Rotate the eikonometer until the lines on the two scales are parallel,

and make the various adjustments to ensure that two lines on the

eikonometer scale coincide with two lines on the stage micrometer.

For the sake of illustration it may be assumed that five of the

divisions on the stage micrometer accurately fill one of the divisions

of the eikonometer scale; this indicates a magnifying power of 500 as

the constant for that particular optical combination, and a record

should be made of the fact.

The magnification constants of the various other optical combinations

should be similarly made and recorded.

To measure any object subsequently it should be first focussed carefully

in the ordinary way.

The eikonometer should then be applied to the eyepiece and the size of

the object read off on the eikonometer scale as millimetres, and the

actual size calculated by dividing the observed size by the

magnification constant for the particular optical combination employed

in the observation.

(c) By means of the filar micrometer.

[Illustration: FIG. 58.--Ramsden's Filar micrometer.]

[Illustration: FIG. 59.--Ramsden's micrometer field, a, fixed wire;

b, reference wire (fixed); c, travelling wire.]

The ~Filar~ or cobweb Micrometer (Ramsden's micrometer) eyepiece (Fig. 58)

consists of an ocular having a fine "fixed" wire stretching horizontally

across the field (Fig. 59), a vertical reference wire--fixed--adjusted

at right angles to the first; and a fine wire, parallel to the reference

wire, which can be moved across the field by the action of a micrometer

screw; the drum head is divided into one hundred parts, which

successively pass a fixed index as the head is turned. In the lower part

of the field is a comb with the intervals between its teeth

corresponding to one complete revolution of this screw-head.

As in the previous method, the value of each division of the micrometer

scale (i. e., the comb) must first be determined for each optical

combination. This is effected as follows:

1. Place the filar micrometer and the stage micrometer in their

respective positions.

2. Rotate the screw of the filar micrometer until the movable wire

coincides with the fixed one, and the index marks zero on the drum head.

(If when the drum head is at zero the two wires do not exactly coincide

they must be adjusted by loosening the drum screw and resetting the

drum.)

3. Focus the scale of each micrometer accurately, and make the lines on

them parallel.

4. Rotate the head of the micrometer screw until the movable line has

transversed one division of the stage micrometer. Note the number of

complete revolutions (by means of the recording comb) and the fractions

of a revolution (by means of scale on the head of the micrometer screw),

which are required to measure the 0.01 mm.

5. Make several such estimations and average the results.

6. Note the optical combination employed in this experiment and record

it carefully, together with the micrometer value in terms of mu.

7. Repeat this process for each of the different optical combinations

and record the results.

To measure an object by this method, simply note the number of

revolutions and fractions of a revolution of the screw-head required to

traverse such object from edge to edge, and express the result as

_micra_ by reference to the recorded values for that particular optical

combination.

_Microscope Illuminant._--In tropical and subtropical regions diffuse

daylight is the best illuminant. In temperate climes however daylight of

the desirable quantity is not always available, and recourse must be

had to oil lamps, gas lamps--preferably those with incandescent

mantles--and electricity; and of these the last is undoubtedly the best.

A handy lamp holder which can be manufactured in the laboratory is shown

in Fig. 60. It consists of a base board weighted with lead to which is

attached the ordinary domestic lamp holder, and behind this is fastened

a curved sheet-iron reflector. An obscured metal filament lamp of about

16 candle power gives the most suitable light, and if monochromatic

light is needed, the blue grease pencil is streaked over the side of the

lamp nearest the microscope; the current is switched on and when the

glass bulb is warm, rubbing with a wad of cotton-wool will readily

distribute the blue greasy material in an even film over the ground

glass.

[Illustration: FIG. 60.--Electric microscope lamp.]

FOOTNOTES:

[1] Its importance will be realised, however, when it is stated in the

words of the late Professor Abbe: "The numerical aperture of a lens

determines all its essential qualities; the brightness of the image

increases with a given magnification and other things being equal, as

the square of the aperture; the resolving and defining powers are

directly related to it, the focal depth of differentiation of depths

varies inversely as the aperture, and so forth."

[2] Made by Mr. Otto Baumbach, 10, Lime Grove, Manchester.

V. MICROSCOPICAL EXAMINATION OF BACTERIA AND OTHER MICRO-FUNGI.

APPARATUS AND REAGENTS USED IN ORDINARY MICROSCOPICAL EXAMINATION.

The following comprises the essential apparatus and reagents for routine

work with which each student should be provided.

1. India-rubber "change-mat" upon which cover-glasses may be rested

during the process of staining.

