In a Physics Laboratory

_The Problem_

You are a physicist investigating the properties of semiconductors,

which are materials used to make transistors. When you apply a voltage

to one specimen of silicon (a semiconductor), it doesn’t behave quite

like the others that you’ve studied. The electrical properties of this

odd specimen are unusual and interesting and could lead to a new type of

transistor. What makes this specimen different from the others? Very

small amounts of impurities can cause large changes in the electrical

properties of semiconductors. You would like to obtain a chemical

analysis of the material, but your colleagues in chemistry tell you they

would have to dissolve a good size part of your sample to analyze it and

you are reluctant to give it up. How do you do it?

_The Solution_

You decide to try neutron activation analysis. You realize you won’t be

able to detect all the elements, but many of those that might affect

semiconductor performance could be detected quite easily.

What will you need? A source of neutrons to activate the material and a

gamma-ray spectrometer to measure the radiation from the material

afterwards. This spectrometer detects and measures gamma rays and sorts

them according to their energy. You find that your friend down the hall,

who is a nuclear physicist, has a gamma-ray spectrometer that

incorporates a lithium-drifted germanium crystal as a detector and a

pulse height analyzer. The germanium detector is a device that senses

the gamma rays that enter it and gives electrical signals related to the

energy of the gamma rays. It was invented only a few years ago and has a

very fine resolution. That is, it can easily “pick out” gamma rays that

are only slightly different in energy. For example, for gamma rays with

energies of approximately 1 MeV (million electron volts), it is not

unusual to distinguish between gamma rays that differ by only 2 or 3

tenths of a percent. The pulse height analyzer is an electronic device

that sorts the electrical pulses from the detector according to their

energy.

[Illustration: Gamma-ray detector]

A lithium-drifted germanium-crystal gamma-ray detector. The large

container is a reservoir of liquid nitrogen that keeps the detector

cooled to a temperature of -196° Centigrade (321° below zero,

Fahrenheit). The lead brick shield keeps out most of the gamma rays

that come from naturally radioactive materials in the room. The

plastic slots hold cards upon which the samples are mounted for

counting. Sometimes the detector is arranged vertically and samples

are placed on shelves above it.

[Illustration: Gamma-ray detector]

What about the neutrons for the irradiation? Although there isn’t a

suitable nuclear reactor[9] in your city, there is one at a university

only an hour away by jet. Since it may take a few hours to get the

sample to the counter after irradiation, you won’t be able to look for

short-lived activation products, i.e., those with half-lives of up to an

hour. However, this will exclude only a few elements from detection.

[Illustration: Pulse-height analyzer]

A pulse-height analyzer used for gamma-ray spectrometry. A gamma-ray

spectrum is displayed on the television screen. Data is printed out

automatically on the electric typewriter and also may be plotted as a

graph on the paper to the left. In other systems, data may be coded

onto punched paper tape as well. Such tape may be “read” by a computer

that can be programmed to use the data to calculate what radioactive

isotopes are present and their quantities.

Now you are ready to begin the analysis. This will be a qualitative

analysis since you are merely looking for a significantly different

element in that silicon crystal. How much of it is present is only of

secondary interest. Therefore, if you find anything different, you will

rely on an approximate calculation to tell you “how much”.

[Illustration: Nuclear reactor]

This is called a “swimming pool” reactor because the nuclear fuel,

built into metal rods, is held in a framework at the bottom of a deep

pool of water. The water serves as a shield to protect workers from

the radiation and also helps the reactor “go” by slowing down neutrons

to make them more likely to interact with the target atoms. “Swimming

pool” reactors are frequently used for neutron activation analysis and

typically provide neutron fluxes of over 10¹³ (10 million million)

neutrons per square centimeter per second.

[Illustration: Quartz capsules]

These sealed quartz capsules contain samples to be irradiated in a

nuclear reactor. They are about to be placed in the aluminum can,

which will be sealed and positioned at the end of an aluminum pole,

close to the core of a “swimming pool” reactor. Often samples are

placed in plastic tubes and are carried in and out of a reactor by air

pressure in a pneumatic tube system.

