HEAT-ABSORBING CIRCUIT.

To bring away the heat produced by the subject, it is highly desirable

that a constant flow of water of even temperature be secured. Direct

connection with the city supply is not practicable, owing to the

variations in pressure, and hence in constructing the laboratory

building provision was made to install a large tank on the top floor,

fed with a supply controlled by a ball-and-cock valve. By this

arrangement the level in the tank is maintained constant and the

pressure is therefore regular. As the level of the water in the tank is

approximately 9 meters above the opening in the calorimeter, there is

ample pressure for all purposes.

[Illustration: FIG. 14.--Schematic diagram of water circuit for

heat-absorbers of calorimeter. A, constant-level tank from which water

descends to main pipe supplying heat-absorbers; _a_, valve for

controlling supply from tank A; B, section of piping passing into cold

brine; _b_, valve controlling water direct from large tank A; _c_, valve

controlling amount of water from cooling section B; C, thermometer at

mixer; D, electric heater for ingoing water; E, thermometer for ingoing

water; _d d d_, heat-absorbers inside calorimeter; F, thermometer

indicating temperature of outcoming water; G, can for collecting water

from calorimeter; _f_, valve for emptying G.]

The water descends from this tank in a large 2-inch pipe to the ceiling

of the calorimeter laboratory, where it is subdivided into three 1-inch

pipes, so as to provide for a water-supply for three calorimeters used

simultaneously, if necessary, and eliminate the influence of a variation

in the rate of flow in one calorimeter upon the rate of flow in another.

These pipes are brought down the inner wall of the room adjacent to the

refrigeration room and part of the water circuit is passed through a

brass coil immersed in a cooling-tank in the refrigeration room. By

means of a by-pass, water of any degree of temperature from 2° C. to 20°

C. may be obtained. The water is then conducted through a pipe beneath

the floor to the calorimeter chamber, passed through the absorbers, and

is finally measured in the water-meter.

A diagrammatic sketch showing the course of the water-current is given

(fig. 14), in which A is the tank on the top floor controlled by the

ball cock and valve, and _a_ is the main valve which controls this

supply to the cooler B, and by adjusting the valve _b_ and valve _c_

any desired mixture of water can be obtained. A thermometer C gives a

rough idea of the temperature of the water, so as to aid in securing the

proper mixture. The water then passes under the floor of the calorimeter

laboratory and ascends to the apparatus D, which is used for heating it

to the desired temperature before entering the calorimeter. The

temperature of the water as it enters the calorimeter is measured on an

accurately calibrated thermometer E, and it then passes through the

absorber system _d d d_ and leaves the calorimeter, passing the

thermometer F, upon which the final temperature is read. It then passes

through a pipe and falls into a large can G, placed upon scales. When

this can is filled the water is deflected for a few minutes to another

can and by opening valve _f_ the water is conducted to the drain after

having been weighed.

_Brine-tank._--The cooling system for the water-supply consists of a

tank in which there is immersed an iron coil connected by two valves to

the supply and return of the brine mains from the central power-house.

These valves are situated just ahead of the valves controlling the

cooling device in the refrigeration room and permit the passage of brine

through the coil without filling the large coils for the cooling of the

air in the calorimeter laboratory. As the brine passes through this

coil, which is not shown in the figure, it cools the water in which it

is immersed and the water in turn cools the coil through which the

water-supply to the calorimeter passes. The brass coil only is shown in

the figure. The system is very efficient and we have no difficulty in

cooling the water as low as 2° C. As a matter of fact our chief

difficulty is in regulating the supply of brine so as not to freeze the

water-supply.

_Water-mixer._--If the valve _b_ is opened, water flows through this

short length of pipe much more rapidly than through the long coil, owing

to the greater resistance of the cooling coil. In conducting these

experiments the valve c is opened wide and by varying the amount to

which the valve _b_ is opened, the water is evenly and readily mixed.

The thermometer C is in practice immersed in the water-mixer constructed

somewhat after the principle of the mixer inside the chamber described

on page 21. All the piping, including that under the floor, and the

reheater D, are covered with hair-felt and well insulated.

_Rate-valves._--It has been found extremely difficult to secure any form

of valve which, even with a constant pressure of water, will give a

constant rate of flow. In this type of calorimeter it is highly

desirable that the rate of flow be as nearly constant as possible hour

after hour, as this constant rate of flow aids materially in maintaining

the calorimeter at an even temperature. Obviously, fluctuations in the

rate of flow will produce fluctuations in the temperature of the ingoing

water and in the amount of heat brought away. This disturbs greatly the

temperature equilibrium, which is ordinarily maintained fairly constant.