2. Squares of blotting paper about 10 cm., for drying cover-slips and

slides.

(The filter paper known as "German lined"--a highly absorbent, closely

woven paper, having an even surface and no loose "fluff" to adhere to

the specimens--is the most useful for this purpose.)

[Illustration: FIG. 61.--Disinfectant Jar.]

3. Glass jar filled with 2 per cent. lysol solution for the reception of

infected cover-glasses and infected pipettes, etc.

4. A square glazed earthenware box with a loose lining containing 2 per

cent. lysol solution for the reception of infected material and used

slides. The bottom of the lining is perforated so that when full the

lining and its contents can be lifted bodily out of the box, when the

disinfectant solution drains away and the slides, etc., can easily be

emptied out. The empty lining is then returned to the box with its

disinfectant solution (Fig. 61).

5. Bunsen burner provided with "peep-flame" by-pass.

6. Porcelain trough holding five or six hanging-drop slides (Fig. 62).

[Illustration: FIG. 62.--Hanging-drop slides: a, Double cell seen from

above; b, single cell seen from the side.]

The best form of hanging-drop slide is a modification of Boettcher's

glass ring slide, and is prepared by cementing a circular cell of tin,

13 to 15 mm. diameter, and 1 to 2 mm. in height, to the centre of a 3 by

1 slip by means of Canada balsam. It is often extremely convenient to

have two of these cells cemented close together on one slide (Fig. 62,

a).

Another form of hanging-drop slide is made in which a

circular or oval concavity or "cell" is ground out of the

centre of a 3 by 1 slip. These are more expensive, less

convenient to work with, and are more easily contaminated by

drops of material under examination, and should be carefully

avoided.

7. Three aluminium rods (Fig. 63), each about 25 cm. long and carrying a

piece of 0.015 gauge platino-iridium wire 7.5 cm. in length. The end of

one of the wires is bent round to form an oval loop, of about 1 mm. in

its short diameter, and is termed a loop or an oese; the terminal 3 or 4

mm. of another wire is flattened out by hammering it on a smooth iron

surface to form a "spatula"; the third is left untouched or is pointed

by the aid of a file. These instruments are used for inoculating culture

tubes and preparing specimens for microscopical examination.

[Illustration: FIG. 63.--Ends of platinum rods. a, loop; b, spatula;

c, needle.]

The method of mounting these wires may be described as follows:

Take a piece of aluminium wire 25 cm. long and about 0.25 cm. in

diameter, and drill a fine hole completely through the wire about a

centimetre from one end. Sink a straight narrow channel along one side

of the wire, in its long axis, from the hole to the nearest end, shallow

at first, but gradually becoming deeper.

On the opposite side of the wire make a short cut, 2 mm. in length,

leading from the hole in the same direction. [The use of a fine dental

drill and small circular saw, worked by a dental motor facilitates the

manufacture of these aluminium handled instruments.]

Now pass one end of the platinum wire through the hole, turn up about 2

mm. at right angles and press the short piece into the short cut. Turn

the long end of the wire sharply, also at right angles, and sink it into

the long channel so that it emerges from about the centre of the cut end

of the aluminium wire (Fig. 63). A few sharp taps with a watch maker's

hammer will now close in the sides of the two channels over the wire and

hold it securely.

[Illustration: FIG. 64.--Platinum rod in aluminium handle--method of

mounting.

The platinum wire may be fused into the end of a piece of glass rod, but

such a handle is vastly inferior to aluminium and is not to be

recommended.]

8. Two pairs of sharp-pointed spring forceps (10 cm. long), one of which

must be kept perfectly clean and reserved for handling clean

cover-slips, the other being for use during staining operations.

9. A box of clean 3 by 1 glass slips.

10. A glass capsule with tightly fitting (ground on) glass lid,

containing clean cover-slips in absolute alcohol.

11. One of Faber's "grease pencils" (yellow, red, or blue) for writing

on glass.

12. A wooden rack (Fig. 65) with twelve drop-bottles (Fig. 66) each 60

c.c. capacity, containing

Aniline water.

Gentian violet, saturated alcoholic solution.

Lugol's (Gram's) iodine.

Absolute alcohol.

Methylene-blue, }

Fuchsin, basic, } saturated alcoholic solution.

Neutral red, 1 per cent. aqueous solution.

Leishman's modified Romanowsky stain.

Carbolic acid, 5 per cent. aqueous solution.

Acetic acid, 1 per cent. solution.

Sulphuric acid, 25 per cent. solution.