You carefully scrape off a small amount of material, weigh it on a

sensitive balance, and put it into a short piece of pure quartz tubing.

You do the same with an ordinary piece of silicon for comparison and

then seal both tubes with an oxygen-gas torch. Although the tubes are

both ¼ inch in diameter and about 1 inch long, the first tube is just

slightly longer so you will be able to determine which is which after

the irradiation.

Off it goes to the reactor in a carefully wrapped package along with

instructions to irradiate the tubes for 12 hours in a neutron flux of

about 10¹³ neutrons per square centimeter per second and to return them

as quickly as possible after they are removed from the reactor.

The following week, the samples are delivered about 4 hours after they

were removed from the reactor. Working quickly but carefully, you note

that they are radioactive but easily handled by ordinary laboratory

techniques. You break the quartz tubes one at a time and attach each of

the two pieces of silicon to a card with self-sticking tape. Then you

place each card, in turn, on a holder close to the gamma-ray detector

for a period of 10 minutes. A spectrum, which is a graph of the quantity

of radiation recorded in each increment of energy over the range

observed for each of the samples, is plotted automatically at the end of

the counting period and you may now compare the compositions of the two

samples. (See the figure on the next two pages.)

The two spectra are virtually identical except that the suspect sample

has one obviously different peak in channel 157 and a somewhat smaller

peak in channel 183. Referring to an energy calibration curve for the

pulse height analyzer, you find that these channels correspond to 0.559

and 0.657 MeV respectively. A search of a table of nuclides, arranged by

gamma-ray energy, reveals that this combination is emitted by

arsenic-76, which would be the activation product for arsenic. Other

data also indicate that for arsenic there should be a number of smaller

peaks, including some corresponding to energies of 1.216, 1.228, 0.624,

and 1.441 MeV. A closer look at the spectrum of the suspect sample

reveals that these are also present.

Finally, noting that the half-life of arsenic-76 is approximately 27

hours, you wait a day and count the sample again in the same position as

the previous count. A decrease in the heights of the 0.559 and 0.657 MeV

peaks, by a little less than half in 24 hours, confirms that arsenic is

the unusual element in this sample. It may not be the only impurity

causing the peculiar behavior of this semiconductor, but it does seem a

likely candidate.

[Illustration: Graph: “Counts in 20 minutes per 3.8 KeV channel” _vs_

“Channel Number”]

Element Counts Channel Number

Gold 1300 125

Sodium 900 140

Antimony 650 150

Antimony 550 160

Scandium 220 240

Iron 150 300

Scandium 60 310

Cobalt 110 320

Iron 45 350

Cobalt 50 360

Sodium 900 370

Sodium 140 425

(values estimated from the graph)

The gamma-ray spectrum obtained after activation of a sample of “pure”

silicon having “ordinary” properties of this type of semiconductor.

Only very small quantities of various trace impurities are indicated.

[Illustration: Graph: “Counts in 20 minutes per 3.8 KeV channel” _vs_

“Channel number”]

Element Counts Channel Number

Gold

Sodium

0.559 Arsenic-75 1500 150

Antimony

0.624 Arsenic-75 190 165

0.657 Arsenic-75 135 175

Scandium

Iron

Scandium

Cobalt

1.216 Arsenic-75 100 330

1.228 Arsenic-75 60 335

Iron

Cobalt

Sodium

1.44 Arsenic-75 25 380

Sodium

(distinct values estimated from graph)

The gamma-ray spectrum obtained after activation of a sample of

silicon having “unusual” electrical properties. While most of the

spectrum is identical with that obtained from the ordinary material,

there is an interesting difference.

Using the equation given on page 12, the approximate known values for

half-life, sample weight, neutron flux, and periods of irradiation and

decay after irradiation, and an estimated value for the number of

arsenic-76 atoms measured by the gamma-ray spectrometer, you calculate

that the arsenic content of the sample is approximately 44 parts per

million (ppm). (See appendix.)

With this information as a starting point, you are now ready to proceed

with further research on the properties of your semiconductor, e.g., if

you double the concentration of arsenic, how will that affect its

properties?