Just before the water enters the reheater D it is caused to pass through

a rate-valve, which at present consists of an ordinary plug-cock. At

present we are experimenting with other types of valves to secure even

greater constancy, if possible.

_Electric reheater._--In order to control absolutely the temperature of

the water entering at E, it is planned to cool the water leaving the

water-mixer at C somewhat below the desired temperature, so that it is

necessary to reheat it to the desired point. This is done by passing a

current of electricity through a coil inserted in the system at the

point D. This electric reheater consists of a standard "Simplex" coil,

so placed in the copper can that the water has a maximum circulation

about the heater. The whole device is thoroughly insulated with

hair-felt. By connecting the electric reheater with the rheostat on the

observer's table, control of the quantity of electricity passing through

the coil is readily obtained, and hence it is possible to regulate the

temperature of the ingoing water to within a few hundredths of a degree.

The control of the amount of heat brought away from the chamber is made

either by (1) increasing the rate of flow or (2) by varying the

temperature of the ingoing water. Usually only the second method is

necessary. In the older form of apparatus a third method was possible,

namely, by varying the area of the absorbing surface of the cooling

system inside of the chamber. This last method of regulation, which was

used almost exclusively in earlier experiments, called for an elaborate

system of shields which could be raised or lowered at will by the

operator outside, thus involving an opening through the chamber which

was somewhat difficult to make air-tight and also considerably

complicating the mechanism inside the chamber. The more recent method of

control by regulating the temperature of the ingoing water by the

electric reheater has been much refined and has given excellent service.

_Insulation of water-pipes through the wall._--To insulate the

water-pipes as they pass through the metal walls of the calorimeter and

to prevent any cooling effect not measured by the thermometers presented

great difficulties. The device employed in the Middletown chamber was

relatively simple, but very inaccessible and a source of more or less

trouble, namely, a large-sized glass tube embedded in a large round

wooden plug with the annular space between the glass and wood filled

with wax. An attempt was made in the new calorimeters to secure air

insulation by using a large-sized glass tube, some 15 millimeters

internal diameter, and passing it through a large rubber stopper,

fitting into a brass ferule soldered between the zinc and copper walls.

(See N, fig. 25.) So far as insulation was concerned, this arrangement

was very satisfactory, but unfortunately the glass tubes break readily

and difficulty was constantly experienced. An attempt was next made to

substitute hard-rubber tubing for the glass tube, but this did not prove

to be an efficient insulator. More recently we have used with perfect

success a special form of vacuum-jacketed glass tube, which gives the

most satisfactory insulation. However, this system of insulation is

impracticable when electric-resistance thermometers are used for

recording the water-temperature differences and can be used only when

mercurial thermometers exclusively are employed. The electric-resistance

thermometers are constructed in such a way, however, as to make

negligible any inequalities in the passage of heat through the

hard-rubber casing. This will be seen in the discussion of these

thermometers.

_Measuring the water._--As the water leaves the respiration chamber it

passes through a valve which allows it to be deflected either into the

drain during the preliminary period, or into a small can where the

measurements of the rate of flow can readily be made, or into a large

tank (G, fig. 14) where the water is weighed. The measurement of the

water is made by weight rather than by volume, as it has been found that

the weighing may be carried out with great accuracy. The tank, a

galvanized-iron ash-can, is provided with a conical top, through an

opening in which a funnel is placed. The diagram shows the water leaving

the calorimeter and entering the meter through this funnel, but in

practice it is adjusted to enter through an opening on the side of the

meter. After the valve _f_ is tightly closed the empty can is weighed.

When the experiment proper begins the water-current is deflected so as

to run into this can and at the end of an hour the water is deflected

into a small can used for measuring the rate of flow. While it is

running into this can, the large can G is weighed on platform scales to

within 10 grams. After weighing, the water is again deflected into the

large can and that collected in the small measuring can is poured into G

through the funnel. The can holds about 100 liters of water and

consequently from 3 to 8 one-hour periods, depending upon the rate of

flow, can be continued without emptying the meter. When it is desired to

empty the meter at the end of the period, the water is allowed to flow

into the small can, and after weighing G, the valve _f_ is opened. About

4 minutes are required to empty the large can. After this the valve is

again closed, the empty can weighed, and the water in the small

measuring-can poured into the large can G through the funnel. The scales

used are the so-called silk scales and are listed by the manufacturers

to weigh 150 kilograms. This form of scales was formerly used in

weighing the man inside the chamber.[7]

Respiration Calorimeters for Studying the Respiratory Exchange and Energy Transformations of Man · 第 48 / 120 题