Xylol.

[Illustration: FIG. 65.--Staining rack, rubber change mat and lysol

pot.]

[Illustration: FIG. 66.--Drop bottle.]

[Illustration: FIG. 67.--Canada balsam pot.]

And two pots with air-tight glass caps (Fig. 67), each provided with a

piece of glass rod and filled respectively with Canada balsam dissolved

in xylol, and sterile vaseline.

METHODS OF EXAMINATION.

Bacteria, etc., are examined microscopically.

1. In the living state, unstained, or stained.

2. In the "fixed" condition (i. e., fixed, killed,

and stained by suitable methods).

The preparation of a specimen from a tube cultivation for examination by

these methods may be described as follows:

~1. Living, Unstained.~--(a) _"Fresh" Preparation._--

1. Clean and dry a 3 by 1 glass slip and place it on one of the squares

of filter paper. Deposit a drop of water (preferably distilled) or a

drop of 1 per cent. solution of caustic potash, on the centre of the

slip, by means of the platinum loop.

[Illustration: FIG. 68.--Holding tubes for removing bacterial growth, as

seen from the front.]

TECHNIQUE OF OPENING AND CLOSING A CULTURE TUBE.

2. Remove the tube cultivation from its rack or jar with the

left hand and ignite the cotton-wool plug by holding it to

the flame of the Bunsen burner. Extinguish the flame by

blowing on the plug, whilst rotating the tube on its long

axis, its mouth directed vertically upward, between the

thumb and fingers. (This operation is termed "flaming the

plug," and is intended to destroy any micro-organisms that

may have become entangled in the loose fibres of the

cotton-wool, and which, if not thus destroyed, might fall

into the tube when the plug is removed and so accidentally

contaminate the cultivation.)

3. Hold the tube at or near its centre between the ends of

the thumb and first two fingers of the left hand, and allow

the sealed end to rest upon the back of the hand between the

thumb and forefinger, the plug pointing to the right. Keep

the tube as nearly in the horizontal position as is

consistent with safety, to diminish the risk of the

accidental entry of organisms (Fig. 68).

4. Take the handle of the loop between the thumb and

forefinger of the right hand, holding the instrument in a

position similar to that occupied by a pen or a paint-brush,

and sterilise the platinum portion by holding it in the

flame of a Bunsen burner until it is red hot. Sterilise the

adjacent portion of the aluminium handle by passing it

rapidly twice or thrice through the flame. After sterilising

it, the loop must not be allowed to leave the hand or to

touch against anything but the material it is intended to

examine, until it is finished with and has been again

sterilised.

5. Grasp the cotton-wool plug of the test-tube between the

little finger and the palm of the right hand (whilst still

holding the loop as directed in step 4), and remove it from

the mouth of the tube by a "screwing" motion of the right

hand.

6. Introduce the platinum loop into the tube and hold it in

this position until satisfied that it is quite cool. (The

cooling may be hastened by touching the loop on one of the

drops of moisture which are usually to be found condensed on

the interior of the glass tube, or by dipping it into the

condensation water at the bottom; at the same time care must

be taken in the case of cultures on solid media to avoid

touching either the medium or the growth.)

7. Remove a small portion of the growth by taking up a drop

of liquid, in the case of a fluid culture, in the loop; or

by touching the loop on the surface of the growth when the

culture is on solid medium; and withdraw the loop from the

tube without again touching the medium or the glass sides of

the tube.

8. Replace the cotton-wool plug in the mouth of the tube.

9. Replace the tube cultivation in its rack or jar.

10. Mix the contents of the loop thoroughly with the drop of water on

the 3 by 1 slide.

11. Again sterilise the loop as directed in step 4, and replace it in

its stand.

12. Remove a cover-slip from the glass capsule by means of the

cover-slip forceps, rest it for a moment on its edge, on a piece of

filter paper to remove the excess of alcohol, then pass it through the

flame of the Bunsen burner. This burns off the remainder of the alcohol,

and the cover-slip so "flamed" is now clean, dry, and sterile.

13. Lower the cover-slip, still held in the forceps, on to the surface

of the drop of fluid on the 3 by 1 slip, carefully and gently, to avoid

the inclusion of air bubbles.

14. Examine microscopically (_vide infra_).

During the microscopical examination, stains and other reagents may be

run in under a cover-slip by the simple method of placing a drop of the

reagent in contact with one edge of the cover-glass and applying the

torn edge of a piece of blotting paper to the opposite side. The reagent

may then be observed to flow across the field and come into contact with

such of the micro-organisms as lie in its path.

