ELECTRIC-RESISTANCE THERMOMETERS FOR THE WATER-CURRENT.

The measurement of the temperature differences of the water-current by

the electric-resistance thermometer was tried a number of years ago by

Rosa,[10] but the results were not invariably satisfactory and in all

the subsequent experimenting the resistance thermometer could not be

used with satisfaction. More recently, plans were made to incorporate

some of the results of the rapidly accumulating experience in the use of

resistance thermometers and consequently an electric-resistance

thermometer was devised to meet the conditions of experimentation with

the respiration calorimeter by Dr. E. F. Northrup, of the Leeds &

Northrup Company, of Philadelphia. The conditions to be met were that

the thermometers should take rapidly the temperature of the ingoing and

outcoming water and that the fluctuations in temperature difference as

measured by the resistance thermometers should be controlled for

calibration purposes by the differences in temperature of the mercurial

thermometers.

[Illustration: FIG. 16.--Details of resistance thermometers for

water-circuit. Upper part of figure shows a sketch of the outside of the

hard-rubber case. In lower part is a section showing interior

construction. Flattened lead tube wound about central brass tube

contains the resistance wire. A is enlarged part of the case forming a

chamber for the mercury bulb. Arrows indicate direction of flow on

resistance thermometer for ingoing water.]

For the resistance thermometer, Dr. Northrup has used, instead of

copper, pure nickel wire, which has a much higher resistance and thus

enables a much greater total resistance to be inclosed in a given space.

The insulated nickel wire is wound in a flattened spiral and then passed

through a thin lead tube flattened somewhat. This lead tube is then

wound around a central core and the flattened portions attached at such

an angle that the water passing through the tubes has a tendency to be

directed away from the center and against the outer wall, thus insuring

a mixing of the water. Space is left for the insertion of the mercurial

thermometer. With the thermometer for the ingoing water, it was found

necessary to extend the bulb somewhat beyond the resistance coil, so

that the water might be thoroughly mixed before reaching the bulb and

thus insure a steady temperature. Thus it was found necessary to enlarge

the chamber A (fig. 16) somewhat and the tube leading out of the

thermometer, so that the bulb of the thermometer itself could be placed

almost directly at the opening of the exit tube. Under these conditions

perfect mixing of water and constancy of temperature were obtained.

In the case of the thermometer which measured the outcoming water, the

difficulty was not so great, as the outcoming water is somewhat nearer

the temperature of the chamber, and the water as it leaves the

thermometer passes first over the mercurial thermometer and then over

the resistance thermometer. By means of a long series of tests it was

found possible to adjust these resistance thermometers so that the

variations in resistance were in direct proportion to the temperature

changes noted on the mercurial thermometers. Obviously, these

differences in resistance of the two thermometers can be measured

directly with the Wheatstone bridge, but, what is more satisfactory,

they are measured and recorded directly on a special type of automatic

recorder described beyond.

Respiration Calorimeters for Studying the Respiratory Exchange and Energy Transformations of Man · 第 55 / 120 题