top, for example,

compensated exactly by the heat which passes in at the bottom, and while

with the top section there would be a large plus deflection on the

galvanometer, thus indicating that the air around the zinc wall was too

cold and that heat was passing out, there would be a corresponding minus

deflection on the bottom section, indicating the reverse conditions. The

two may exactly balance each other, but it has been found advantageous

to consider each section as a unit by itself and to attempt delicate

temperature control of each individual unit. This has been made possible

by the electrical connections, as shown on the diagram.

Respiration Calorimeters for Studying the Respiratory Exchange and Energy Transformations of Man · 第 57 / 120 题