The non-toxic basic dyes most generally employed for the intra-vitam

staining of bacteria are

Neutral red, }

Quinoleine blue }

Methylene green } in 0.5 per cent. aqueous solutions.

Vesuvin, }

_Negative Stain_ (Burri).--By this method of demonstration the

appearances presented by dark ground illumination (by means of a

paraboloid condenser) are closely simulated, since minute particles,

bacteria, blood or pus cells etc. stand out as brilliantly white or

colourless bodies on a dark grey-brown background.

_Reagent required:_

Any one of the liquid waterproof black drawing inks (Chin-chin, Pelican,

etc.). This is prepared for use as follows:

Measure out and mix:

Liquid black ink, 25 c.c.

Tincture of iodine 1 c.c.

Allow the mixture to stand 24 hours, centrifugalise thoroughly, pipette

off the supernatant liquid to a clean bottle and then add a crystal of

thymol or one drop of formalin as a preservative.

METHOD.--

1. With the sterilised loop deposit one drop of the liquid ink close to

one end of a 3 by 1 slide.

2. With the sterilised loop deposit a drop of the fluid culture (or of

an emulsion from a solid culture) by the side of the drop of ink (Fig.

69, a); mix the two drops thoroughly by the aid of the loop.

3. Sterilise the loop.

4. Hold the slide firmly on the bench with the thumb and forefinger of

the left hand applied to the end nearest the drop of fluid.

5. Take another clean 3 by 1 slide in the right hand and lower its short

end obliquely (at an angle of about 60 deg.) transversely on to the mixed

ink and culture on the first slide, and allow the fluid to spread across

the slide and fill the angle of incidence.

6. Maintaining the original angle, draw the second slide firmly and

evenly along the first toward the end farthest from the left hand (Fig.

69, b).

7. Throw the second slide into a pot of disinfectant; allow the first

slide to dry in the air.

[Illustration: FIG. 69.--Spreading negative film.]

8. Place a drop of immersion oil on the centre of the film, lower the

1/12-inch objective into the oil and examine microscopically without the

intervention of a cover-slip.

(The film of ink may be covered with a long cover-glass and xylol balsam

as a permanent preparation.)

(<b) _Hanging-drop Preparation._--

1. Smear a layer of sterile vaseline on the upper surface of the ring

cell of a hanging-drop slide by means of the glass rod provided with the

vaseline bottle, and place the slide on a piece of filter paper.

2. "Flame" a cover-slip and place it on the filter paper by the side of

the hanging-drop slide.

3. Place a drop of water on the centre of the cover-slip by means of the

platinum loop.

4. Obtain a small quantity of the material it is desired to examine, in

the manner detailed above (pages 74-76, steps 2 to 11 must be followed

in their entirety and with the strictest exactitude whenever tube

contents are being handled), and mix it with the drop of water on the

cover-slip.

5. Raise the cover-slip in the points of the forceps and rapidly invert

it on to the ring cell of the hanging-drop slide, so that the drop of

fluid occupies the centre of the ring. (Carefully avoid contact between

the drop of fluid and either the ring cell or the layer of vaseline.

Should this happen, the now _infected_ hanging-drop slide and its

cover-slip must be dropped into the pot of lysol and a new preparation

made.)

6. Press the cover-slip firmly down into the vaseline on to the top of

the ring cell. (This spreads out the vaseline into a thin layer, and

besides ensuring the adhesion of the cover-slip, seals the cells and so

retards evaporation.)

7. Examine microscopically.

The examination of a "fresh" specimen or a "hanging-drop" preparation is

directed to the determination of the following data:

1. The nature of the bacteria present--e. g., cocci, bacilli, etc.

2. The purity of the cultivation; this can only be determined when gross

morphological differences exist between the organisms present.

3. The presence or absence of spores; when present, spores show their

typical refrangibility exceedingly well by this method.

4. The presence or absence of mobility. In a hanging-drop specimen some

form of movement can practically always be observed, and its character

must be carefully determined by noting the relative positions of

adjacent micro-organisms.

(a) Brownian or molecular movement. Minute particles of solid matter

(including bacteria), when suspended in a fluid, will always show a

vibratory movement affecting the entire field, but never altering the

relative positions of the bacteria. (Cocci exhibit this movement, but

with the exception of the Micrococcus agilis, the cocci are non-motile.)

(b) Streaming movement. This is due to currents set up in the hanging

drop as a result of jarring of the specimen or of evaporation, or to the

fact that the cover-slip is not perfectly level, and although the

relative positions of the bacteria may vary, still the flowing movement

of large numbers of organisms in some one direction will usually be

sufficient to demonstrate the nature of this motion.

(c) Locomotive movement, or ~true motility~, is determined by observing

some one particular bacillus changing its position in the field

independently of, and in a direction contrary to, other organisms

present.

When the examination is completed and the specimen finished with, the

"fresh specimen"--i. e., the slide with the cover-slip attached--must

be dropped into the lysol pot. In the hanging-drop specimen, however,

the cover-slip only is infected, and this may be raised from the ring

cell by means of forceps and dropped into the disinfectant.

_Permanent Staining of the Hanging-drop Specimen._--Occasionally it is

necessary to fix and stain a hanging-drop preparation. This may be done

as follows:

1. Remove the cover-slip from the cell by the aid of the forceps.

2. If the drop is small, fix it by dropping it face downward, whilst

still wet, on to the surface of some Gulland's solution or corrosive

sublimate solution (_vide_ page 82) in a watch-glass. If the drop is

large, place it face upward on the rubber mat, cover it with an inverted

watch-glass, and allow it to dry. Then fix it in the alcohol and ether

solution (_vide_, page 82).

3. Dip the cover-glass into a beaker containing hot water in order to

remove some of the vaseline adhering to it.

4. Wash successively in alcohol, xylol, ether, and alcohol, to remove

the last traces of grease.

5. Wash in water.

6. Stain, wash, dry, and mount as for an ordinary cover-slip film

preparation (_vide_ pages 83-85).

~2. Killed, Stained.~--In this method three distinct processes are

necessary:

"Preparing" and "fixing" the film.

Staining.

Mounting.

_Preparing the Film._--

1. Flame a cover-slip and place it on a piece of filter paper.

2. Place a drop of water on the centre of the cover-slip by means of

platinum loop.

3. Obtain a small quantity of the material to be examined upon a

sterilised platinum loop (see pages 74-76, steps 2 to 11) and mix it

with the drops of water on the cover-slip.

4. Spread the drop of emulsion evenly over the cover-slip in the form of

a square film to within 1 mm. of each edge of the cover-slip.

5. Allow it to dry completely in the air.

_Fixing._--Fix by passing the cover-slip, held in the fingers, three or

four times through the flame of a Bunsen burner.

In some instances (e. g., when the films after staining are intended

for micrometric observations) it is almost essential to fix by exposure

to a uniform temperature of 115 deg. C., for twenty minutes. This is best

done in a carefully regulated hot-air oven.

Fixation may also be effected by immersing in some fixative fluid, such

as one of the following:

1. Absolute alcohol, for five to fifteen minutes.

{ equal parts, for five to thirty

2. Absolute alcohol, { minutes (e. g., for blood or

Ether, { milk).

3. Osmic acid, 1 per cent. aqueous solution, for thirty seconds.

4. Corrosive sublimate, saturated aqueous solution, for five minutes.

5. Corrosive sublimate (Lang), for five minutes. This solution is

prepared by dissolving:

Sodium chloride 0.75 gramme

Hydrarg. perchloride 12.00 grammes

Acetic acid 5.00 grammes

In distilled water 100.00 c.c.

Filter.

6. Gulland's solution, for five minutes. This solution is prepared by

mixing:

Absolute alcohol 25.0 c.c.

Ether 25.0 c.c.

Corrosive sublimate, 20 per cent. alcoholic solution 0.4 c.c.

7. Formalin 10 per cent. aqueous solution (= 4 per cent. aqueous

solution of formaldehyde since formalin is a 40 per cent. solution of

the gas in water).

Either of these methods of fixation coagulates the albuminous material

and ensures perfect adhesion of the film to the cover-slip.

_Clearing._--Wash the cover-slip thoroughly in running water and proceed

with the staining.

If the film has been prepared from broth, liquefied gelatine, or pus or

other morbid exudations, saturate the film after fixation with acetic

acid 2 per cent. and allow it to act for two minutes.

Wash with alcohol, then let the alcohol remain on the cover-slip for two

minutes. (This will "clear" the groundwork and give a much sharper and

cleaner film than would otherwise be obtained.)

If the film has been prepared from blood or bloodstained fluid, treat

with acetic acid 2 per cent. for two minutes after fixation. Wash with

water, dry, and proceed with the staining. (This will remove the

haemoglobin and facilitate examination.)

The elements of bacteriological technique · 第 4 / 33 段